Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 [55] 56 57 58 59 60 61 62 63 64 65 66 67 68
Ser No.DateLoc FromTitleLoc ToNotes
LI-7700 TG1-0155 2014-09-10 East End Gas Analyzer CH4 Calibration East End Gas Analyzer CH4 Calibration Location: East End
Last calibration: 2014-03-07
RSSI=80%, Optical RH=23%
Ambient P=101.23kPa, Tair=29.9C
Zero Air at 1400psi, 2nd CH4 4.995 at 700
CH4 Zero Offset: 0.085ppm
CH4 Span Offset: -0.195ppm
Notes: Very hot day with no wind. Offsets larger than usual.
LI-7700 TG1-0223 2014-09-10 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2014-03-07
RSSI=82%, Optical RH=12%
Ambient P=100.88kPa, Tair=31.5C
Zero Air at 1400psi, 2nd CH4 4.995 at 750
CH4 Zero Offset: 0.025ppm
CH4 Span Offset: -0.085ppm
Notes: Very hot day with no wind. Span offset a little large.
LI-7700 TG1-0211 2014-09-10 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell
Last calibration: 2014-03-07
RSSI=69%, Optical RH=22%
Ambient P=kPa, Tair=25.9C
Zero Air at 1400psi, 2nd CH4 4.995 at 800
CH4 Zero Offset: 0.07ppm
CH4 Span Offset: -0.075ppm
Notes: Very warm day with no wind. Offsets seem a little large.
LI-7500A 75H-2180 2014-09-08 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed on the West Pond tower with 7550 sn AIU-0666. There was no 7550-7500A pair on the tower due to issues when we tried to swap the head on 9/4.
LI-7550 AIU-0666 2014-09-08 Lab none Installed West Pond Interface LI-7550 Installed on the West Pond tower with head sn 75H-2180. There was now 7550-7500A pair on the tower because of issues with swapping the head on 9/4. This box has a new door gasket.
LI-7700 TG1-0155 2014-02-25 Twitchell Alfalfa Gas Analyzer CH4 Installed East End Gas Analyzer CH4 Installed on the East End Tower with its wash box to replace sn TG1-0203 that was returned to Frank. This instrument came directly from Alfalfa.
LI-7700 TG1-0155 2014-02-25 Twitchell Alfalfa Gas Analyzer CH4 Removed East End Gas Analyzer CH4 Removed from the Alfalfa tower along with its wash box to be installed on the East End Tower to replace sn TG1-0203 that was returned to Frank
LI-7700 TG1-0223 2012-05-02 Sherman Island Gas Analyzer CH4 Installed West Pond Gas Analyzer CH4 Installed at West Pond on the folding boom at the top of the tower.
LI-7700 TG1-0223 2012-05-02 Sherman Island Gas Analyzer CH4 Removed West Pond Gas Analyzer CH4 Removed from the Sherman tower for use at West pond. Replaced by LGR (sn 08-114).
LI-7700 TG1-0357 2014-06-13 Lab none Installed Mayberry-Portable Installed on the Mayberry Portable tower to replace sn TG1-224 that was having line lock problems.
LI-7550 AIU-0614 2014-09-04 West Pond Interface LI-7550 Removed Lab none Removed from the West Pond tower. In trying to swap the head, this box went into a reboot loop and we could not get it to complete a boot. This box also needs its connector circuit board upgraded to prevent condensation problems.
LI-7500A 0065 2014-09-04 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed for calibration. Problems with the 7550 box sn AIU-0614 prevented installation of a new 7500.
LI-7500A 75H-2180 2014-09-02 Lab none Calibration Lab none Last calibration: 2014-03-26
AGC=94.8 SDM=1 DAC1=- DAC2=-
Ambient P=99.73kPa, Tair=23.3C, Tdew=14.27C
Zero Air at 1000psi, God CO2 421.22 at 1350 Tdewspan=15.25
CO2 zero 0.9147 -> 0.9143 Offset: -1.66ppm
CO2 span 1.0173 -> 1.0134 Offset: 1.98ppm
H2O zero 0.8803 -> 0.8812 Offset: 5.3ppm
H2O span 0.9946 -> 1.0193 Offset: -0.5ppm
Notes:
LI-7500A 75H-2182 2014-08-25 Lab none Calibration Lab none Last calibration: 2014-02-28
AGC=83.4 SDM=1 DAC1=- DAC2=-
Ambient P=99.77kPa, Tair=23.79C, Tdew=12.26C
Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.25
CO2 zero 0.8815 -> 0.8816 Offset: 0.90ppm
CO2 span 1.0064 -> 1.0074 Offset: -0.54ppm
H2O zero 0.8476 -> 0.8463 Offset: -7.7ppm
H2O span 0.9875 -> 1.0039 Offset: -0.33ppm
Notes:
CR23X 2071 2011-01-24 Borrowed Borrowed Lab none Borrowed from Robin Weber in Allen Goldstein's lab.
CR23X 2071 2014-08-29 Lab none Upgrade Lab none The operating system was updated from Rev9 to Rev19.
GMP221 K1340006 2014-08-27 Lab none Installed Twitchell Alfalfa Installed in profile B at 8cm to replace sn F4960018 due to zero reading.
GMT221 F4960018 2014-08-27 Twitchell Alfalfa Removed Lab none Removed from profile B at 8cm due to reading zero. Replaced by sn K1340006.
GMP221 Y2009100 2014-08-27 Lab none Installed Twitchell Alfalfa Installed in profile B at 2cm to replace sn F4960019 due to zero reading.
GMT221 F4960019 2014-08-27 Twitchell Alfalfa Removed Lab none Removed from profile B at 2cm due to reading zero. Replaced by sn Y2009100.
GMP222 Z2720014 2014-08-27 Lab none Installed Twitchell Alfalfa Installed in Forced Diffusion probe 1 to replace sn Z2720011 due to zero reading
GMP222 Z2720011 2014-08-27 Twitchell Alfalfa Removed Lab none Removed from Forced Diffusion probe 1 because it was reading zero. Replaced by sn Z2720014
LI-7500A 75H-2182 2014-08-27 Lab none Installed Twitchell Alfalfa Gas Analyzer CO2/H2O Installed on the Twitchell Alfalfa tower to replace sn 75H-2180 for calibration.
LI-7500A 75H-2180 2014-08-27 Twitchell Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell Alfalfa tower for calibration. Replaced by sn 75H-2182.
CR10X 35475 2014-07-21 Borrowed Returned Lab none Returned to the lab by Leilei who had borrowed it with the Picarro / Forerunner soil respiration experiment for use in Marin.
CR10X 35475 2014-06-26 Lab none Installed Sherman Island Installed with the Picarro / Forerunner soil raspiration campaign to control ventilation fans in the big box enclosure.
CR10X 35475 2014-06-13 Mayberry-Portable Removed Lab none Removed from the portable tower and replace with sn that was programed and wired for a TC profile and conductivity sensor.
LI-7500A 0042 2014-06-26 Lab none Installed Sherman Island Gas Analyzer CO2/H2O Installed on the Sherman tower to replace sn 0976 for calibration.
LI-7500 0976 2014-06-26 Sherman Island Gas Analyzer CO2/H2O Removed Lab none Removed from the Sherman tower for calibration. Replaced by sn 0042.
LI-7550 AIU-1285 2014-08-19 Lab none Installed Mayberry Interface LI-7550 Installed on the Mayberry tower with 7700 sn TG1-0224 and 7500 sn 75H-2514 to replace the Ameriflux emergency response GHG system.
LI-7700 TG1-0224 2014-08-19 Lab none Installed Mayberry Gas Analyzer CH4 Installed on the Mayberry tower with 7500 sn 75H-2514 and 7550 sn AIU-1285to replace the Ameriflux emergency response GHG system.
LI-7550 AIU-0915 2014-08-19 Mayberry Interface LI-7550 Removed Returned none This is part of the emergency response GHG system borrowed from Ameriflux. Removed from the Mayberry Tower with 7700 sn TG1-0352 and 7500 sn 75H-2339 and returned to Stephen at LBNL Ameriflux.
LI-7700 TG1-0352 2014-08-19 Mayberry Gas Analyzer CH4 Removed Returned Anemometer This is part of the emergency response GHG system borrowed from Ameriflux. Removed from the Mayberry Tower with 7500 sn 75H-2339, 7550 sn AIU-0915 and returned to Stephen at LBNL Ameriflux.
LI-7500A 75H-2514 2014-08-19 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry tower with 7700 sn TG1-0224 and 7550 sn AIU-1285 to replace the Ameriflux emergency response GHG system.
