Equipment and Calibration History

View Equipment |

View history entry 2794 for Licor LI-7700 sn: TG1-0423

Date: 2017-05-31
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2016-10-11 RSSI=80.7%, Optical RH=10% Ambient P=100.3kPa, Tair=20.4C Zero Air at 1500psi, NOAA Air 1.89919 at 1600 CH4 Zero Offset: 0.017ppm CH4 Span Offset: -0.021ppm Notes: Replaced desiccant and flushed cavity with zero air. Optics RH was 26% before replacement and 10% after replacement.
Entered by: