Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 [25] 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68
Ser No.DateLoc FromTitleLoc ToNotes
S2-411 1053 2020-07-15 Lab none Initial settings Lab none Set up as SDI-12 address G for NDVIin
S2-411 1049 2020-07-15 Lab none Initial settings Lab none Set up as SDI-12 address G for NDVIin
RPI0W RPI0W-005 2020-07-15 Lab none Tested Lab none This RTC clock does not work, keep resetting the clock to 2017-12-10, which is strange because usually the clock resets back to 12-31-1999 (especially if the battery is dead). I'm borrowing the RTC clock from RPIW-003, which is still in the lab. We should order a new RTC to replace it.

I set up the access network as RPI0W-005.
AM16/32B 27703 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
PS-30M 0127000A 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
KD140SX-UFBS 11ZV0M0862 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
KD140SX-UFBS 11ZV0M0861 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
KD140SX-UFBS 11ZV0M0860 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
HMP45AC T2130005 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
TE525MM 49187-109 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
ML2x 377/010 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
ML2x 54/016 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
ML2x 377/008 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
NR01 1570 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
SRS-Ni NDVI 882103061 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
SRS-Ni NDVI 882103036 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
SRS-Pi PRI 913003007 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
SRS-Pi PRI 913003006 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
PQS 1 110338 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
PQS 1 110339 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
HFP01 6498 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
HFP01 6491 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
HFP01 6342 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
CS451 20010046 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
WM 1590 153802 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
LI-7700 TG1-0357 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
LI-7500A 75H-2668 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
LI-7550 AIU-1557 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
AM25T 6767 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
CR1000 47108 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
AM16/32B 11662 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
NL115 6763 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
SD130BN 0030F4D2148D 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
AM16/32B 27703 2020-07-09 Lab none Loaned Loaned none Given to Sabina of Hydrofocus with the Sherman Barn system. This was an extra mux to replace the old, failing mux sn 11662.
PS-30M 0127000A 2020-07-09 Lab none Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
KD140SX-UFBS 11ZV0M0862 2020-07-09 Lab none Loaned Loaned none Given to Sabina of Hydrofocus with the Sherman Barn system
KD140SX-UFBS 11ZV0M0861 2020-07-09 Lab none Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
KD140SX-UFBS 11ZV0M0860 2020-07-09 Lab none Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
HMP45AC T2130005 2020-07-09 Sherman Barn RH & Temp Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
TE525MM 49187-109 2020-07-09 Sherman Barn Precipitation Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
ML2x 377/010 2020-07-09 Sherman Barn Soil Moisture Probe C Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus. This sensor was just below the surface and in the southern profile (relative to the other 2 profiles).
ML2x 54/016 2020-07-09 Sherman Barn Soil Moisture Probe B Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus. This sensor was near the surface, between the other two soil profiles.
ML2x 377/008 2020-07-09 Sherman Barn Soil Moisture Probe A Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus. This sensor was near the surface in the northern profile (relative to the other 2 profiles).
NR01 1570 2020-07-09 Sherman Barn Net Radiometer 4-way Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
SRS-Ni NDVI 882103061 2020-07-09 Sherman Barn Narrow Band NDVI Reflected Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
SRS-Ni NDVI 882103036 2020-07-09 Sherman Barn Narrow Band NDVI Incoming Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
SRS-Pi PRI 913003007 2020-07-09 Sherman Barn Narrow Band PRI Reflected Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
SRS-Pi PRI 913003006 2020-07-09 Sherman Barn Narrow Band PRI Incoming Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
PQS 1 110338 2020-07-09 Sherman Barn PAR Sensor Reflected Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
PQS 1 110339 2020-07-09 Sherman Barn PAR Sensor Incoming Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
HFP01 6498 2020-07-09 Sherman Barn Heat Flux Plate C Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus. This sensor was in the southern soil profile relative to the other two profiles.
HFP01 6491 2020-07-09 Sherman Barn Heat Flux Plate B Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus. This sensor was in the middle soil profile, located between the north and south profile. The sensor was very near the surface, with only litter covering it.
HFP01 6342 2020-07-09 Sherman Barn Heat Flux Plate A Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus. This sensor was in a northern position relative to the other soil profiles.
CS451 20010046 2020-07-09 Sherman Barn Water Level Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
WM 1590 153802 2020-07-09 Sherman Barn Anemometer Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
LI-7700 TG1-0357 2020-07-09 Sherman Barn Gas Analyzer CH4 Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
LI-7500A 75H-2668 2020-07-09 Sherman Barn Gas Analyzer CO2/H2O Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
LI-7550 AIU-1557 2020-07-09 Sherman Barn Interface LI-7550 Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
AM25T 6767 2020-07-09 Sherman Barn Multiplexer Thermocouples Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
CR1000 47108 2020-07-09 Sherman Barn Data Logger Met Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
AM16/32B 11662 2020-07-09 Sherman Barn Multiplexer Met Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
NL115 6763 2020-07-09 Sherman Barn Interface NL115 Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
SD130BN 0030F4D2148D 2020-07-09 Sherman Barn Camera Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
S2-431 1061 2020-07-13 Lab none Tested Lab none I tried to talk to the sensor with the new Apogee SDI-12 to USB dongle and Apogee software - no luck. The same method worked just fine foe the new NDVI sensors. Next will be to return them to Apogee for evaluation.
S2-431 1063 2020-07-13 Lab none Tested Lab none I tried to talk to the sensor with the new Apogee SDI-12 to USB dongle and Apogee software - no luck. The same method worked just fine foe the new NDVI sensors. Next will be to return them to Apogee for evaluation.
PS-15M 12501588 2020-07-07 Lab none Installed Tonzi none Installed in the Soil CO2 box under the tree at Tonzi. I swapped out an old 1st gen 30A model sn 186580399.
PS-30M 186580399 2020-07-07 Tonzi Removed Lab none Removed from the Soil CO2 box under the tree. This is an old 1st generation model. I put in a newer 15A 2nd model sn 12501588 since the power requirements here are small.
PS-15M 12501584 2020-07-07 Lab none Installed Tonzi none Installed at Tonzi to replace sn 0127000A for use elsewhere. Since updating the solar power to 24V the 30A model is not required.
PS-30M 0127000A 2020-07-07 Tonzi none Removed Lab none Removed from the Floor solar system. I needed a 30A model for the tower going to Steve and Sabina. Since updating this system 24V the amp load has decreased below 15A. Replaced by sn 12501584.
PS-30M 0135002X 2018-01-11 Lab none Installed Mayberry none Installed at MB to replace sn 04350908, which was full of ants.
MNKID-B KB15214 2020-05-21 Lab none Installed Sherman Wetland Solar Controller Installed at Sherman Wetland to replace sn KB04456 that has a blank LCD for testing in the lab.
MNKID-B KB04456 2020-05-21 Sherman Wetland Solar Controller Removed Lab none Removed from Sherman Wetland Because its LCD is blank and there were some power wobbles a couple days. Replaced by sn KB15214.
GMP222 Z2720012 2008-03-06 Lab none Installed Tonzi CO2 Probe Tree1 Profile -0.16 Installed in the Wast under tree profile at -16cm. mV and ppm readings in air were recorded before installation.
GMP222 Z2720011 2008-03-06 Lab none Installed Tonzi CO2 Probe Tree1 Profile -0.08 Installed in the Wast under tree profile at -8cm. mV and ppm readings in air were recorded before installation.
GMP222 X5050014 2008-03-06 Lab none Installed Tonzi CO2 Probe Tree1 Profile -0.04 Installed in the Wast under tree profile at -4cm. mV and ppm readings in air were recorded before installation.
GMP222 Z2720013 2008-03-06 Lab none Installed Tonzi CO2 Probe Tree1 Profile -0.02 Installed in the Wast under tree profile at -2cm. mV and ppm readings in air were recorded before installation.
GMP222 X1920001 2008-03-06 Lab none Installed Tonzi CO2 Probe Open Profile -0.16 Installed in the East open grass profile at -16cm. mV and ppm readings in air were recorded before installation.
GMP222 Z2720014 2008-03-06 Lab none Installed Tonzi CO2 Probe Open Profile -0.08 Installed in the East open grass profile at -8cm. mV and ppm readings in air were recorded before installation.
GMP222 X4840007 2008-03-06 Lab none Installed Tonzi CO2 Probe Open Profile -0.04 Installed in the East open grass profile at -4cm. mV and ppm readings in air were recorded before installation.
GMP222 X3830001 2008-03-06 Lab none Installed Tonzi CO2 Probe Open Profile -0.02 Installed in the East open grass profile at -2cm. mV and ppm readings in air were recorded before installation.
HFP01SC 00243 2014-12-10 Twitchell Rice Heat Flux Plate A Removed Lab none Removed from TW and brought back to the lab for troubleshooting
LI-7500A 0042 2000-12-14 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 0035 for calibration: H2O = 0-1000mmol/m3, CO2 = 13-20mmol/m3.
LI-7700 TG1-0100 2010-07-08 Loaned none Installed Sherman Island Gas Analyzer CH4 Open-path Moved from Frank's tower in the middle of East pond to Sherman Island.
CS450 20010770 2020-06-16 Sherman Wetland-Temp1 Water Level Notes Sherman Wetland-Temp1 Water Level Top of pipe to ground = 74cm
Top of pipe to water level = 130cm
Water level reading = 0.45psi ~ 32 cm
Top of pipe to sensor = Did not measure, calculated to be 162cm, which is 14cm greater from last measurement 2 months ago. Grr
PAR-LITE 040412 2018-09-28 Lab none Calibration Lab none Calibrated in the lab by sticking it out the south-facing window in Hilgard 205 with the reference sensor sn 161073. The factory calibration was 5.58 µV/(µmol/m2·s) and the new calibration was 5.74 µV/(µmol/m2·s). For details see the RT report "PAR comparison 09-2018" https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=14
PAR-LITE 010159 1999-02-09 Factory none Calibration Factory none Initial factory calibration 5.16 uV/(umol/m2/s).
PQS 1 140455 2018-09-28 Lab none Calibration Lab none Calibrated in the lab by sticking it out the south-facing window in Hilgard 205 with the reference sensor sn 161073. The factory calibration was 8.15 µV/(µmol/m2·s) and the new calibration was 8.47 µV/(µmol/m2·s). For details see the RT report "PAR comparison 09-2018" https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=14
PAR-LITE 010159 2018-09-28 Lab none Calibration Lab none Calibrated in the lab by sticking it out the south-facing window in Hilgard 205 with the reference sensor sn 161073. The factory calibration was 5.16 µV/(µmol/m2·s) and the new calibration was 5.59 µV/(µmol/m2·s). For details see the RT report "PAR comparison 09-2018" https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=14
LI-7500A 75H-2514 2020-05-22 Lab Calibration Lab Last calibration: 2020-03-02
AGC=100.5 SDM=0 DAC1=- DAC2=-
Ambient P=100.3kPa, Tair=23.2C, Tdew=9.7C
Zero Air at 750psi, AMP CA04396 408.97 at 600 Tdewspan=16.0
CO2 zero 0.9051 -> 0.9048 Offset: -1.2ppm
CO2 span 1.0104 -> 1.0090 Offset: 0.63ppm
H2O zero 0.8393 -> 0.8405 Offset: 8.4ppm
H2O span 1.0173 -> 1.0148 Offset: 0.04ppm
Notes:
LI-7500A 75H-2180 2020-05-22 Lab Calibration Lab Last calibration: 2020-02-03
AGC=99.5 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=23.2C, Tdew=9.6C
Zero Air at 750psi, AMP CA04396 408.97 at 600 Tdewspan=16.0
CO2 zero 0.9137 -> 0.9140 Offset: 1.4ppm
CO2 span 1.0319 -> 1.0317 Offset: 0.18ppm
H2O zero 0.8884 -> 0.8899 Offset: 9.5ppm
H2O span 1.0125 -> 1.0184 Offset: -0.13ppm
Notes:
LI-7500A 75H-2669 2020-05-22 Lab Calibration Lab Last calibration: 2020-02-04
AGC=100.7 SDM=0 DAC1=- DAC2=-
Ambient P=100.2kPa, Tair=23.5C, Tdew=8.3C
Zero Air at 750psi, AMP CA04396 408.97 at 600 Tdewspan=16.0
CO2 zero 0.9184 -> 0.9185 Offset: 0.44ppm
CO2 span 1.016 -> 1.0124 Offset: 1.78ppm
H2O zero 0.9006 -> 0.8999 Offset: -4.3ppm
H2O span 1.0101 -> 1.0094 Offset: 0.01ppm
Notes:
miniDOT 1220 2018-02-28 Mayberry DO Sensor E Moved Mayberry DO Sensor E We're guessing this is when the sensor was moved from -62 to -45cm.
