Equipment and Calibration History

Page: 1 2 3 4 [5] 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68
Ser No.DateLoc FromTitleLoc ToNotes
EXO2 22C103416 2024-04-04 Hill Slough Water Level Removed Lab none Removed from the Hill Slough channel and brought back to the lab for calibration (and to install the new auto wiper that arrived)
LI-7500A 75H-2180 2024-03-25 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at Bouldin Corn to replace sn 0065 for calibration.
LI-7500A 0065 2024-03-25 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 75H-2180.
CS450 20010026 2024-03-25 West Pond Water Level Removed Lab none Removed from WP and brought back to the lab. This sensor is needed at Mayberry, so we are slowly decommissioning West Pond.
HMP45AC W0510026 2024-03-26 Tonzi RH & Temp Floor Removed Lab none Removed for Tonzi Floor met due to age and poor RH data. Replaced by new Hygrovue a few weeks ago.
D1 92243047 2024-03-26 Lab none Installed Tonzi-ptden Dendrometer LiLPine Installed on the little pine next to cluster oak 1 at breast height just below the first set of branches
LI-7500A 0073 2024-03-26 Lab none Installed Tonzi-Tower Gas Analyzer CO2/H2O Installed at Tonzi-Tower to replace sn 75H-2669 for calibration.
LI-7500A 75H-2669 2024-03-26 Tonzi-Tower Gas Analyzer CO2/H2O Removed Lab none Removed from Tonzi-Tower for calibration. Replaced by sn 0073.
LI-7500A 75H-2176 2024-03-26 Lab none Installed Tonzi Gas Analyzer CO2/H2O Floor Installed at Tonzi Floor to replace sn 75H-2667 for calibration.
LI-7500A 75H-2667 2024-03-26 Tonzi Gas Analyzer CO2/H2O Floor Removed Lab none Removed from Tonzi Floor for calibration. Replaced by sn 75H-2176.
HMP45AC W0730037 2024-03-26 Vaira RH & Temp Lab none Removed from Vaira. A Hygrovue was installed a few weeks ago for cross comparison. This HMP45 will probably not be used again.
LI-7500A 0042 2024-03-26 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0035 for calibration.
LI-7500A 0035 2024-03-26 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0042.
eosFD FD20160001 2024-03-26 Lab none Installed Vaira Forced Diffusion North Installed in the north location at Vaira after factory repair.
LI-7500A 75H-2180 2024-03-23 Lab Calibration Lab Last calibration: 2024-01-03
AGC=101.4 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.0kPa, Tair=24.65C, Tdew=11.42C
Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00
CO2 zero 0.9141 -> .9137 Offset: -1.89ppm
CO2 span 1.0406 -> 1.0356 Offset: -2.97ppm
H2O zero 0.8965 -> .8951 Offset: -8.2ppm
H2O span 1.0383 -> 1.0393 Offset: -.01ppm
Notes: Used "524.52" as span point in software, otherwise, the post-span reading was too high
LI-7500A 75H-2176 2024-03-23 Lab Calibration Lab Last calibration: 2023-11-03
AGC=102.1 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.1kPa, Tair=25.05C, Tdew=10.57C
Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00
CO2 zero 0.8678 -> .8682 Offset: 1.94ppm
CO2 span 1.0116 -> 1.0079 Offset: 2.58ppm
H2O zero 0.8552 -> .8494 Offset: -33.2ppm
H2O span .9231 -> .9236 Offset: -.01ppm
Notes: Larger H2O zero offset than usual
LI-7500A 0073 2024-03-23 Lab none Calibration Lab none Last calibration: 2023-10-05
AGC=103 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.2kPa, Tair=25.2C, Tdew=10.4C
Zero Air at 325psi, AMP CA04119 524.77 at 550 Tdewspan=20.00
CO2 zero 0.9167 -> .9159 Offset: -3.14ppm
CO2 span 1.0064 -> 1.0067 Offset: -4.95ppm
H2O zero 0.8296 -> .8323 Offset: 15.6ppm
H2O span 1.0109 -> 1.0195 Offset: -.19ppm
Notes: Slightly larger CO2 span offset than usual. Painted some chips in the paint with nail polish
LI-7500A 0042 2024-03-23 Lab none Calibration Lab none Last calibration: 2023-10-05
AGC=100.8 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.08kPa, Tair=24.9C, Tdew=9.91C
Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00
CO2 zero 0.9473 -> .9483 Offset: 3.82ppm
CO2 span 1.0033 -> 1.0098 Offset: -4.49ppm
