Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 [71]
Ser No.DateLoc FromTitleLoc ToNotes
HMP45AC U3030041 2009-12-21 Lab none Tested Lab none I tested the HMP45AC on a CR1000 against HMP45C sn U2120013. No problems found. See file HMP45A_CalCheck.xls.
NR-LITE 041556 2009-10-28 Lab none Installed Tonzi-Tower The NRLite was installed on the top of the Tonzi tower by Joe for a calibration of the permanent tower NRLite.
NR-LITE 041556 2005-05-01 Factory Calibration Lab none Initial factory calibration: 14.2 uV/W.m2
NR-LITE 041556 2009-12-22 Tonzi-Tower Removed Lab none The instrument was removed from the top of the Tonzi tower by Siyan. It had been installed there for a calibration of the permanent NRLite on the tower.
LI-7500A 0041 2009-12-09 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed for the Twitchell tower and returned to the lab.
LI-7500A 0041 2009-12-14 Lab none Upgrade Lab none Last cal Sep 15, 2009
AGC = 55, Embedded software v3.0.0 upgraded to v3.0.1
Internal chemicals changed.
LI-7500A 0035 2009-12-03 Lab none Calibration Lab none AGC = 58, Pressure = 100.8kPa, Temperature = 20.2C
span CO2 = 594.7ppm, span H2O = 16.7C from dewpoint generator

Note: Scrubs on N2 tank causing error - removed.

Last calibration Aug 26, 2009
Last CO2 zero = 0.9344, span = 1.0120
Last H2O zero = 0.7295, span = 0.9425

New CO2 zero = 0.9342, span = 1.0118
New H2O zero = 0.7335, span = 0.9722

Configuration downloaded and saved
LI-7500A 0035 2009-11-30 Lab none Upgrade Lab none Embedded software v3.0.0 upgraded to v3.0.1
Internal chemical swapped.
GMP221 Y4110008 2005-05-16 Tonzi Removed Lab none Removed from Tonzi - location at Tonzi unknown
GMP221 Y0850004 2005-05-16 Tonzi CO2 Probe Open Profile -0.16 Removed Lab none Removed from Tonzi in the east open area set at 16cm deep
GMP221 Y0710013 2003-03-17 Lab none Installed Tonzi CO2 Probe Tree1 Profile -0.16 Installed at Tonzi in the west under canopy set at 16cm deep.
GMP221 Y0850004 2003-02-17 Lab none Installed Tonzi CO2 Probe Open Profile -0.16 Installed at Tonzi in the east open set at 16cm deep.
LI-7500A 0042 2009-12-08 Lab none Calibration Lab none Embedded software updated to v3.0.1
Last calibration Sept 15 2009
AGC=67, Pressure=101.01kPa Temperature=17.3C
CO2 span=597.4ppm, H2O span=14.5C
Precal CO2 zero=0.9609, span=0.9818
Postcal CO2 zero=0.9615, span=0.9897
Precal H2O zero=0.7855, span=1.0086
Postcal H2O zero=0.7844, span=0.9773
CO2 zero off by 2.2ppm, span off by ~7ppm
H2O span off by ~0.7C
LI-7500A 0042 2009-12-07 Lab none Upgrade Lab none Internal chemicals changed.
AGC = 69, Embedded software v3.0.0
Last calibration Sep 15 2009
GMP221 Y4110008 2009-12-01 Factory Repair Factory Vaisala factory repair and calibration 0-3% - certificate on file.
Problem: Got probe error 10 and 11. No IR source.
Action: Replaced probe circuit board and adjusted unit.
GMP221 X1520002 2009-12-01 Factory Repair Factory Vaisala factory repair and calibration 0-2% - certificate on file.
Problem: Got probe error 10 and 11. No IR source.
Action: Replaced probe circuit board and adjusted unit.
GMP221 Y1720006 2009-12-01 Factory Repair Factory Vaisala factory repair and calibration 0-2% - certificate on file.
Problem: Got probe error 05, 10, and 11. No IR source and memory error.
Action: Replaced probe circuit board and adjusted unit.
GMP221 Y2009100 2009-12-01 Factory Repair Factory Vaisala factory repair and calibration 0-3% - certificate on file.
Problem: Got probe error 05 and 11. No IR source and memory error.
Action: Replaced probe circuit board and adjusted unit.
GMP222 Y2009109 2009-12-01 Factory Repair Factory Vaisala factory repair and calibration 0-1% - certificate on file.
Problem: Got probe error 10 and 11. No IR source
Action: Replaced probe circuit board and adjusted unit.
GMP222 Z2720015 2009-12-01 Factory Repair Factory Vaisala factory repair and calibration 0-1% - certificate on file.
Problem: Got probe error 05 and 10. Memory error and no IR source.
Action: Replaced probe circuit board and adjusted unit
GMP221 Y0710013 2009-12-01 Lab none Calibration Lab none Vaisala factory calibration 0-3% - certificate on file.
Problem: Probe label is 3%. probe is spanned to 2% and is not scaled correctly.
Action: Re-spanned probe to 3%, udated coefficients and adjusted unit
GMP221 Y0850004 2009-12-01 Factory Calibration Factory Vaisala factory calibration 0-2% - certificate on file.
Problem: none
Action: Updated coefficients and adjusted unit.
GMP222 Y2030049 2009-12-01 Factory Calibration Factory Vaisala factory calibration 0-1% - certificate on file.
Problem: none
Action: Updated coefficients and adjusted unit.
GMP222 Y1950005 2009-12-01 Lab none Calibration Lab none Vaisala factory calibration 0-1% - certificate on file
Problem: Out of tolerance.
Action: Updated coefficients and adjusted unit.
HMP45AC U3030041 2009-12-01 Factory Calibration Lab none Vaisala factory calibration - certificate on file.
Problem: The probe was out of tolerance. The modified probe cable was not working. Intermittent readings.
Action: Probe was calibrate.
0...100% = 0...1V, -40...+60C = 0...1V
LI-7500A 0065 2009-12-03 Lab none Calibration Lab none AGC = 62, Pressure = 99.73kPa, Temperature = 21.63C
span CO2 = 594.7ppm, span H2O = 16.7C from dewpoint generator
Last calibration Aug 26, 2009
Last CO2 zero = 0.8738, span = 0.9939
Last H2O zero = 0.6414, span = 1.0055

New CO2 zero = 0.8739, span = 0.9931
New H2O zero = 0.6378, span = 0.9992

Configuration downloaded and saved
LI-7500A 0065 2009-11-30 Lab none Upgrade Lab none Current embedded software v3.0.0 upgraded to v3.0.1. Internal chemicals changed.