Equipment and Calibration

View History

Equipment Entry

Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-7700
Serial No:TG1-0424
Tag No:NULL
Acquisition Date:2015-09-10
Project:CDFW
Location on 2023-08-08:Lab, NULL
Notes: GHG-1 package comprised of: sn TG1-0424, sn 75H-2669, sn AIU-1570

Documents

Calibration: cal_LI-7700_snTG1-0424.pdf
Manual: Collecting_LI-7700_Ethernet_Data.pdf
Manual: LI-7700_Manual_V2_201103.pdf
Manual: LI-7700_Manual_V3_201312.pdf
PackingSlip: ps_Licor_20150910.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6833 2024-08-28 Lab none Calibration Loaned none Calibration Location: Lab
Last calibration: 2024-02-26
RSSI=75%, Optical RH=4%
Ambient P=100.08kPa, Tair=25.7C
Zero Air at 1400psi, AMP CA03923 3.3865 at 1500
CH4 Zero Offset: .046ppm
CH4 Span Offset: -.0225ppm
Notes: Spray nozzle had been offset so RSSI~20 and Haley was refilling reservoir and manually cleaning every 2 weeks. We replace the spiky hat and spray nozzle after calibration.
6575 2024-02-26 Lab Calibration Lab Calibration Location: Lab
Last calibration: 2023-09-05
RSSI=77%, Optical RH=5%
Ambient P=100.6kPa, Tair=23.0C
Zero Air at 350psi, AMP CA04119 4.223 at 600
CH4 Zero Offset: -.063ppm
CH4 Span Offset: .009ppm
Notes: BWW sensor. After installing it again, I put the spiky hat and spray nozzle on again. The was reservoir was empty; Haley refilled it.
6402 2023-09-05 Lab Calibration Lab Calibration Location: Lab
Last calibration: 2023-01-19
RSSI=80%, Optical RH=23%
Ambient P=100.0kPa, Tair=25.0C
Zero Air at 800psi, AMP CA04119 4.2238 at 850
CH4 Zero Offset: .028ppm
CH4 Span Offset: -.0458ppm
Notes: Calibrated zero on 2023-09-05 and span on 2023-09-07 (P=100.3 kPa, T=24.1C). After calibration, swapped drierite and flushed headspace with zero air until optics RH=10%.
6362 2023-08-08 Gilbert Tract Gas Analyzer CH4 Removed Lab none Removed from Gilbert Tract for calibration. Replaced by sn TG1-0223.
6169 2023-01-19 Gilbert Tract Calibration Gilbert Tract Calibration Location: Gilbert Tract
Last calibration: 2022-06-09
RSSI=86%, Optical RH=23%
Ambient P=102.3kPa, Tair=8.0C
Zero Air at 1350psi, AMP CA04396 1.9317 at 1250
CH4 Zero Offset: .001ppm
CH4 Span Offset: .0017ppm
Notes: I wanted to flush the headspace with zero air, but the tubing did not quite reach. I swapped the desiccant.
5896 2022-06-09 Gilbert Tract Calibration Gilbert Tract Calibration Location: Gilbert Tract
Last calibration: 2021-11-30
RSSI=74%, Optical RH=16%
Ambient P=101kPa, Tair=21.6C
Zero Air at 450psi, AMP CA04119 4.2238 at 1450
CH4 Zero Offset: -.01ppm
CH4 Span Offset: .0698ppm
Notes: Breezy day. Did not change drierite or flush headspace with zero air.
5611 2021-11-30 Gilbert Tract Gas Analyzer CH4 Calibration Gilbert Tract Gas Analyzer CH4 Calibration Location: Gilbert Tract
Last calibration: 2021-06-01
RSSI=85%, Optical RH=22%
Ambient P=102.1kPa, Tair=15.2C
Zero Air at 100psi, AMP CA04396 1.9317 at 1800
CH4 Zero Offset: .032ppm
CH4 Span Offset: -.0197ppm
Notes: Relatively still day with a lot of visitors to the site. The zero air tank pressure may have been >100psi, but the regulator was wonky. Zero was stable for 2min after applying zero. Swapped out the desiccant one for one with new drierite after calibration. I didn't flush out the headspace because I wanted to preserve the zero air for more calibrations that day.
5490 2021-09-02 Sherman Wetland Gas Analyzer CH4 Moved Gilbert Tract Gas Analyzer CH4 Moved from SW to GT with the decomissioning of SW and the beginning of GT
5277 2021-05-26 Sherman Wetland Gas Analyzer CH4 Calibration Sherman Wetland Gas Analyzer CH4 Calibration Location: Sherman Wetland
Last calibration: 2020-11-16
RSSI=80%, Optical RH=15%
Ambient P=101kPa, Tair=27.3C
AMP FB04161 at 100psi, AMP CA04119 4.223 at 2050
CH4 Zero Offset: -0.022ppm
CH4 Span Offset: 0.017ppm
Notes: Breezy day. Zero gas was an AMP cal gas in a baby tank that we borrowed for the day. Not much zero air in the tank but it was enough.
4944 2020-11-12 Sherman Wetland Gas Analyzer CH4 Calibration Sherman Wetland Gas Analyzer CH4 Calibration Location: Sherman Wetland
Last calibration: 2020-05-21
RSSI=82%, Optical RH=12%
Ambient P=101.8kPa, Tair=14.7C
Zero Air at 700psi, AMP CA3923 3.121 at 700
CH4 Zero Offset: -0.084ppm
CH4 Span Offset: 0.002ppm
Notes: Still day
4484 2020-05-21 Sherman Wetland Gas Analyzer CH4 Calibration Sherman Wetland Gas Analyzer CH4 Calibration Location: Sherman Wetland
Last calibration: 2019-10-31
RSSI=84%, Optical RH=22%
Ambient P=101.5kPa, Tair=24.1C