LI-7500A 75H-2339 2014-08-19 Mayberry Gas Analyzer CO2/H2O Removed Returned none This is part of the emergency response GHG system borrowed from Ameriflux. Removed from the Mayberry Tower with 7700 sn TG1-0352 and 7550 sn AIU-0915 and returned to Stephen at LBNL Ameriflux.
LI-7500A 0073 2014-06-25 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 0035 for calibration.
LI-7500A 0035 2014-06-25 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira tower for calibration. Replaced by sn 0073.
LI-7550 AIU-0397 2014-04-28 Mayberry Gas Analyzer CO2/H2O Installed Mayberry-Portable Gas Analyzer CO2/H2O Installed on the Mayberry Portable tower with LI-7500 sn 75H-2176 and LI-7700 sn TG1-0224. This tower is mounted in the back of Dennis's truck and will be moved between two locations on a roughly weekly basis.
LI-7700 TG1-0224 2014-08-18 Factory Returned Lab none Returned from Licor. Sporadic line lock problems were finally traced to the laser. Licor was unable to determine why the laser failed and replace the laser for free. Licor calibrated the sensor.
LI-7700 TG1-0224 2014-07-01 Lab none Shipped Factory This sensor has been having sporadic line lock problems both in the field and in the lab. Shipped to Licor for diagnosis and repair.
LI-7700 TG1-0224 2014-06-13 Mayberry-Portable Removed Lab none Removed from the portable tower due to intermittent line lock problems. Brought to the lab for further testing. Replaced by sn TG1-0357.
LI-7700 TG1-0224 2014-04-28 Mayberry Gas Analyzer CH4 Installed Mayberry-Portable Gas Analyzer CH4 Installed on the Mayberry Portable tower with LI-7500 sn 75H-2176 and LI-7550 sn AIU-0397. This tower is mounted in the back of Dennis's truck and will be moved between two locations on a roughly weekly basis.
LI-7500A 75H-2176 2014-04-28 Mayberry Gas Analyzer CO2/H2O Installed Mayberry-Portable Gas Analyzer CO2/H2O Installed on the Mayberry Portable tower with LI-7700 sn TG1-0224 and LI-7550 sn AIU-0397. This tower is mounted in the back of Dennis's truck and will be moved between two locations on a roughly weekly basis.
LI-7550 AIU-0915 2014-04-23 Borrowed Borrowed Lab none Borrowed from the Ameriflux project at LBNL as part of their emergency response GHG system. Came with LI-7500A sn 75H-2339 and LI-7700 sn TG1-0352.
LI-7700 TG1-0352 2014-04-23 Borrowed Borrowed Lab none Borrowed from the Ameriflux project at LBNL as part of their emergency response GHG system. Came with LI-7500A sn 75H-2339 and LI-7550 sn AIU-0915. LI-7700 sn TG1-0352
LI-7500A 75H-2339 2014-04-23 Borrowed Borrowed Lab none Borrowed from the Ameriflux project at LBNL as part of their emergency response GHG system. Came with LI-7700 sn TG1-0352 and LI-7550 sn AIU-0915.
LI-7500A 75H-2176 2014-04-28 Mayberry Gas Analyzer CO2/H2O Removed Mayberry-Portable Gas Analyzer CO2/H2O Removed from the permanent Mayberry tower and installed on the portable tower. Replaced by Ameriflux sn 75H-2339
LI-7500A 75H-2150 2014-06-26 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed on the Twitchell tower to replace sn 75H-2182 for calibration.
LI-7500A 75H-2182 2014-06-26 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell tower for calibration. Replaced by sn 75H-2150.
WM 1590 120906 2012-05-07 Lab none Installed Twitchell Rice Anemometer Installed at the rice to replace sn 000024
FMA RMT-200 05-0010 2012-01-25 Twitchell Rice Calibration Twitchell Rice With the new external filter I calibrated the LGR. The cal tank has <200psi of 3.276ppm CH4. Last calibration was Dec 20, 2011.
Pre Cal: 8:49PST, 29.61C, 156.07Torr, 13.871usec, 3.282ppm, 1Hz

The first try at calibration failed somehow ending in a value (3.317ppm) farther from the span gas. A second calibration seemed to work fine and ended up right on.

When I left the LGR read
9:56PST, 23.93C, 153.81Torr, 13.060usec, 10Hz, 2.311ppm
FMA RMT-200 05-0010 2011-12-18 Twitchell Rice Calibration Twitchell Rice All was running. The LGR read:
9:04PST, 31.36C, 97.41Torr, 13.973usec, 10Hz, 2.964ppm

I calibrated the LGR with secondary tank with 3.276ppm CH4 at 250psi. Last cal was 2011-11-22
Pre cal: 9:12PST, 30.18C, 124.54Torr, 13.972usec, 1Hz, 3.200ppm
Pressure was a little low but okay. After cal the LGR read 3.271ppm. I adjusted the bypass valve to bring the pressure to 146Torr in high flow mode. I returned the LGR to 10Hz before leaving.
FMA RMT-200 05-0010 2011-11-22 Twitchell Rice Calibration Twitchell Rice We calibrated the LGR with a 3.276ppm tank with 400psi. Last cal was Oct 10, 2011.
9:56PST, 9.53C, 156.07Torr, 14.341usec, 1Hz, 3.357ppm
After cal it was right on. When we left it was back in high flow at 10Hz and pressure was about 141Torr, I did not adjust the bypass valve.
FMA RMT-200 05-0010 2011-10-27 Lab none Installed Twitchell Rice Installed to replace sn 08-0114 since in slow flow mode there was zero flow.
FMA RMT-200 08-0114 2011-08-12 Twitchell Rice Calibration Twitchell Rice The LGR pressure was really low 89.28 and in slow flow it was even lower. I checked the external 2um filter and it seemed clean. I checked the internal 2um filter and it was completely stopped up. I replaced it with a cleaned one. I blew the replacement filter clear with the cal tank but it still brought the ring time down after I got it all back together. The pressure was better and I had to close down the bypass valve to bring it in range.
9:00PST, 24.16C, 144.56Torr, 13.300usec, 10Hz, 1.831ppm

The low flow pressure was still a bit low during calibration:
9:06PST, 24.05C, 125.31Torr, 13.302usec, 1Hz, 3.270ppm

The calibration was almost right on with the 3.267ppm tank at 650psi. Last cal was July 1, 2011. The pressure fell to 111.7 by the end of the calibration.

I also replace the 40um filter strainer. It was very dirty and had some spider parts in it.

