View history entry 1762 for Licor LI-7700 sn: TG1-0211

Date: 2015-03-24
From: Twitchell Rice - Gas Analyzer CH4
To: Twitchell Rice - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Twitchell Last calibration: 2014-09-10 RSSI=56%, Optical RH=25% Ambient P=102.5kPa, Tair=18.8C Zero air at 1200psi, 2nd CH4 4.995 at 600 CH4 Zero Offset: -0.05ppm CH4 Span Offset: 0.065ppm Notes: Very windy
Entered by: