Rec No. | Date | Loc From | Title | Loc To | Notes |
5560 |
2021-11-08 |
Lab none |
Returned |
Returned none |
Returned to Sebastien of AMP in exchange for a freshly calibrated reference PAR sn 212225 |
5570 |
2021-11-05 |
Cal Kit PAR Sensor Incoming |
Removed |
Lab none |
Removed from BA tower at the conclusion of the rad calibration and brought back to the lab. This sensor will go back to AMP. |
5536 |
2021-10-21 |
Cal Kit PAR Sensor Reflected |
Moved |
Cal Kit PAR Sensor Incoming |
Moved from GT Reflected to BA Incoming as part of the radiometer cal kit to begin the calibration at BA |
5718 |
2021-10-07 |
Cal Kit PAR Sensor Incoming |
Moved |
Cal Kit PAR Sensor Reflected |
Moved at GT from incoming to reflected to continue inter-comparison |
5458 |
2021-09-15 |
Cal Kit PAR Sensor Reflected |
Moved |
Cal Kit PAR Sensor Incoming |
Moved from MB to the new Gilbert Tract GT site to continue the radiometer intercomparison there. |
5426 |
2021-08-17 |
Cal Kit PAR Sensor Reflected |
Moved |
Cal Kit PAR Sensor Incoming |
Moved from EE to MB after completion of radiometer intercomparison at EE |
5390 |
2021-07-21 |
Cal Kit PAR Sensor Reflected |
Moved |
Cal Kit PAR Sensor Incoming |
Moved from WP to EE at the conclusion of the radiometer inter-comparison at WP |
5347 |
2021-06-22 |
Cal Kit PAR Sensor Reflected |
Moved |
Cal Kit PAR Sensor Incoming |
Moved from HS to WP as part of the radiometer calibration kit at the conclusion of the 2-week intercomparison at HS. |
5344 |
2021-06-16 |
Cal Kit PAR Sensor Incoming |
Moved |
Cal Kit PAR Sensor Reflected |
Moved at HS from the incoming position to the reflected position |
5322 |
2021-06-09 |
Cal Kit PAR Sensor Reflected |
Moved |
Cal Kit PAR Sensor Incoming |
Moved from BC to HS as part of rad cal kit |
5294 |
2021-06-02 |
Cal Kit PAR Sensor Incoming |
Moved |
Cal Kit PAR Sensor Reflected |
Moved from the incoming position to the reflected position for continuing inter-comparison. |
5309 |
2021-05-27 |
Lab none |
Installed |
Cal Kit PAR Sensor Incoming |
Installed at BC as part of the cal kit in the incoming position |
5150 |
2021-03-02 |
Cal Kit PAR Sensor Reflected |
Removed |
Lab none |
Removed from TZ Tower top with the calibration kit |
5305 |
2021-02-10 |
Cal Kit PAR Sensor Incoming |
Moved |
Cal Kit PAR Sensor Reflected |
Moved at TZ from the incoming position to the reflected position to continue inter-comparison. |
5059 |
2021-01-21 |
Lab none |
Installed |
Cal Kit PAR Sensor Incoming |
Installed at TZ in the incoming position as part of the inter-comparison kit to test PAR, AirT and RH on the top of the Tower. |
5307 |
2021-01-17 |
Cal Kit PAR Sensor Reflected |
Removed |
Lab none |
Removed from WP with the cal kit |
5306 |
2020-12-16 |
Cal Kit PAR Sensor Incoming |
Moved |
Cal Kit PAR Sensor Reflected |
Moved from the incoming position to the reflected position to continue the inter-comparison at WP |
4976 |
2020-11-19 |
Lab none |
Installed |
Cal Kit PAR Sensor Incoming |
This is the Ameriflux standard installed at WP for cross comparison with existing sensors |
5310 |
2020-06-04 |
Factory none |
Calibration |
Factory none |
y=mx+b 149.48 0; The date for this entry is made up but allows this calibration to be used as needed. After discrepancies with an inter-comparison on Tonzi Tower in the spring to 2021 and a conversation with Sebastien Biraud on 2021-05-05 we learned that the OSU calibration for this senor was no good and the subsequent calibration that this entry represents should be used instead. The OSU calibration uses a lamp and fails for certain PAR sensors for unknown reasons. This calibration is done at LBL outside under ambient conditions cross compared with ten other PAR sensors. |
4942 |
2020-01-10 |
Factory none |
Calibration |
Factory none |
y=mx+b 131.80 0 Calibrated by Chad Hanson at OSU as part of the Ameriflux standard program 7.59 uV/umol/m2/s |