View history entry 1685 for Licor LI-7550 sn: AIU-0397

Date: 2015-01-30
From: Factory -
To: Lab - none
Title: Returned
Notes: Returned from Licor on RMA 00186965. Licor says: Inspected the unit and found the LI-7550 would not power up. Found multiple power supplies shorted due to ants getting inside the LI-7550. Removed ants, replaced the defective analog board and subjected the unit to environmental tests.
Entered by: Joe Verfaillie