Site | Date | Author | Summary |
East End |
(351) 2020-12-16 12:50:00 |
Daphne Szutu |
Regular data collection/cleaning, hunting day, 7700 spin problems?, Apogee_in still has gaps
|
East End |
(338) 2020-12-03 13:10:00 |
Daphne Szutu |
Regular data collection/cleaning, swapped 7500 for calibration, adjusted eddy clock -7min, moved SHF plates to new channel to avoid gaps, tested precip bucket--needs more testing
|
East End |
(324) 2020-11-19 14:05:00 |
Joe Verfaillie |
Tested tipping bucket, calibrated 7700, some NaNs in SHFP1 and Apogee NDVI but no obvious problem, leveled diffuse PAR, fixed leak in 7700 wash tubing
|
East End |
(308) 2020-11-03 10:35:00 |
Daphne Szutu |
Regular data collection/cleaning, water level a little lower, empty wash reservoir and dusty lower 7700 mirror
|
East End |
(296) 2020-10-22 11:50:00 |
Daphne Szutu |
Regular data collection/cleaning, a little more water under the tower, adjusted water level sensor height, changed eddy clock -3min, lots of spiderwebs and cattail fluff
|
East End |
(279) 2020-10-05 09:30:00 |
Daphne Szutu |
Regular data collection/cleaning, hazy air, swap out battery bank for 4 new AGM batteries, water level a little higher (finally), removed eosGP
|
East End |
(273) 2020-09-29 08:30:00 |
Joe Verfaillie |
Testing solar power battery bank, system running on one new battery temporarily
|
East End |
(267) 2020-09-23 11:45:00 |
Daphne Szutu |
Regular data collection/cleaning, tower losing overnight/predawn power, still dry at the tower, many wasps
|
East End |
(254) 2020-09-10 11:50:00 |
Joe Verfaillie |
Wetland is dry and yellow, Tested batteries - one weak battery removed from battery bank, replaced HMP fan but need correct voltage fan, took apart Pi-cam to clean view.
|
East End |
(245) 2020-09-01 15:40:00 |
Daphne Szutu |
Regular data collection/cleaning, hazy air, water level still low, some power problems--lost a few days of pre-dawn data
|
East End |
(241) 2020-08-28 15:05:00 |
Daphne Szutu |
Quick stop to update CR1000 program for Apogee NDVI sensors, bad air quality from wildfire smoke
|
East End |
(234) 2020-08-21 10:00:00 |
Daphne Szutu |
Regular data collection and cleaning, terrible air quality because of wildfire smoke, water level is still low, swapped 7500 for calibration
|
East End |
(218) 2020-08-05 10:45:00 |
Daphne Szutu |
Regular data collection/cleaning, no water under tower, installed water level sensor, measured water level, installed new Apogee NDVI sensors
|
East End |
(203) 2020-07-21 11:25:00 |
Daphne Szutu |
Regular data collection/cleaning, no water under the tower, no power at datalogger--fixed, lost 1 day of met data, removed water level sensor for repair, installed Picam as phenocam, soil samples for Kuno
|
East End |
(183) 2020-07-01 09:35:00 |
Daphne Szutu |
Regular data collection/cleaning, camera stolen again, no water at the tower
|
East End |
(161) 2020-06-09 09:37:00 |
Joe Verfaillie |
Standard data collection, water level pretty low
|
East End |
(150) 2020-05-29 09:50:00 |
Joe Verfaillie |
Replaced diffuse PAR motor, calibrated conductivity sensor, restarted camera.
|
East End |
(140) 2020-05-19 14:45:00 |
Daphne Szutu |
Regular data collection/cleaning, 7700 calibration, swapped 7500 for calibration, changed eddy clocks -7 min, df PAR motor dead--needs to be replaced, no photos on camera card
|
East End |
(122) 2020-05-01 12:45:00 |
Daphne Szutu |
Regular data collection/cleaning, breezy, diffuse PAR not working--fixed, water level up a little, first site visit in a month.
|
East End |
(94) 2020-04-03 13:40:00 |
Daphne Szutu |
Regular data download/cleaning, 7550 time sync error, first site visit in 3 weeks
|
East End |
(71) 2020-03-11 10:15:00 |
Daphne Szutu |
Replaced boardwalk wood, moved met sensors to new pluggable strip, 7700 clock sync error, low water level
|
East End |
(65) 2020-03-05 12:05:00 |
Daphne Szutu |
Regular data collection/cleaning, adjusted 7550 clock +10 min, 7700 not connected--fixed, smoky horizons
|
East End |
(51) 2020-02-20 12:25:00 |
Joe Verfaillie |
Standard data collection, swapped 7500, serviced Minidot, wasps trying to escape
|
East End |
(37) 2020-02-06 13:40:00 |
Daphne Szutu |
Regular data collection/cleaning, gaps in PARin--moved to different mux channel, uploaded updated CR1000 program
|
East End |
(23) 2020-01-23 10:20:00 |
Daphne Szutu |
Regular data collection/cleaning, PARin had gaps to keep an eye on
|
East End |
(7) 2020-01-07 12:55:00 |
Daphne Szutu |
Regular data download/collection, cold and foggy, serviced water sensors, updated 7700 wash spin settings, 7550 time sync error-fixed, some camera data missing
|