Equipment and Calibration History

View Equipment |

View history entry 3898 for Licor LI-7700 sn: TG1-0421

Date: 2019-04-11
From: Bouldin Alfalfa - Gas Analyzer CH4
To: Bouldin Alfalfa - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Bouldin Alfalfa Last calibration: 2018-11-06 RSSI=83%, Optical RH=17% Ambient P=101.3kPa, Tair=19.2C Zero Air M4002 at 1600psi, AMP CA03923 3.121 at 1600 CH4 Zero Offset: 0.037ppm CH4 Span Offset: 0.007ppm Notes: Breezy day, upgraded firmware from v1.0.23 to v1.0.27 before calibration
Entered by: