Equipment and Calibration History

View Equipment |

View history entry 5278 for Licor LI-7700 sn: TG1-0421

Date: 2021-05-26
From: West Pond -
To: West Pond -
Title: Calibration
Notes: Calibration Location: West Pond Last calibration: 2020-11-19 RSSI=80%, Optical RH=12% Ambient P=101.2kPa, Tair=29.1C AMP FB04161 at 800psi, AMP CA04119 4.223 at 2100 CH4 Zero Offset: -0.024ppm CH4 Span Offset: 0.043ppm Notes: Relatively still day. Zero gas was an AMP cal gas in a baby tank that we borrowed for the day. Zero offset and span are a bit higher than I usually see on 7700s. A lot of bounce in concentrations from Dennis and Robert talking and wiggling the upper boardwalk.