Zero Air at 750psi, AMP CA3923 3.121 at 800
CH4 Zero Offset: 0.006ppm
CH4 Span Offset: -.013ppm
Notes: Breezy day. Swapped optics drierite after calibration and used cal gas to dry head space. New optics RH was 9%.
4130 2019-10-31 Sherman Wetland Gas Analyzer CH4 Calibration Sherman Wetland Gas Analyzer CH4 Calibration Location: Sherman Wetland
Last calibration: 2019-04-11
RSSI=78%, Optical RH=19%
Ambient P=102.4kPa, Tair=18.1C
Zero Air at 1000psi, AMP CA03923 3.121 at 1000
CH4 Zero Offset: 0.005ppm
CH4 Span Offset: 0.005ppm
Notes: Zero had random large spikes that turned out to be a tiny insect crawling around the bottom mirror - removed before span. Otherwise, very tight calibration.
3900 2019-04-11 Sherman Wetland Gas Analyzer CH4 Calibration Sherman Wetland Gas Analyzer CH4 Calibration Location: Sherman Wetland
Last calibration: 2018-10-24
RSSI=77%, Optical RH=13%
Ambient P=102.0kPa, Tair=15.4C
Zero Air M4002 at 1800psi, AMP CA03923 3.121 at 1900
CH4 Zero Offset: 0.051ppm
CH4 Span Offset: 0.024ppm
Notes: Windy day 3m/s to 7m/s
3884 2019-02-12 Sherman Wetland Gas Analyzer CH4 Upgrade Sherman Wetland Gas Analyzer CH4 Upgraded firmware (embedded software) from 1.0.23 to 1.0.27. Licor change log: "Expanded a buffer in the DSP to prevent phase changes in the DSP data caused when the buffer filled up. This fixes the synchronization loss and resultant data loss between the LI-7550 and LI-7700."
3772 2018-10-25 Lab none Installed Sherman Wetland Gas Analyzer CH4 Installed at SW to replace sn TG1-0423, which was brought back to the lab for power board replacement and for calibration. Heater settings changed from Bouldin settings (top: 25%, 2 degC, 18:00-11:00; bottom: 25%, 30%, 18:00-11:00) to Sherman/Twitchell settings (top: 5%, 2 degC, 18:00-9:00; bottom: 5%, 30%, 18:00-9:00).
3756 2018-10-24 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2018-04-13
RSSI=82.9%, Optical RH=9%
Ambient P=100.6kPa, Tair=27.8C
Zero Air at 1500psi, AMP CH4 1.9497 at 1150
CH4 Zero Offset: -0.015ppm
CH4 Span Offset: 0.043ppm
Notes: First calibration after replacing power board - no problems found.
3755 2018-10-22 Lab none Upgrade Lab none I swapped out power board pn 9677-079 per Licor's instructions. According to Licor the original boards Rev 0, could be damaged by exposure to voltage above 24VDC. The new board Rev 0A (V3) should not have that problem.
3748 2018-10-18 Bouldin Alfalfa Gas Analyzer CH4 Removed Lab none Removed from Bouldin Alfalfa for power board upgrade and to swap in for other two 7700s needing upgrade.
3414 2018-04-19 Lab none Installed Bouldin Alfalfa Gas Analyzer CH4 Installed on Bouldin Alfalfa to replace sn TG1-0223 for calibration
3396 2018-04-13 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-10-11
RSSI=75%, Optical RH=4%
Ambient P=101.8kPa, Tair=24.2C
Zero Air at 500psi, LBL Air 1.9497 at 1500
CH4 Zero Offset: .008ppm
CH4 Span Offset: -.0417ppm
Notes: Started "new" ultra pure zero air tank that had 500 psi left in it.
3391 2018-04-05 West Pond Gas Analyzer CH4 Removed Lab none Removed from WP and brought back to the lab for calibration. It was replaced by sn TG1-0224.
3150 2017-10-13 Lab none Installed West Pond Gas Analyzer CH4 Installed on West Pond to replace sn TG1-0223 for calibration
3139 2017-10-11 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-03-22
RSSI=76.7%, Optical RH=4%
Ambient P=100.5kPa, Tair=23.7C
Zero Air at 500psi, AMP CH4 1.9497 at 2000
CH4 Zero Offset: 0.042ppm
CH4 Span Offset: -0.001ppm
Notes:
3110 2017-10-04 East End-Temp3 Gas Analyzer CH4 Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
2883 2017-06-23 Bouldin Alfalfa Gas Analyzer CH4 Installed East End-Temp3 Gas Analyzer CH4 Moved from Bouldin Alfalfa to be installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site.
2676 2017-03-23 Lab none Installed Bouldin Alfalfa Gas Analyzer CH4 Installed at Bouldin Alfalfa to replace TG1-0421.
2660 2017-03-22 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2015-07-23
RSSI=84.6%, Optical RH=26%
Ambient P=100.7kPa, Tair=23.2C
Zero Air at 1700psi, NOAA CH4 1.89919 at 1750
CH4 Zero Offset: 0.05ppm
CH4 Span Offset: -0.004ppm
Notes: Changed dessicant. Previous calibration was at Licor.
2139 2015-12-23 Twitchell Rice-Trailer Removed Lab none Removed with the temporary tower set up near the building at the rice project for the LGR COS experiment.
2115 2015-12-08 Lab none Installed Twitchell Rice-Trailer Installed near the trailer at Twitchell Rice as a temporary site to test a COS analyzer from LGR