Equipment and Calibration History

History for Licor LI-7700 TG1-0421

Rec No.DateLoc FromTitleLoc ToNotes
6904 2024-10-10 Lab none Installed Tonzi Gas Analyzer CH4 Tower Installed on top of the Tower by adding a new mount using L-brackets and a rail fitting removing existing Allen screws and reusing the sockets for standard 1/4-20 bolts. Wash box and spiked hat also installed.
6903 2024-10-08 Lab Calibration Lab Calibration Location: Lab
Last calibration: 2024-02-16
RSSI=84.5%, Optical RH=18%
Ambient P=100.05kPa, Tair=30.0C
Zero Air at 1250psi, AMP FB04035 1.9156 at 700
CH4 Zero Offset: -0.024ppm
CH4 Span Offset: 0.0294ppm
Notes: Returned from Yampah in Elk Horn Slough. Kind of muddy all over - cleaned up. Cables to come back later.
6902 2024-10-08 Loaned none Returned Lab none Returned to the Lab by Robert from UCSC. J.J. brought Ethernet, power, wash cables and tubing back on 10/11.
6570 2024-02-16 Lab Calibration Lab Calibration Location: Lab
Last calibration: 2023-02-02
RSSI=81%, Optical RH=12%
Ambient P=100.98kPa, Tair=21.5C
Zero Air at 450psi, AMP CA04119 4.2238 at 675
CH4 Zero Offset: .032ppm
CH4 Span Offset: .0058ppm
Notes: Calibration before loaning this sensor to J.J. to use at Elkhorn Slough Yampah tower while they're waiting for replacement/repair
6569 2024-02-16 Lab none Loaned Loaned none Loaned sensor to Sylvain and J.J. (UCSC) to use at the Elkhorn Slough Yampah tower while they wait for a replacement sensor/repairs. Spiky hat was not loaned. Cables were loaned: 7700 yellow power cable, 7700 green Ethernet cable, wash black power cable, wash black tubing without spray nozzle (spray nozzle loaned previously for UCSC to use at Elkhorn Porter tower)
6531 2024-01-11 Hill Slough Gas Analyzer CH4 Removed Lab none Removed from HS and brought back to the lab for calibration. It was replaced by sn TG1-0155
6328 2023-07-11 Lab none Installed Hill Slough Gas Analyzer CH4 Installed at Hill Slough to replace sn TG1-0211 for calibration.
6315 2023-07-10 Factory none Calibration Lab none Received from the factory back at the lab. 13.8 hours of labor went into repairing the extensive water damage and instrument calibration. 18 parts were replaced.
6233 2023-03-28 Lab none Shipped Factory none Shipped back to Licor for repair. Expecting a 4-6 months turnaround time
6232 2023-03-24 Loaned none Removed Lab none Removed from the BWW urban tower because it was not working. Joe opened it and found a bunch of water inside the bottom can. There is corrosion on the boards and will need to be returned to Licor for repair.
6223 2023-02-02 Lab none Installed Loaned none Re-installed on the BWW tower after calibration in the lab
6184 2023-02-02 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2022-06-01
RSSI=79%, Optical RH=23%
Ambient P=101.3kPa, Tair=19.5C
Zero Air at 1400psi, AMP CA04119 4.2238 at 1275
CH4 Zero Offset: -.011ppm
CH4 Span Offset: .0138ppm
Notes: First calibration since sensor was installed on BWW urban tower 7 months ago. After calibration, swapped desiccant and flushed headspace with zero air until new optics RH=10%. Sensor removed this morning and will go back up on the BWW urban tower this afternoon.
5912 2022-07-15 Lab none Loaned Loaned none Installed on the BWW urban tower and updated the GHG columns. I also connected the HMP back to the 7700 auxiliary input. However, I couldn't remember the correct coefficients for the HMP. I tried the same multiplier and offset we use for CRBasic, but that didn't produce reasonable data. I e-mailed Joe for help and he fixed it the following day.

This sensor still needs a wash box and wash power cable. These will have to come from BA after our next visit. The current wash tubing is a spliced one, so I could also replace that with new black tubing.
5908 2022-07-05 Factory none Returned Lab none Received back in the lab after repair and calibration at Licor. They replaced the fuse and fuse holder and calibrated the instrument.
5909 2022-06-01 Factory Calibration Factory Calibration Location: Factory
Last calibration: 2021-11-30
RSSI=0%, Optical RH=0%
Ambient P=0kPa, Tair=0C
0 at 0psi, 0 0 at 0
CH4 Zero Offset: 10.1875ppm
CH4 Span Offset: .000103546ppm
Notes: Calibrated at Licor factory after replacing fuse holder and fuse. Not sure of the offset units (but probably not in ppm?)
5907 2022-04-28 Lab none Shipped Factory none Sent back to Licor for inspection/repair around this date.
