Equipment and Calibration History

View Equipment |

View history entry 6304 for Licor LI-7700 sn: TG1-0423

Date: 2023-06-15
From: East End -
To: East End -
Title: Calibration
Notes: Calibration Location: East End Last calibration: 2023-01-18 RSSI=82%, Optical RH=16% Ambient P=100.9kPa, Tair=26.7C Zero Air at 1150psi, AMP CA04119 4.2238 at 1000 CH4 Zero Offset: -.06ppm CH4 Span Offset: .0138ppm Notes: Optics RH=16%, which is pretty good, so we did not flush the headspace or change its desiccant.
Entered by:
Field notes