Before leaving the LGR read:
9:34PST, 24.55C, 144.81Torr, 13.291usec, 10Hz, 1.827ppm
FMA RMT-200 05-0010 2010-05-13 Twitchell Rice Calibration Twitchell Rice The LGR was reporting:
8:27PDT, 22.89C, 120.58Torr, 16.781usec, 10Hz, 1.858ppm

I adjusted the bypass valve to correct the pressure. I calibrated the LGR with 3.276ppm tank with 700psi. Last cal was March 22, 2011
Precal: 8:47PDT. 24.11C, 156.50Torr, 16.776usec, 1Hz, 3.290ppm
After cal 3.276ppm
Before I left the LGR read:
8:52PDT, 24.49C, 150.77Torr, 16.774usec, 10Hz, 1.883ppm
FMA RMT-200 05-0010 2011-03-22 Twitchell Rice Calibration Twitchell Rice Pre cal: 8:42PST, 17.16C, 156.57Torr, 17.042usec, 1Hz, 3.347ppm � about 70ppb off.
Post cal: 3.277ppm

Before we left the LGR read:
8:55PST, 17.42C, 149.22Torr, 17.068usec, 10Hz, 2.176ppm
FMA RMT-200 08-0114 2010-12-02 Twitchell Rice Calibration Twitchell Rice Pre cal:
7:43PST, 24.90C, 142.80Torr, 16.064usec, 1Hz, 2.660ppm - off by 616ppb a lot.
Post cal CH4 was 3.263ppm

And back in operation after adjusting the bypass valve because pressure was too high:
7:53PST, 23.33C, 143.96Torr, 16.089usec, 10Hz, 3.266ppm
FMA RMT-200 08-0114 2010-09-08 Twitchell Rice Calibration Twitchell Rice I adjusted the bypass valve. When Frank arrived we did a calibration. Last cal was done on Aug 6, 2010. The field tank is 3.276ppm CH4
Precal: 9:52PST, 25.54C, 111.2Torr, 16.112usec, 1Hz, 3.377ppm

Low flow pressure seems low here too but I did not check the filters. Being 100ppb off seems a lot, last cal uses a standard closer to 1.8ppm though. Post cal CH4 read 3.276ppm.

When we left the LGR read:
10:02PST, 25.49C, 144.34Torr, 16.080usec, 10Hz, 1.840ppm
FGGA 09-0010 2010-06-22 Twitchell Rice Calibration Twitchell Rice Initial readings from the LGR were:
9:57, 29.18C, 129.50Torr, 13.902usec, 10Hz, 1.865ppm, 14819ppm, 385.91ppm

I tried calibrating and notice very low pressure in low flow mode � 74.89Torr. I completed the calibration, but did not like it. I checked the exterior 2um filter and it seemed okay. I also checked the internal 2um filter and it seemed good as well. In checking the interior I noticed a loose connection. The connector that has to be disconnected to clean the mirror was left disconnected last time I cleaned the mirrors. I�m asking LGR what this connection does. It doesn�t look significant but after connecting it the pressure improved somewhat. And I did a second calibration.

Precal: last cal June 3, 3.276ppm CH4, 433.0ppm CO2, 1800psi
10:54, 134.32Torr, 3.217+-0.003ppm CH4, 416+-2ppm CO2

Post cal: 10:57, 130.0Torr, 3.276+-0.002ppm CH4, 433+-1ppm CO2

Last readings from LGR:
11:02, 28.57C, 144.47Torr, 13.86usec, 10Hz, 1.888ppm, 15195ppm, 400.44ppm
FGGA 09-0010 2010-06-03 Twitchell Rice Calibration Twitchell Rice Calibrated with 3.276ppm CH4 and 433.0ppm CO2 tank 1750psi. Last cal May 20, 2010
Precal: 10:26, 129.6Torr dropping to 119Torr, 3.30+-0.05ppm CH4, 430+-1ppm CO2
Postcal: 3.26+-0.001ppm CH4, 429+-1ppm CO2

System back at 10Hz.
FGGA 09-0010 2010-05-20 Twitchell Rice Calibration Twitchell Rice calibrated the LGR:

Initial readings:
9:35, 19.99C, 138.24Torr, 14.071usec, 10Hz, 1.8530ppm, 8187ppm, 406.97ppm

bypass valve adjusted to 143.55Torr. Calibrated with 3.276ppm CH4, 433.0ppm CO2 tank with 1800psi

Pre cal: 9:41, 138.16Torr, 1Hz, 3.220+-0.005ppm, 444+-1ppm
Post cal: 9:48, 3.278+-0.002ppm, 432+-1ppm

Ambient conditions after calibration:
9:50, 20.02C, 143.90Torr, 14.069usec, 10Hz, 1.8610ppm, 8332ppm, 390.55ppm
FGGA 09-0010 2010-05-05 Twitchell Rice Calibration Twitchell Rice Twitchell: cal gas 3.276 ppm CH4, 433ppm CO2, mirror 13.9 us, 145 Torr, 37C
CH4 pre-cal: 3.342
CO2 pre-cal: 441
CH4 post-cal: 3.273
CO2 post-cal: 433
FGGA 09-0010 2010-04-22 Twitchell Rice Calibration Twitchell Rice The LGR would not line lock CH4/H2O when power was restored. Initially the cell temperature was about 10C and I let it warm up to 18C with no lock and many restarts. Finally I called LGR and got their maintenance password to change line lock settings from 0.165 to 0.185 and this did the trick.

Initial readings
11:35, 18.50C, 144.60Torr, 13.940usec, 10Hz, 2.149ppm, 12039ppm, 398.07ppm

Calibrated with 3.276ppm CH4, 433.0ppm CO2 tank, last cal Mar 11, 2010
Precal 11:40, 138.24Torr, 3.270+-0.002ppm CH4, 425.0+-1ppm CO2
Postcal: 11:44, 138Torr, 3.279+-0.002ppm CH4, 433.5+-1ppm CO2
FGGA 09-0012 2010-02-18 Twitchell Rice Calibration Twitchell Rice Last calibration: 2010-01-14 00:00:00
Cell Pressure=39.54kPa, Cell Temperature=27.98C, Ring Time=17.17usec
2nd standard at 0psi: CH4: 3.276ppm, CO2: 433ppm, H2O: 0ppm
CH4 = 3.5032ppm -> 3.316ppm Offset: -0.227ppb
CO2 = 448.8ppm -> 441ppm Offset: -15.8ppm
Notes: External filter seemed plugged up, swapped it with cleaned filter before calibration. Values improved, but I still don't agree with cal tank. I think the very low pressure is a problem. Maybe the internal filter is plugged up. I will do some check in the lab and then possibly swap out the LGR.
FGGA 09-0012 2009-08-06 Lab none Installed Twitchell Rice Installed Frank's FGGA at the rice
FGGA 09-0012 2010-01-14 Twitchell Rice Calibration Twitchell Rice Last calibration: 2009-03-03 00:00:00
Cell Pressure=148.63kPa, Cell Temperature=28.27C, Ring Time=17.277usec
2nd Standard at 0psi: CH4: 3.276ppm, CO2: 433ppm, H2O: 0ppm
CH4 = 3.368ppm -> 3.275ppm Offset: -0.092ppb
CO2 = 432.1ppm -> 433.4ppm Offset: 0.9ppm
Notes: LGR pressure very low ~16Torr, but very easily adjusted using bypass valve.
LI-7500A 0041 2009-07-31 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed at Rice. Can't figure out which sn it replaced though....
FMA RMT-200 08-0114 2009-08-06 Twitchell Rice Moved Sherman Island Gas Analyzer CH4 Closed-path Removed from Rice and replaced with Frank's FGGA analyzer (sn 09-0012). Installed to replace LGR sn 05-0010 for maintenance.
FMA RMT-200 08-0114 2009-06-04 Lab none Installed Twitchell Rice Installed at the Rice.
WMP 1352 000024 2009-04-03 Lab none Installed Twitchell Rice Anemometer Installed at the Rice. Z= 3.05 m at middle of path & North spar pointing to 318 magnetic.

Need to apply Schotanus correction since model 1352.
USB2000-VIS-NIR USB2G4461 2014-07-08 Factory Returned Lab none Returned from Ocean Optics. They fixed the coax connector where the fiber optics connects to the box. They did a "level two repair" which seems to include a calibration (document on file).
USB2000-VIS-NIR USB2G4461 2014-06-13 Lab none Shipped Factory Sent to Ocean Optics for repair on RMA 1415292. The coax connector where the fiber optics connects to the box broke off.
LI-6400 PSC-0785 2014-06-25 Lab none Calibration Lab none Madeleine and I calibrated the 6400 in the lab
Zero: CO2 1.45 to 0.0
H2O -0.21 to -0.001

Span CO2 with 421.22 CO2 gas
--did not need to span--left at 421.0

H2O span (dew point set at 15.26C)
14.63 to 15.26C
LI-7500A 0073 2014-06-24 Lab none Calibration Lab none Last calibration: 2014-02-28
AGC=60 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.46kPa, Tair=22.08C, Tdew=12.44C
Zero Air at 1200psi, God CO2 421.22 at 1500 Tdewspan=15.26
CO2 zero 0.9578 -> 0.9580 Offset: 0.65ppm
CO2 span 1.0009 -> 1.0018 Offset: -0.72ppm
H2O zero 0.7355 -> 0.7389 Offset: 21.0ppm
H2O span 0.9972 -> 1.0138 Offset: -0.28ppm
Notes:
LI-7500A 0042 2014-06-24 Lab none Calibration Lab none Last calibration: 2014-02-28
AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.62kPa, Tair=21.30C, Tdew=11.17C
Zero Air at 1200psi, God CO2 421.22 at 1500 Tdewspan=15.26
CO2 zero 0.9506 -> 0.9506 Offset: -0.12ppm
CO2 span 1.0078 -> 1.0074 Offset: 0.18ppm
H2O zero 0.8506 -> 0.8506 Offset: 0.3ppm
H2O span 0.9853 -> 0.9960 Offset: -0.19ppm
Notes:
LI-7500A 75H-2150 2014-06-24 Lab none Calibration Lab none Last calibration: 2013-12-18
AGC=95.08 SDM=1 DAC1=- DAC2=-
Ambient P=100.15kPa, Tair=21.78C, Tdew=11.81C
Zero Air at 1200psi, God CO2 421.22 at 1500 Tdewspan=15.26
CO2 zero 0.9012 -> 0.9012 Offset: 0.07ppm
CO2 span 1.0016 -> 1.0011 Offset: 0.28ppm
H2O zero 0.8643 -> 0.8652 Offset: 5.8ppm
H2O span 0.9987 -> 1.0087 Offset: -0.21ppm
Notes:
GMP221 Y2009100 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*1.233 + -902.3; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
GMP221 Y0850004 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*0.743 + -1252; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
GMP221 Y1720006 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*0.6676 + -1028; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
GMP222 Z2720011 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*1.176 + -287.8; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
GMP221 Y0710013 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*1.028 + -18.33; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
GMT221 G0150037 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*1.017 + -200.7; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
GMP222 H1630066 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*1.746 + -27.77; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
GMP222 Z2720014 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*1.248 + 0.9176; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
GMP222 Z2720014 2014-04-08 Lab none Calibration Lab none Trueppm = Probeppm*1.027 + 0.2028; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.1C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP221 Y2009100 2014-04-08 Lab none Calibration Lab none Trueppm = Probeppm*1.047 + -347.2; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.1C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP221 K1340006 2014-04-08 Lab none Calibration Lab none Trueppm = Probeppm*1.002 + -37.28; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.1C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP221 K1340007 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.01 + -11.96; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP221 K1340005 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*0.9908 + -5.527; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP222 K1350103 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.022 + 63.48; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP222 K1350102 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.02 + 41.22; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP222 K1310104 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.008 + 43.09; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP222 K1350101 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.014 + 51.82; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP222 K1310103 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.094 + 0.7501; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP222 K1310102 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.014 + 42.61; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
GMP222 Y1950005 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*1.073 + -174.1; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.
GMP222 Z2720015 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*0.9937 + 44.53; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.
GMT222 G0230062 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*1.004 + 19.27; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.
GMP221 Y1720006 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*0.9255 + -857.4; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.