5787 2022-04-15 Lab none Tested Lab none Plugged it in in the lab. The power LED was blinking intermittently and the ethernet and status LEDs were off. The fuse and fuse connector were totally rotten and the inside of the sensor body had condensation and a few tsp of water. We didn't see any obvious corrosion on the other circuit boards and no obvious leaks/cracks in the sensor. We wiped up the water and will let it air dry before sending it back to Licor.
5790 2022-04-13 West Pond Gas Analyzer CH4 Removed Lab none Removed from West Pond and brought back to the lab. I couldn't connect to this sensor. The power LED light was blinking intermittently. Ethernet and Status LEDs were off. I removed the power cable as well since there was a crack in the plastic cable collar at the sensor. The wash power, wash tubing, and Ethernet cable are still there.
5612 2021-11-30 West Pond Calibration West Pond Calibration Location: West Pond
Last calibration: 2021-06-01
RSSI=87%, Optical RH=18%
Ambient P=101.4kPa, Tair=18.6C
Zero Air at 100psi, AMP CA04396 1.9317 at 1750
CH4 Zero Offset: .050ppm
CH4 Span Offset: .0023ppm
Notes: Relatively still day. The zero air tank pressure may have been >100psi, but the regulator was wonky. Zero was stable for 2min after applying zero.
5278 2021-05-26 West Pond Calibration West Pond Calibration Location: West Pond
Last calibration: 2020-11-19
RSSI=80%, Optical RH=12%
Ambient P=101.2kPa, Tair=29.1C
AMP FB04161 at 800psi, AMP CA04119 4.223 at 2100
CH4 Zero Offset: -0.024ppm
CH4 Span Offset: 0.043ppm
Notes: Relatively still day. Zero gas was an AMP cal gas in a baby tank that we borrowed for the day. Zero offset and span are a bit higher than I usually see on 7700s. A lot of bounce in concentrations from Dennis and Robert talking and wiggling the upper boardwalk.
4982 2020-11-19 West Pond Calibration West Pond Calibration Location: West Pond
Last calibration: 2020-05-19
RSSI=80%, Optical RH=8%
Ambient P=102.0kPa, Tair=17.0C
Zero Air at 500psi, AMP CA3923 3.121 at 500
CH4 Zero Offset: 0.005ppm
CH4 Span Offset: 0.002ppm
Notes: Still day. This sensor was originally at BA before we moved it to WP to calibrate. The sensor usually at WP (sn TG1-0224) was removed and brought back to the lab because its serial number was not recognized in the 7700 or 7550 software.
4970 2020-11-19 Bouldin Alfalfa Gas Analyzer CH4 Moved West Pond Gas Analyzer CH4 Removed from Bouldin Alfalfa for calibration. Replaced by sn TG1-0211. Later we discovered that the 7700 at WP was behaving weirdly so we swapped this instrument in there are calibrated it there.
4456 2020-05-19 Bouldin Alfalfa Calibration Bouldin Alfalfa Calibration Location: Bouldin Alfalfa
Last calibration: 2019-10-22
RSSI=76%, Optical RH=4%
Ambient P=101.2kPa, Tair=19.0C
Zero Air at 900psi, AMP CA3923 3.121 at 975
CH4 Zero Offset: -0.082ppm
CH4 Span Offset: 0.004ppm
Notes: Breezy day
4114 2019-10-22 Bouldin Alfalfa Gas Analyzer CH4 Calibration Bouldin Alfalfa Gas Analyzer CH4 Calibration Location: Bouldin Alfalfa
Last calibration: 2019-04-11
RSSI=84%, Optical RH=22%
Ambient P=101.8kPa, Tair=20.9C
Zero Air at 1300psi, AMP CA03923 3.121 at 1400
CH4 Zero Offset: 0.022ppm
CH4 Span Offset: 0.001ppm
Notes: Swapped optics desiccant after calibration using cal gas to dry head space. New RH 7%
3910 2019-04-11 Bouldin Alfalfa Gas Analyzer CH4 Upgrade Bouldin Alfalfa Gas Analyzer CH4 Upgraded LI-7700 sn TG1-0421 firmware from 1.0.23 to 1.0.27. To connect, Joe had to change the laptop network subnet from 192.168.2.5 to 192.168.1.5 (to match the LI-7700 IP address of 192.168.1.77), even though the subnet mask was 255.255.255.0. We then restarted the instrument and calibrated LI-7700 sn TG1-0421.
3898 2019-04-11 Bouldin Alfalfa Gas Analyzer CH4 Calibration Bouldin Alfalfa Gas Analyzer CH4 Calibration Location: Bouldin Alfalfa
Last calibration: 2018-11-06
RSSI=83%, Optical RH=17%
Ambient P=101.3kPa, Tair=19.2C
Zero Air M4002 at 1600psi, AMP CA03923 3.121 at 1600
CH4 Zero Offset: 0.037ppm
CH4 Span Offset: 0.007ppm
Notes: Breezy day, upgraded firmware from v1.0.23 to v1.0.27 before calibration
3885 2019-02-12 Bouldin Alfalfa Gas Analyzer CH4 Tested Bouldin Alfalfa Gas Analyzer CH4 Tried to upgrade firmware from 1.0.23 to 1.0.27, but there was an error, "Fatal: Network Error: Connection Timed out."
3834 2018-11-07 Lab none Installed Bouldin Alfalfa Gas Analyzer CH4 Installed at BA. The previous LI-7700 sn TG1-0424 was removed for a power board swap.
3793 2018-11-06 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2018-05-14
RSSI=81%, Optical RH=13%
Ambient P=100.2kPa, Tair=26.7C
Zero Air at 1350psi, AMP Air 1.9497 at 950
CH4 Zero Offset: -0.015ppm
CH4 Span Offset: 0.0083ppm
Notes: First calibration after power board replacement
3792 2018-11-05 Lab none Upgrade Lab none I swapped out power board pn 9677-079G per Licor's instructions. The new power board is pn 9977-079. According to Licor, "the original board needs a new resistor for the "under-voltage" problem (potential to drop out when heater and washers run simultaneously while incoming power is lower than usual, say 11-10 VDC)". The new board Rev 0A (V3) should not have that problem.
3786 2018-11-01 East End Gas Analyzer CH4 Removed Lab none Removed from EE and brought back to the lab for a power board replacement and calibration. It was replaced by sn TG1-0423
3469 2018-05-16 Lab none Installed East End Gas Analyzer CH4 Installed on the East End tower to replace sn TG1-0357 for calibration
3451 2018-05-14 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-10-16
RSSI=78%, Optical RH=8%
Ambient P=100.2kPa, Tair=21.8C
Zero Air at 325psi, LBL Air 1.9497 at 1400
CH4 Zero Offset: 0.024ppm
CH4 Span Offset: -.0067ppm
Notes:
3452 2018-05-10 Sherman Wetland Gas Analyzer CH4 Removed Lab none Removed from Sherman Wetland for calibration. Replaced by sn TG1-0423
3165 2017-10-26 Lab none Installed Sherman Wetland Gas Analyzer CH4 Installed at SW to replace sn TG1-0224.
3146 2017-10-16 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-07-13
RSSI=77%, Optical RH=3%
Ambient P=100.4kPa, Tair=29.7C
Zero Air at 500psi, AMP Air 1.9497 at 2000
CH4 Zero Offset: 0.020ppm
CH4 Span Offset: -0.0137ppm
Notes:
3105 2017-10-04 Bouldin Corn Gas Analyzer CH4 Removed Lab none Removed from the Bouldin Corn tower for calibration and testing in the lab for drop outs when washing problem. Replaced by sn TG1-0211.
2927 2017-07-14 Lab none Installed Bouldin Corn Gas Analyzer CH4 Installed at Bouldin Corn tower
2931 2017-07-13 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-04-03
RSSI=72.1%, Optical RH=3%
Ambient P=100.3kPa, Tair=22.4C
Zero Air at 1400psi, NOAA Air 1.89919 at 1500
CH4 Zero Offset: 0.010ppm
CH4 Span Offset: 0.0519ppm
Notes: First calibration after replacing the circuit power board. Ambient CH4 = 1.98
2926 2017-07-13 Lab none Repair Lab none Replaced the power supply circuit board (9977-073) in the LI-7700. The analyzer was having a few problems before this repair: atmosphere pressure was about 40-50 kpa, spin motor did not work, fluid pump did not work. Li-cor emailed the replacement board as a warranty replacement (no cost). We opened up the bottom enclosure plate and the enclosure cylinder to replace the board.

After the repair, the pressure read 100.1 kPa and the spin motor and fluid pump both worked.
2925 2017-07-12 Bouldin Corn Gas Analyzer CH4 Removed Lab none Brought back to the lab to replace its power board. Power board provided by Li-cor.
2716 2017-04-26 Lab none Installed Bouldin Corn Gas Analyzer CH4 Installed at 5.2m agl the Bouldin Corn site with the initial set up of the tower.
2681 2017-04-03 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2016-08-11
RSSI=77.7%, Optical RH=4%
Ambient P=100.2kPa, Tair=26.7C
Zero Air at 1650psi, NOAA CH4 1.89919 at 1700
CH4 Zero Offset: 0.021ppm
CH4 Span Offset: -0.035ppm
Notes:
2675 2017-03-23 Bouldin Alfalfa Gas Analyzer CH4 Removed Lab none Removed from Bouldin Alfalfa and brought back to the lab for calibration. It was replaced by TG1-0424.
2419 2016-08-12 Lab none Installed Bouldin Alfalfa Gas Analyzer CH4 Installed with the initial set up of the site.
2391 2016-08-11 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2015-08-10
RSSI=80.6%, Optical RH=27%
Ambient P=100.6kPa, Tair=22.2C
zero air at 500psi, AMP CH4 1.9966 at 400
CH4 Zero Offset: .029ppm
CH4 Span Offset: 0.0006ppm
Notes: First lab calibration since receiving this instrument. The desiccant was replaced - instrument was sitting idle in lab for almost a year.