H2O zero 0.8474 -> .8451 Offset: -13.6ppm
H2O span 1.0075 -> 1.02 Offset: -.28ppm
Notes: H2O span was oscillating within 1ppm (but average value was stable), but this went away after adding more DI water to the dewpoint generator reservoir. CO2 span seems larger than usual. Painted some chips in the paint with nail polish
EXO pH-ORP 22D102641 2024-01-25 Lab none Installed Hill Slough pH-ORP Installed in the Hill Slough channel after calibration
EXO Wiped Cond/Temp 23D104030 2024-01-25 Lab none Installed Hill Slough Conductivity Sensor Installed in the Hill Slough channel after calibration
EXO fDOM 22E100649 2024-01-25 Lab none Installed Hill Slough fDOM Installed in the Hill Slough channel after calibration
EXO Turbidity 22E106072 2024-01-25 Lab none Installed Hill Slough Turbidity Installed in the Hill Slough channel after calibration
EXO2 22C103416 2024-01-25 Lab none Installed Hill Slough Water Level Installed in the Hill Slough channel after calibration
LI-7500A 0041 2004-04-07 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0065 for calibration
LI-7500A 0065 2004-04-07 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0041
LI-7500A 0041 2001-03-22 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 0065 for calibration.
LI-7500A 0065 2001-03-22 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0041.
D1 92243042 2024-03-07 Lab none Installed Tonzi-ptden Dendrometer Pine11833 Install on a large pine id 11833 to the south of the tower near cluster tree 7.
HFP01SC 00068 2024-03-07 Tonzi Heat Flux Plate D Moved Tonzi Heat Flux Plate D I moved this plate back down to -10cm where I found it before fixing its cable a couple weeks ago.
HFP01 000138 2024-03-07 Tonzi Heat Flux Plate A Maintenance Tonzi Heat Flux Plate A We dug this plate up to replace its cable that had all its insulation rotted off. The plate was 2-5cm below the surface – it was tilted and ground surface was not very flat. We put it back flat at 3cm and the new cable in conduit.
D1 92243044 2024-03-07 Lab none Installed Tonzi-ptden Dendrometer OakLK Installed on Liukang's tree (92 in leaf water potential data)
D1 92243043 2024-03-07 Lab none Installed Tonzi-ptden Dendrometer Oak102 Installed leaf water potential oak 102
D1 92243045 2024-03-07 Lab none Installed Tonzi-ptden Dendrometer C5 Installed on Cluster Oak 5.
SONIC-ANEMO-SPE10 022365359 2024-03-07 Lab none Installed Gilbert Tract Weather Station Installed as an accessory to the Arable Mark 3 sn D006946
Mark 3 D006946 2024-03-07 Lab none Installed Gilbert Tract Weather Station Installed at Gilbert Tract on the radiometer boom.
SONIC-ANEMO-SPE10 022365416 2024-02-09 Lab none Installed Vaira Weather Station B Accessory for the Arable Mark 4 sn D006920.
Mark 3 D006920 2024-02-09 Lab none Installed Vaira Weather Station B Installed at Vaira to replace sn B301042. I believe the older B unit was a prototype and the newer D unit is production grade
Mark 3 B301042 2024-02-09 Vaira Weather Station B Removed Lab none Removed from Vaira. Replaced by sn D006920 at Arable's request. I believe the older B unit was a prototype and the newer D unit is production grade
HFP01SC 00068 2024-02-16 Tonzi Heat Flux Plate D Maintenance Tonzi Heat Flux Plate D I dug this plate up. It was at about -10cm with voids below the plate. I soldered on a new cable amd cut off the calibration cable because the insulation on both was completely rotten. I put it back at -2cm.
EXO2 22C103416 2024-03-06 Factory none Received Lab none Received wiper sn 13A100277 (firmware 1.0.86) from YSI after repair and calibration. Came with an extra brush and a small allen key for installing the brush. The factory inspection reported that water had leaked into the probe housing and all internal components were corroded. They replaced the wiper under warranty. YSI recommends sending the wiper back to the factory every year to have wiper motor o-rings replaced and lubed to prevent water leakage.
