Equipment and Calibration History

View Equipment |

View history entry 22 for Licor LI-7500A sn: 0035

Date: 2009-12-03
From: Lab - none
To: Lab - none
Title: Calibration
Notes: AGC = 58, Pressure = 100.8kPa, Temperature = 20.2C span CO2 = 594.7ppm, span H2O = 16.7C from dewpoint generator Note: Scrubs on N2 tank causing error - removed. Last calibration Aug 26, 2009 Last CO2 zero = 0.9344, span = 1.0120 Last H2O zero = 0.7295, span = 0.9425 New CO2 zero = 0.9342, span = 1.0118 New H2O zero = 0.7335, span = 0.9722 Configuration downloaded and saved
Entered by: