Equipment and Calibration History

View Equipment |

View history entry 6123 for Licor LI-7500A sn: 1594

Date: 2022-11-28
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2022-04-08 AGC=98.3 SDM= DAC1= DAC2= Ambient P=100.3kPa, Tair=20.9C, Tdew=4.05C Zero Air at 1800psi, AMP CA04119 524.77 at 1350 Tdewspan=16.00 CO2 zero 0.8643 -> .8643 Offset: -.38ppm CO2 span 1.0097 -> 1.0168 Offset: -4.93ppm H2O zero 0.8891 -> .8889 Offset: -1.1ppm H2O span 1.0106 -> 1.0136 Offset: -.06ppm Notes: Set H2O span on 2022-11-29 using 16.00C. Ambient T=22.4 and Dew Point = 0.77C.
Entered by: