Equipment and Calibration

View History

Equipment Entry

Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-7700
Serial No:TG1-0423
Tag No:
Acquisition Date:2015-09-10
Project:CDFW
Location on 2019-02-12:East End,
Notes: GHG-1 package comprised of: sn TG1-0423, sn 75H-2668, sn AIU-1569

Documents

Calibration: cal_LI-7700_snTG1-0423.pdf
Manual: Collecting_LI-7700_Ethernet_Data.pdf
Manual: LI-7700_Manual_V2_201103.pdf
Manual: LI-7700_Manual_V3_201312.pdf
PackingSlip: ps_Licor_20150910.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6809 2024-08-07 East End Calibration East End Calibration Location: East End
Last calibration: 2024-01-11
RSSI=70%, Optical RH=4%
Ambient P=101.1kPa, Tair=35.1C
Zero Air at 1700psi, AMP CA03923 3.3865 at 1850
CH4 Zero Offset: -.028ppm
CH4 Span Offset: -0.0185ppm
Notes: Hot day
6529 2024-01-11 East End Calibration East End Calibration Location: East End
Last calibration: 2023-06-15
RSSI=79%, Optical RH=23%
Ambient P=102.2kPa, Tair=16.0C
Zero Air at 1800psi, AMP CA03923 3.3865 at 1725
CH4 Zero Offset: .015ppm
CH4 Span Offset: .0015ppm
Notes: After calibration, flushed headspace with zero air until optics RH=9% and swapped in a bottle with fresh drierite.
6304 2023-06-15 East End Calibration East End Calibration Location: East End
Last calibration: 2023-01-18
RSSI=82%, Optical RH=16%
Ambient P=100.9kPa, Tair=26.7C
Zero Air at 1150psi, AMP CA04119 4.2238 at 1000
CH4 Zero Offset: -.06ppm
CH4 Span Offset: .0138ppm
Notes: Optics RH=16%, which is pretty good, so we did not flush the headspace or change its desiccant.
6168 2023-01-18 East End Calibration East End Calibration Location: East End
Last calibration: 2022-06-09
RSSI=77%, Optical RH=21%
Ambient P=101.9kPa, Tair=12.6C
Zero Air at 1450psi, AMP CA04396 1.9317 at 1200
CH4 Zero Offset: .025ppm
CH4 Span Offset: .0023ppm
Notes: After calibration, changed desiccant and flushed headspace with zero air until optics RH=14%.
5895 2022-06-09 East End Calibration East End Calibration Location: East End
Last calibration: 2021-12-01
RSSI=70%, Optical RH=17%
Ambient P=101kPa, Tair=29.5C
Zero Air at 350psi, AMP CA04119 4.2238 at 1550
CH4 Zero Offset: .015ppm
CH4 Span Offset: -.0098ppm
Notes: Hot day, some breeze. Did not change drierite or flush headspace with zero air.
5613 2021-12-01 East End Calibration East End Calibration Location: East End
Last calibration: 2021-06-01
RSSI=83%, Optical RH=15%
Ambient P=101.2kPa, Tair=19.2C
Zero Air at 1200psi, AMP CA04396 1.9317 at 1800
CH4 Zero Offset: .007ppm
CH4 Span Offset: .0023ppm
Notes: Relatively still day. Did not change drierite or flush headspace with zero air. Zero air tank pressure uncertain--wonky regulator.
5279 2021-05-26 East End Calibration East End Calibration Location: East End
Last calibration: 2020-11-19
RSSI=80%, Optical RH=13%
Ambient P=101.5kPa, Tair=29.2C
AMP FB04161 at 500psi, AMP CA04119 4.223 at 2050
CH4 Zero Offset: 0ppm
CH4 Span Offset: 0.008ppm
Notes: Relatively still day. Zero gas was an AMP cal gas in a baby tank that we borrowed for the day. Zero was right on.
4981 2020-11-19 East End Calibration East End Calibration Location: East End
Last calibration: 2020-05-19
RSSI=77%, Optical RH=12%
Ambient P=102.4kPa, Tair=18C
Zero Air at 500psi, AMP CA3923 3.121 at 500
CH4 Zero Offset: -0.05ppm
CH4 Span Offset: -0.002ppm
Notes: Breezy day
4454 2020-05-19 East End Gas Analyzer CH4 Calibration East End Gas Analyzer CH4 Calibration Location: East End
Last calibration: 2019-10-22
RSSI=83%, Optical RH=20%
Ambient P=101.5kPa, Tair=22.5C
Zero Air at 900psi, AMP CA3923 3.121 at 850
CH4 Zero Offset: 0.01ppm
CH4 Span Offset: 0.006ppm
Notes: Swapped optics desiccant (indicating drierite) after calibration. Used zero air cal gas to dry out head space. New RH ~7%.
4117 2019-10-22 East End Gas Analyzer CH4 Calibration East End Gas Analyzer CH4 Calibration Location: East End
Last calibration: 2019-04-11
RSSI=83%, Optical RH=17%
Ambient P=102.0kPa, Tair=24.6C
Zero Air at 1300psi, AMP CA03923 3.121 at 1400
CH4 Zero Offset: 0.023ppm
CH4 Span Offset: -0.001ppm
Notes:
3899 2019-04-11 East End Gas Analyzer CH4 Calibration East End Gas Analyzer CH4 Calibration Location: East End
Last calibration: 2018-10-30
RSSI=73%, Optical RH=11%
Ambient P=101.9kPa, Tair=16.5C
Zero Air M4002 at 1700psi, AMP CA03923 3.121 at 1800
CH4 Zero Offset: 0.036ppm
CH4 Span Offset: 0.013ppm
Notes: Windy day
3883 2019-02-12 East End Gas Analyzer CH4 Upgrade East End Gas Analyzer CH4 Upgraded firmware (embedded software) from 1.0.23 to 1.0.27. Licor change log: "Expanded a buffer in the DSP to prevent phase changes in the DSP data caused when the buffer filled up. This fixes the synchronization loss and resultant data loss between the LI-7550 and LI-7700."
3787 2018-11-01 Lab none Installed East End Gas Analyzer CH4 Installed at EE to replace sn TG1-0421, which was brought back to the lab for a power board replacement and calibration.
3781 2018-10-30 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2018-05-08
RSSI=80%, Optical RH=8%
Ambient P=100.8kPa, Tair=27.97C
Zero Air at 1500psi, AMP Air 1.9497 at 1000
CH4 Zero Offset: -0.015ppm
CH4 Span Offset: -0.0027ppm
Notes: First calibration after power board was replaced
3780 2018-10-29 Lab none Upgrade Lab none I swapped out power board pn 9677-073G per Licor's instructions. The new power board is pn 9977-079. According to Licor the original boards Rev 0, could be damaged by exposure to voltage above 24VDC. The new board Rev 0A (V3) should not have that problem.

Note: This instrument was replaced with 2 original screws and 2 long screws. The long screws were used to help re-assemble the instrument, and I forgot to swap them for the original, shorter, screws.
3771 2018-10-25 Sherman Wetland Gas Analyzer CH4 Removed Lab Gas Analyzer CH4 Removed from SW and brought back to the lab for a power board swap and calibration. It was replaced by sn TG1-0424.
3453 2018-05-10 Lab none Installed Sherman Wetland Data Logger Met Installed on Sherman Wetland to replace sn TG1-0421 for calibration
3449 2018-05-08 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-10-23
RSSI=73%, Optical RH=4%
Ambient P=100.8kPa, Tair=23.9C
Zero Air at 400psi, LBL Air 1.9497 at 1400
CH4 Zero Offset: 0.016ppm
CH4 Span Offset: 0.0067ppm
Notes:
3447 2018-05-03 Mayberry Gas Analyzer CH4 Removed Lab Gas Analyzer CH4 Removed from MB and brought back to the lab for calibration. It was replaced by TG1-0423.
3181 2017-10-26 Lab none Installed Mayberry Gas Analyzer CH4 Installed at MB to replace sn TG1-0357
3156 2017-10-23 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-05-31
RSSI=74.6%, Optical RH=3%
Ambient P=101.3kPa, Tair=27.9C
Zero Air at 425psi, AMP Air 1.9497 at 1850
CH4 Zero Offset: -0.055ppm
CH4 Span Offset: 0.0197ppm
Notes:
3167 2017-10-19 East End-Temp1 Gas Analyzer CH4 Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
2811 2017-06-01 Lab none Installed East End-Temp1 Gas Analyzer CH4 Installed at open water temporary tower in East End wetland
2794 2017-05-31 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2016-10-11
RSSI=80.7%, Optical RH=10%
Ambient P=100.3kPa, Tair=20.4C
Zero Air at 1500psi, NOAA Air 1.89919 at 1600
CH4 Zero Offset: 0.017ppm
CH4 Span Offset: -0.021ppm
Notes: Replaced desiccant and flushed cavity with zero air. Optics RH was 26% before replacement and 10% after replacement.
2673 2017-03-23 Sherman Wetland Gas Analyzer CH4 Removed Lab none Removed from Sherman Wetland and brought back to the lab for calibration. It was replaced by TG1-0224.
2493 2016-10-11 Sherman Wetland Gas Analyzer CH4 Calibration Sherman Wetland Gas Analyzer CH4 Calibration Location: Sherman Wetland
Last calibration: 2016-05-20
RSSI=78.7%, Optical RH=23%
Ambient P=101.6kPa, Tair=16.4C
zero air at 850psi, AMP CH4 4.030 at 1900
CH4 Zero Offset: 0.026ppm
CH4 Span Offset: 0.023ppm
Notes: Windy.
2285 2016-05-24 Lab none Installed Sherman Wetland Gas Analyzer CH4 Installed on the Sherman Wetland tower at the initial set up of the site.
2276 2016-05-20 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2015-07-23 (Factory)
RSSI=79.4%, Optical RH=18%
Ambient P=100.08kPa, Tair=22.6C
Zero Air at 500psi, 17-07-2013 CH4 Tank 4.995 at 500
CH4 Zero Offset: 0.03ppm
CH4 Span Offset: 0.006ppm
Notes: Uneventful calibration. Previous calibration was factory.