RPI0W RPI0W-007 2023-05-31 Lab none Installed Bouldin Corn none Installed in the corn field at BC for the 2023 growing season
RPI0W RPI0W-002 2023-05-31 Lab none Installed Bouldin Corn none Installed in the corn field at BC for the 2023 growing season
SRS-Ni NDVI 2046503357 2024-02-28 Bouldin Corn Narrow Band NDVI Reflected Removed Lab none Removed from BC tower because most of the SRS NDVI sensors installed in the Delta have been drifting over the past few years
SRS-Ni NDVI 882103020 2024-02-28 Bouldin Corn Narrow Band NDVI Incoming Removed Lab none Removed from BC tower because most of the SRS NDVI sensors installed in the Delta have been drifting over the past few years
SRS-Ni NDVI 2046503344 2024-02-28 East End Narrow Band NDVI Reflected Removed Lab none Removed from EE tower because most of the SRS NDVI sensors installed in the Delta have been drifting over the past few years
SRS-Ni NDVI 2046503390 2024-02-28 East End Narrow Band NDVI Incoming Removed Lab none Removed from EE tower because most of the SRS NDVI sensors installed in the Delta have been drifting over the past few years
SRS-Ni NDVI 882103059 2024-02-28 Gilbert Tract Narrow Band NDVI Incoming Removed Lab none Removed from GT tower because most of the SRS NDVI sensors have been drifting in the past few years
SRS-Ni NDVI 2046503404 2024-02-28 Gilbert Tract Narrow Band NDVI Reflected Removed Lab none Removed from GT tower because most of the SRS NDVI sensors have been drifting in the past few years
LI-7550 AIU-1570 2024-02-28 West Pond Interface LI-7550 Removed Lab none Removed from WP as part of its decomissioning
WM 1590 120908 2024-02-28 West Pond Anemometer Removed Lab none Removed from WP as part of its decomissioning
eosFD FD20160001 2023-10-19 Lab none Shipped Factory none Shipped to Eosense. The data has been super noisy since mid-March.
eosFD FD20160001 2024-03-04 Factory none Returned Lab none Returned from Eosense after repair and calibration. CO2 Sensor Replacement 0.00, Recalibration 0.00,
Flux Calibration .00
SRS-Ni NDVI 1234903795 2024-02-27 Vaira Narrow Band Reflected Removed Lab none Removed from the refleced location at Vaira. These sensors drift randomly and can be replace with broad band NDVI from SW-PAR.
SRS-Ni NDVI 1334903221 2024-02-27 Vaira Narrow Band Incoming Removed Lab none Removed from the incoming location at Vaira. These sensors drift randomly and can be replace with broad band NDVI from SW-PAR.
HygroVUE10 E5761 2024-02-27 Lab none Installed Vaira RH & Temp Installed at Vaira in the RH/T location. Will cross compare with old HMP45 before removing the old sensor.
HygroVUE10 E5736 2024-02-27 Lab none Installed Tonzi RH & Temp Installed in the Floor RH/T location. Will cross compare with existing HMP45 before removing the old sensor.
GMP343 N2220008 2024-02-27 Lab none Installed Tonzi CO2 Probe Tower Profile 23 Reinstalled in the top location of the Tower co2 profile after testing in the lab. Reinstalled cable.
GMP343 N2210009 2024-02-27 Lab none Installed Tonzi CO2 Probe Tower Profile 6 Reinstalled in the upper-mid location of the Tower co2 profile after testing in the lab.
GMP343 N2210010 2024-02-27 Lab none Installed Tonzi CO2 Probe Tower Profile 1.8 Reinstalled in the lower-mid location of the Tower co2 profile after testing in the lab.
GMP343 N2220007 2024-02-27 Lab none Installed Tonzi CO2 Probe Tower Profile 0.35 Reinstalled in the bottom location of the Tower co2 profile after testing in the lab. New cable.
SONIC-ANEMO-SPE10 022365410 2024-02-14 Lab none Installed Bouldin Corn none Installed at Bouldin Corn with a new, udpated Mark 3, sn D006617.
Mark 3 D006617 2024-02-14 Lab none Installed Bouldin Corn none Installed at Bouldin Corn with a new 2-D sonic, sn 022365410, to replace the older Mark 3, sn B301069.
Mark 3 B301069 2024-02-14 Bouldin Corn Removed Lab none Removed from Bouldin Corn and replaced with the updated Mark 3 sn D006617. This older Mark 3 will be sent back to Loreli.
A547 3586 2024-02-12 West Pond Interface A547 Removed Lab none Removed from WP with conductivity sensor sn 6096 and brought back to the lab. Nothing wrong with the data (probably), I just wanted an extra sensor in the lab to play around with TempCoef.
LI-7700 TG1-0424 2024-02-26 Lab Calibration Lab Calibration Location: Lab
Last calibration: 2023-09-05
RSSI=77%, Optical RH=5%
Ambient P=100.6kPa, Tair=23.0C
Zero Air at 350psi, AMP CA04119 4.223 at 600
CH4 Zero Offset: -.063ppm
CH4 Span Offset: .009ppm
Notes: BWW sensor. After installing it again, I put the spiky hat and spray nozzle on again. The was reservoir was empty; Haley refilled it.
LI-7500A 75H-2150 2024-02-26 Lab Calibration Lab Last calibration: 2023-11-09
AGC=102.8 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.4kPa, Tair=24.1C, Tdew=10.96C
Zero Air at 350psi, AMP CA04119 575 at 524.77 Tdewspan=20.00
CO2 zero 0.9041 -> .9030 Offset: -5.59ppm
CO2 span 1.018 -> 1.0175 Offset: .48ppm
H2O zero 0.8731 -> .8737 Offset: 3.5ppm
H2O span 1.0265 -> 1.0382 Offset: -.25ppm
Notes: BWW tower sensor. One of the mount screws wouldn't come out initially; Joe got it out pulsing with vice grips + WD-40. He cleaned out the white dust inside the bolt holes with a thread cutter. He sprayed the bolts and bolt holes with a anti-corrosion spray before I installed it again. The 7550 temperature sensor was bad. I replaced with a repaired homemade temperature sensor. The HMP fan was not spinning, but it seemed to work fine after I removed a bit of fluff.
SONIC-ANEMO-SPE10 022365394 2024-02-14 Lab none Installed Bouldin Alfalfa Weather Station B Added to the new Arable Mark 3 sn D006812
Mark 3 D006812 2024-02-14 Lab none Installed Bouldin Alfalfa Weather Station B Installed at Bouldin Alfalfa to replace sn B301045
Mark 3 B301045 2024-02-14 Bouldin Alfalfa Weather Station B Removed Lab none Removed from Bouldin Alfalfa. Replaced by sn D006812.
LI-7700 TG1-0421 2024-02-16 Lab Calibration Lab Calibration Location: Lab
Last calibration: 2023-02-02
RSSI=81%, Optical RH=12%
Ambient P=100.98kPa, Tair=21.5C
Zero Air at 450psi, AMP CA04119 4.2238 at 675
CH4 Zero Offset: .032ppm
CH4 Span Offset: .0058ppm
Notes: Calibration before loaning this sensor to J.J. to use at Elkhorn Slough Yampah tower while they're waiting for replacement/repair
LI-7700 TG1-0421 2024-02-16 Lab none Loaned Loaned none Loaned sensor to Sylvain and J.J. (UCSC) to use at the Elkhorn Slough Yampah tower while they wait for a replacement sensor/repairs. Spiky hat was not loaned. Cables were loaned: 7700 yellow power cable, 7700 green Ethernet cable, wash black power cable, wash black tubing without spray nozzle (spray nozzle loaned previously for UCSC to use at Elkhorn Porter tower)
HMP155 T2930114 2024-02-16 Lab none Loaned Loaned none Loaned sensor and cable to Sylvain and J.J. (UCSC) to use at the Elkhorn Slough Yampah tower while they wait for a replacement sensor/repairs.
CS547A 6096 2024-02-12 West Pond Conductivity Sensor Removed Lab none Removed from WP with interface sn 3586 and brought back to the lab. Nothing wrong with the data (probably), I just wanted an extra sensor in the lab to play around with TempCoef.
LI-7500A 0418 2024-02-12 Lab none Installed Gilbert Tract Gas Analyzer CO2/H2O Installed at Gilbert Tract to replace sn 75H-2176 for calibration.
LI-7500A 75H-2176 2024-02-12 Gilbert Tract Gas Analyzer CO2/H2O Removed Lab none Removed from Gilbert Tract for calibration. Replaced by sn 0418.
LI-7500A 0418 2024-02-09 Lab Calibration Lab Last calibration: 2023-10-23
AGC=101.3 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.68kPa, Tair=23.2C, Tdew=5.6C
Zero Air at 475psi, AMP CA04119 524.77 at 700 Tdewspan=20.00
CO2 zero 0.9072 -> .9075 Offset: 1.41ppm
CO2 span 1.0066 -> 1.0063 Offset: .14ppm
H2O zero 0.9331 -> .9317 Offset: -7.5ppm
H2O span 1.0052 -> 1.002 Offset: .08ppm
Notes: Vertical orientation. H2O zero only took 10min to stabilize
CS547A 6097 2024-02-13 Disposed none Notes Disposed none Found this sensor in the dead sensor box in the lab! We were able to find the Kc (cell constant) on its label. This was not previously recorded in the equipment entry, and Campbell did not have any record of it either. Incontrovertible proof that we should never throw anything away.
GMP343 N2210009 2024-02-09 Tonzi CO2 Probe Tower Profile 6 Removed Lab none Removed from the upper mid location on the Tower due to bad readings and to be calibrated
GMP343 N2210010 2024-02-09 Tonzi CO2 Probe Tower Profile 1.8 Removed Lab none Removed from the lower mid location on the Tower due to bad readings and to be calibrated
GMP343 N2220007 2024-02-09 Tonzi CO2 Probe Tower Profile 0.35 Removed Lab none Removed from the lowest location on the Tower due to bad readings, cow chewed cable and to be calibrated
CS655 17131 2024-01-30 Lab none Installed Mayberry none Installed at MB in a tule patch southeast of the tower (~50cm from the staff gauge). This is a test to see if we can continuously track porewater salinity changes.
CS655 17130 2024-01-30 Lab none Installed Mayberry none Installed at MB in the channel in front of the tower. This is a test to see if we can continuously track porewater salinity changes.
HMP155 T2820180 2024-01-25 Factory none Disposed Disposed none Campbell's evaluation:
Inspected. Cap filter is dirty and sticker labels started to peel off suggesting sensor has been deployed in a very humid environment. After tests sensor showed problems to provide correct RH and Temp data. Unfortunately sensor can't be repaired or calibrated. Warranty expired o 08-27-2022. NOTE: We suggest to acquire a new unit such as the HygroVUE10, which offers advantages worth look up, as a replacement for the damaged one.

Dead unit will be returned so that I can take it apart!
HMP155 T2820180 2024-01-17 Lab none Shipped Factory none Sent to Campbell on RMA 37957. RH and AirT are reading zero.
EXO2 22C103416 2024-01-23 Lab none Tested Lab none The wiper sn 22E103340 hasn't worked properly since being installed in mid-December. When we pulled the sonde last week, the brush was parked over the pH sensor and all the sensors were dirty. I brought the sonde back to the lab for routine calibration and tested the brush. The brush did work smoothly as first, but then it became intermittent. Sometimes it didn't work at all and sometimes it worked only in spurts (and the EXO would disconnect from the laptop right before the brush moved). The wiper did not show up under the Sensors or Calibration tab. Nothing changed when I tried plugging the wiper into a different port. I talked to Matt at YSI who thinks something is messed up electronically inside the wiper--it can receive commands (to wipe) but is not talking back to the sonde. I will send the wiper back to YSI for repair.
EXO2 22C103416 2022-09-29 Lab none Calibration Lab none Depth (m), Last cal: 9/23/2022 11:38:38 AM
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,0.109 m,0.000 m,0.000 m,22.000 °C
EXO2 22C103416 2024-01-22 Lab none Calibration Lab none Depth(m),Last cal: 2023-10-18
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,-0.076 m,0.000 m, 0.000 m, 17.888 °C
EXO pH-ORP 22D102641 2024-01-22 Lab none Calibration Lab none pH, Last cal: 2023-10-16
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.01 pH,7.01 pH,7.01 pH,-4.8 mV,19.428 °C
Pt2,10.06 pH,10.04 pH,10.06 pH,-174.7 mV,20.116 °C
Pt3,4.00 pH,4.05 pH,4.01 pH,161.2 mV,21.033 °C
4-7Slope,4-7mVPerDecade,7-10Slope,7-10mVPerDecade
166.0 mV,55.16,169.9 mV,55.71
Note: Calibration pretty close and delta slope and mV per decade are ok. Used opened standards from May 2023.
EXO pH-ORP 22D102641 2024-01-22 Lab none Calibration Lab none ORP (mV), Last Cal: 2023-10-16
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,230.4 mV,227.6 mV,223.3 mV,0.0 mV,21.275 °C
Note: Poured Zobell solution (mixed 2022-09) into small brown plastic bottle for this calibration.
EXO Wiped Cond/Temp 23D104030 2024-01-22 Lab none Calibration Lab none Sp Cond (uS/cm),Last cal: 2023-08-17
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.5
CellConstant 0.46
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,8420.0 µS/cm,8507.6 µS/cm,8420.0 µS/cm, 0.0 µS/cm,20.226 °C
Note: Used 10,000 uS/cm standard from fridge (brought it to room temperature first). Measured temp-compensated conductivity with handheld YSI probe
EXO Turbidity 22E106072 2024-01-22 Lab none Calibration Lab none Turbidity (FNU), Last cal: 2023-10-16
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.4
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 FNU,3.85 FNU,0.00 FNU,0.00 FNU,21.571 °C
Pt2,124.00 FNU,122.47 FNU,124.00 FNU,0.00 FNU,20.745 °C
Pt3,1010.00 FNU, 959.99 FNU, 1010.00 FNU, 0.00 FNU, 20.651 °C
Note: Used the lab's DI water for zero point for this and previous calibrations. I'll try to get some miliQ water for future calibrations since this zero point had a larger than usual offset. Probably doesn't affect our data too much since our turbidity is usually >50 FNU.
EXO fDOM 22E100649 2024-01-23 Lab none Calibration Lab none fDOM (QSU and RFU), Last cal: 2023-08-17
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-2.68 QSU,0.00 QSU,0.00 QSU,20.692 °C
Pt2,300.00 QSU,182.33 QSU,300.08 QSU,0.00 QSU,21,776 °C
Notes: Fresh calibration solution after a factory reset. Previous calibration attempts in the last few days were all really off. Keep an eye on this sensor.

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,0.24 RFU,0.00 RFU,0.00 RFU,20.081 °C
Pt2,100.0 RFU,RFU,122.16 RFU,100.05 RFU,0.00 RFU,22.019 °C
Notes: Fresh calibration solution after a factory reset
EXO pH-ORP 22D102641 2024-01-18 Hill Slough pH-ORP Removed Lab none Removed from HS channel and brought back to the lab for routine calibration
EXO Wiped Cond/Temp 23D104030 2024-01-18 Hill Slough Conductivity Sensor Removed Lab none Removed from HS channel and brought back to the lab for routine calibration
EXO fDOM 22E100649 2024-01-18 Hill Slough fDOM Removed Lab none Removed from HS channel and brought back to the lab for routine calibration
EXO Turbidity 22E106072 2024-01-18 Hill Slough Turbidity Removed Lab none Removed from HS channel and brought back to the lab for routine calibration
EXO2 22C103416 2024-01-18 Hill Slough Water Level Removed Lab none Removed from HS channel and brought back to the lab for routine calibration
LI-7500A 75H-2418 2024-01-23 Lab none Returned Returned none Returned to Stephen Chan at LBL. Returned with AMP LI-7550 sn AIU-1039, a 7550 thermistor, IRGA cable, and IRGA mount.
LI-7550 AIU-1039 2024-01-23 Lab none Returned Returned none Returned to Stephen Chan at LBL now that our 7550 AIU-1866 is back from the factory. AMP 7550 Returned with AMP LI-7500A sn 75H-2418, a 7550 thermistor, IRGA cable, and IRGA mount.
LI-7500A 1594 2024-01-18 Lab none Installed Hill Slough Gas Analyzer CO2/H2O Installed at Hill Slough to replace sn 0418 for calibration.