CS451 70011236 2020-03-11 Mayberry Water Level Notes Mayberry Water Level Before we moved the water level sensor:
Top of well to WL = 71cm
Top of well to probe = 86cm
After we moved the water level sensor:
Top of well to WL = 69cm
Top of well to probe = 91cm (therefore, water level = 22 cm)
Sensor reading = 0.31 psi (x 70.41 psi/cm = 21.8 cm – good match with physical reading!)
Staff gauge = 27cm
CS450 20010770 2020-04-24 Sherman Wetland-Temp1 Water Level Notes Sherman Wetland-Temp1 Water Level Top of well to water level = 68cm
Top of well to ground = 74cm
Sensor psi reading = 1.14psi ~ 80cm
Top of well to sensor = ?? Did not measure, calculated to be 148cm
CS451 20010046 2020-02-06 Sherman Barn Water Level Notes Sherman Barn Water Level Camilo measured water level:
Distance from top of well to water level: 88.1cm (2.89ft)
Distance from top of well to ground: 24.7cm (0.81ft)
Distance from top of well to sensor: ??
Sensor psi reading: 0.653 psi ~ 46cm
2300-000 2300565 2020-06-15 Lab none Maintenance Lab none I set the unit up to log data. The data logging interval is 30sec and the wiper does 2 revolutions. The logging interval limits the number of wiper revolutions but doesn't seem to affect the data output on the serial line which seems to be 1Hz.
S2-431 1061 2020-06-11 Lab none Tested Lab none The two red/far red sensors stopped working on the BA tower. I tested them on the CR1000 in the lab running the BA program and still got no response. I tried the CR1000 terminal to talk to them over SDI-12 and got nothing trying both C1 and C3 ports. I will get a USB to SDI-12 from Apogee soon and will try again with that.
S2-431 1063 2020-06-11 Lab none Tested Lab none The two red/far red sensors stopped working on the BA tower. I tested them on the CR1000 in the lab running the BA program and still got no response. I tried the CR1000 terminal to talk to them over SDI-12 and got nothing trying both C1 and C3 ports. I will get a USB to SDI-12 from Apogee soon and will try again with that.
eosFD FD20180047 2020-06-11 Lab none Tested Lab none This unit was non-responsive in the field. In the lab it started right up and I was able to collect and clear its data. These units have had an issue with not starting up when they are too hot though this is supposed to be fixed in newer units. Maybe this unit has the too-hot-to-start problem and this combined with the power outage observed at Bouldin Corn to knock it out. Maybe I'll ask Camilo to put it back on Monday.
2300-000 2300565 2020-06-11 Lab none Tested Lab none I plugged in the fluorometer in the lab with the manufacturer supplied power supply and serial cable. The unit seems to be working and using less than 3.4A at 12V supplied by the power supply. I'm wondering if there is some short in the cable in the field. The unit also needs its wiper and data output set up before it can be used in the field.