Equipment and Calibration

View History

Equipment Entry

Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-7700
Serial No:TG1-0223
Tag No:NULL
Acquisition Date:2012-02-07
Project:DWR-Wetland
Location on 2023-08-08:Gilbert Tract, NULL
Notes: Part of GHG1 system with LI-7550 sn AIU-0650 and LI-7500A sn 75H-2176

Documents

Calibration: cal_Licor_LI-7700_snTG1-0223_20120119.pdf
Manual: Collecting_LI-7700_Ethernet_Data.pdf
Manual: LI-7700_Manual_V2_201103.pdf
Manual: LI-7700_Manual_V3_201312.pdf
PackingSlip: ps_Licor_20120207.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6805 2024-08-07 Gilbert Tract Calibration Gilbert Tract Calibration Location: Gilbert Tract
Last calibration: 2024-01-11
RSSI=85%, Optical RH=7%
Ambient P=100.5kPa, Tair=30.7C
Zero Air at 1600psi, AMP CA03923 3.3865 at 1800
CH4 Zero Offset: -.011ppm
CH4 Span Offset: -.0085ppm
Notes: Warm, breezy
6528 2024-01-11 Gilbert Tract Calibration Gilbert Tract Calibration Location: Gilbert Tract
Last calibration: 2023-08-07
RSSI=82%, Optical RH=3%
Ambient P=101.8kPa, Tair=9.7C
Zero Air at 1850psi, AMP CA03923 3.3865 at 1750
CH4 Zero Offset: .002ppm
CH4 Span Offset: -.0225ppm
Notes: Breezy
6363 2023-08-08 Lab none Installed Gilbert Tract Gas Analyzer CH4 Installed at Gilbert Tract to replace sn TG1-0424 for calibration.
6356 2023-08-07 Lab Calibration Lab Calibration Location: Lab
Last calibration: 2023-01-18
RSSI=83%, Optical RH=28%
Ambient P=100.1kPa, Tair=27.2C
Zero Air at 900psi, AMP CA04119 4.2238 at 900
CH4 Zero Offset: -.003ppm
CH4 Span Offset: -.0318ppm
Notes: After calibration, flushed optics headspace with zero air until optics RH=11%. Removed used silica desiccant and put in fresh drierite.
6352 2023-07-24 Mayberry Gas Analyzer CH4 Removed Lab none Removed from Mayberry for calibration. Replaced by sn TG1-0211.
6170 2023-01-18 Mayberry Calibration Mayberry Calibration Location: Mayberry
Last calibration: 2022-06-07
RSSI=84%, Optical RH=15%
Ambient P=101.9kPa, Tair=9.9C
Zero Air at 1500psi, AMP CA04396 1.9317 at 1350
CH4 Zero Offset: .011ppm
CH4 Span Offset: .0093ppm
Notes: Did not flush headspace or change desiccant afterwards
6129 2022-12-13 Mayberry Gas Analyzer CH4 Maintenance Mayberry Gas Analyzer CH4 The humidity reading was 31%. I swapped the desiccant bottle refilling it with silica gel desiccant. The humidity didn't change and hopefully will go down. Swap with regular dri-rite at calibration next time.
5914 2022-06-08 Lab none Installed Mayberry Gas Analyzer CH4 Installed at MB after attaching a new fuse holder and fuse. This sensor was calibrated in the lab yesterday.
5888 2022-06-07 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2021-12-03
RSSI=76%, Optical RH=29%
Ambient P=100.0kPa, Tair=21.7C
Zero Air at 625psi, AMP CA04119 4.2238 at 1600
CH4 Zero Offset: -.037ppm
CH4 Span Offset: .0392ppm
Notes: Before this calibration, I replaced fuse and fuse holder. I added some grease to the large o-ring around the base (Licor grease in the small plastic container) before reassembling everything. I followed Joe's advice to use longer helper screws help get the bottom to seat all the way in, and then replace those "helper" screws with the original screws with o-rings. During calibration for the span, I realized at 16:10 that the top heater had been on since 16:00. Probably the zero was calibrated while the heater was on. The span seemed stable from 16:12-16:18 and I don't think turning the heater off really affected it. I spanned the sensor at 16:18. After calibration, I pumped zero air into the headspace until optics RH=10%. I swapped out the old purple desiccant for a bottle of fresh blue desiccant. I added a bit of grease to the desiccant plug as well.
5880 2022-05-27 Mayberry Gas Analyzer CH4 Removed Lab none Removed from Mayberry because water got inside and rotted out the fuse on the bottom of the can. Same as WP a few weeks ago.
5604 2021-12-01 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 Calibration Location: Mayberry
Last calibration: 2021-06-01
RSSI=85%, Optical RH=22%
Ambient P=101.8kPa, Tair=9.3C
Zero Air at 600psi, AMP CA04396 1.9317 at 1650
CH4 Zero Offset: 0.034ppm
CH4 Span Offset: -0.0157ppm
Notes: A lot of fog in the air so it was hard for RSSI to go >60 even when clean. The RSSI increased to 85 when we put the hood on. Relatively still day. After calibration, we dried down the sensor headspace with zero air until optics RH=10%. We swapped out the old drierite for fresh drierite.
5591 2021-11-23 Mayberry Gas Analyzer CH4 Upgrade Mayberry Gas Analyzer CH4 Upgraded the firmware from v1.0.27 to v1.0.29. The sensor restarted succesfully and all seems well.
5276 2021-05-26 Mayberry Calibration Mayberry Calibration Location: Mayberry
Last calibration: 2020-11-16
RSSI=77%, Optical RH=25%
Ambient P=101.0kPa, Tair=27.8C
AMP FB03864 at 1200psi, AMP CA04119 4.223 at 2000
CH4 Zero Offset: -0.028ppm
CH4 Span Offset: -0.003ppm
Notes: Breezy day. Zero gas was an AMP cal gas in a baby tank that we borrowed for the day. After the calibration, we used the zero air to flush the headspace until optics RH = 10%. The drierite was purple, so we swapped in a new bottle with fresh drierite.
4945 2020-11-12 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 Calibration Location: Mayberry
Last calibration: 2020-05-22
RSSI=78%, Optical RH=26%
Ambient P=101.6kPa, Tair=11.6C
Zero Air at 700psi, AMP CA3923 3.121 at 700
CH4 Zero Offset: 0.011ppm
CH4 Span Offset: 0.002ppm
Notes: Still day. Swapped out pink silica gel and put in fresh drierite in the optics chem bottle. Used zero air to flush head space down to 9% RH. Optics RH was 12% when we left the site.
4486 2020-05-21 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 Calibration Location: Mayberry
Last calibration: 2019-10-22
RSSI=83%, Optical RH=26%
Ambient P=101.2kPa, Tair=21.8C
Zero Air at 750psi, AMP CA3923 3.121 at 800
CH4 Zero Offset: 0.004ppm
CH4 Span Offset: -0.003ppm
Notes: Breezy day. Swapped optics drierite after calibration and used cal gas to dry head space. New optics RH was 9%.
4118 2019-10-22 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 Calibration Location: Mayberry
Last calibration: 2019-04-25
RSSI=84.7%, Optical RH=28%
Ambient P=101.6kPa, Tair=29.7C
Zero Air at 1200psi, AMP CA03923 3.121 at 1200
CH4 Zero Offset: -0.075ppm
CH4 Span Offset: 0.007ppm
Notes: Slightly larger zero offset but span seems good. After calibration swapped optics desiccant and dried head space with cal gas. New RH 11%.
3949 2019-04-25 Mayberry Gas Analyzer CH4 Upgrade Mayberry Gas Analyzer CH4 The thermocouple module was broken so we swapped it for a new one. The LI-7700 firmware was upgraded from v1.0.21 to v1.0.27 mostly minor bug fixes see change log https://www.licor.com/env/support/LI-7700/topics/change-log.html
3936 2019-04-25 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 Calibration Location: MB
Last calibration: 2018-10-26
RSSI=80%, Optical RH=12%
Ambient P=100.9kPa, Tair=26C
Zero Air at 1600psi, AMP CA3923 3.121 at 1600
CH4 Zero Offset: 0.013ppm
CH4 Span Offset: 0.015ppm
Notes: TC module swapped before calibration; breezy day
3759 2018-10-25 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 Calibration Location: MB
Last calibration: 2018-04-25
RSSI=81%, Optical RH=8%
Ambient P=101.6kPa, Tair=20.5C
Zero Air at 2000psi, AMP Air 3.121 at 2100
CH4 Zero Offset: 0.021ppm
CH4 Span Offset: 0.018ppm
Notes: Calm day
3448 2018-05-03 Lab none Installed Mayberry Gas Analyzer CH4 Installed at MB to replace sn TG1-0423, which was brought back to the lab for calibration.
3424 2018-04-25 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-10-16
RSSI=84%, Optical RH=3%
Ambient P=100.4kPa, Tair=22.3C
Zero Air at 400psi, LBL Air 1.9497 at 1400
CH4 Zero Offset: 0.013ppm
CH4 Span Offset: -0.0027ppm
Notes:
3413 2018-04-19 Bouldin Alfalfa Gas Analyzer CH4 Removed Lab Gas Analyzer CH4 Removed from Bouldin Alfalfa for calibration. Replaced by sn TG1-0424
3184 2017-10-19 Lab none Installed Bouldin Alfalfa Gas Analyzer CH4 Installed at BA (now that the East End temporary towers are done, we have an available LI-7700 to put up here)
3145 2017-10-16 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-03-23
RSSI=79.2%, Optical RH=2%
Ambient P=100.5kPa, Tair=28.7C
Zero Air at 500psi, AMP Air 1.9497 at 2000
CH4 Zero Offset: -0.015ppm
CH4 Span Offset: -0.0177ppm
Notes:
3149 2017-10-13 West Pond Gas Analyzer CH4 Removed Lab none Removed from West Pond for calibration. Replaced by sn TG1-0424
2671 2017-03-23 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2016-10-11
RSSI=86.2%, Optical RH=16%
Ambient P=102.0kPa, Tair=17.6C
Zero Air at 750psi, Ameriflux CH4 4.0304 at 1100
CH4 Zero Offset: -0.041ppm
CH4 Span Offset: -0.0251ppm
Notes: Changed dessicant and optics RH = 27%. Flushed head with span gas and optics RH dropped to 16%.
2492 2016-10-11 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: west pond
Last calibration: 2016-04-13
RSSI=71%, Optical RH=27%
Ambient P=101.3kPa, Tair=22C
zero air at 700psi, AMP CH4 4.030 at 1200
CH4 Zero Offset: 0.001ppm
CH4 Span Offset: -0.006ppm
Notes: Breezy.
2235 2016-04-13 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2015-09-30
RSSI=84.2%, Optical RH=24%
Ambient P=101.74kPa, Tair=22.97C
Zero Air at 900psi, AMP CH4 1.9966 at 700
CH4 Zero Offset: 0.011ppm
CH4 Span Offset: 0.001ppm
Notes:
1950 2015-09-30 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2015-03-24
RSSI=66.5%, Optical RH=25%
Ambient P=101.3kPa, Tair=18.1C
Zero Air at 1050psi, AMP CH4 1.9966 at 1500
CH4 Zero Offset: 0.013ppm
CH4 Span Offset: -0.0066ppm
Notes: light, sporadic rain - no wind
1763 2015-03-24 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2014-09-10
RSSI=80%, Optical RH=19%
Ambient P=102.3kPa, Tair=20.8C
Zero air at 1200psi, 2nd CH4 4.995 at 600
CH4 Zero Offset: -0.001ppm
CH4 Span Offset: 0.08ppm
Notes:
1545 2014-09-10 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2014-03-07
RSSI=82%, Optical RH=12%
Ambient P=100.88kPa, Tair=31.5C
Zero Air at 1400psi, 2nd CH4 4.995 at 750
CH4 Zero Offset: 0.025ppm
CH4 Span Offset: -0.085ppm
Notes: Very hot day with no wind. Span offset a little large.
1229 2014-03-07 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2013-08-21
RSSI=73.8%, Optical RH=7%
Ambient P=102.4kPa, Tair=20.7C
Zero Air at 1600psi, 2nd CH4 4.995 at 1100
CH4 Zero Offset: 0.01ppm
CH4 Span Offset: 0.001ppm
Notes: Calibrated by Cove and Sara after firmware upgrade to v 1.0.21. Windy day. Windy day.
1050 2013-08-21 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2013-02-19
RSSI=78%, Optical RH=3%
Ambient P=101.1kPa, Tair=21.8C
Zero Air at 1700psi, CH4 in air 4.995 at 1300
CH4 Zero Offset: 0.01ppm
CH4 Span Offset: 0.013ppm
Notes:
931 2013-02-19 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration: 2012-09-06
RSSI=66.8%, Optical RH=1.0%
Ambient P=101.98kPa, Tair=10.8C
Zero Air at 1800psi, 2nd CH4 in Air 4.982 at 1300
CH4 Zero Offset: 0.016ppm
CH4 Span Offset: 0.006ppm
Notes:
839 2012-09-06 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: West Pond
Last calibration:
RSSI=69%, Optical RH=%
Ambient P=kPa, Tair=C
ultra zero air at 1900psi, ch4 in air 4.982 at 950
CH4 Zero Offset: 0.047ppm
CH4 Span Offset: 0.010ppm
Notes: Calibrated on the West Pond tower by Cove and Joe. Last cal is the original done at Licor.
1539 2012-05-02 Sherman Island Gas Analyzer CH4 Installed West Pond Gas Analyzer CH4 Installed at West Pond on the folding boom at the top of the tower.
1538 2012-05-02 Sherman Island Gas Analyzer CH4 Removed West Pond Gas Analyzer CH4 Removed from the Sherman tower for use at West pond. Replaced by LGR (sn 08-114).
645 2012-02-15 Lab none Installed Sherman Island Installed on the Sherman tower and connected directly to the CR1000 by ethernet. Auto washer also installed.