Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-820
Serial No:CGA-556
Tag No:
Acquisition Date:0000-00-00
Project:DOE-TCP
Location on 2018-05-23:Lab,
Notes:

Documents

Manual: LI-820_840_Embedded_CGA103.pdf
Manual: LI-820_Manual.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
3699 2018-05-23 Tonzi Removed Lab none Removed from the Tonzi Tower with the old CO2 profile system. New Vaisala profile system is running
3336 2018-02-21 Lab none Installed Tonzi Installed in the old CO2 profile system on the Tower for comparison of the new GMP343 profile.
3245 2017-12-14 Tonzi Removed Lab none Removed from the Tower co2 profile for cleaning and calibration. The rest of the profile equipment was turned off
3211 2017-11-29 Lab none Installed Tonzi Installed in the Tower CO2 profile system after lab cleaning and calibration
2500 2016-10-13 Tonzi-Tower Calibration Tonzi-Tower none Last calibration was 2016-09-13 Cell T = 52.2, Pressure = 96.2, N2 at 1150psi, CO2 406ppm at 1800
zero offset = -1.5ppm
Span offset = -7.5ppm
Zero 0.9429 -> 0.9424
Span 0.8368 -> 0.8492
2365 2016-07-07 Tonzi Calibration Tonzi Last calibration was 2016-06-16 Cell T = 52.2, Pressure = 91.0
Zero Air offset = -2.0ppm
Span CO2 405ppm offset = 5.0ppm
Zero 0.9472 -> 0.9467
Span 0.8496 -> 0.8369
2322 2016-06-16 Tonzi-Tower Calibration Tonzi-Tower Last calibration was 2016-05-26 Cell T = 52.2, Pressure = 92.0
Zero Air offset = 5.0ppm
Span CO2 405ppm offset = 0.0ppm
Zero 0.9461 -> 0.9472
Span 0.8499 -> 0.8496
2282 2016-05-26 Tonzi-Tower Calibration Tonzi-Tower Last calibration was 2016-02-09
Cell T = 52.2, Pressure = 92.0
Zero Air offset = 16.4ppm
Span CO2 405ppm offset = -19.4ppm
Zero 0.9423 -> 0.9461
Span 0.8220 -> 0.8499

Changed Pressure output range from 1.33331998E3 to 115kPa
2183 2016-02-23 Lab none Installed Tonzi-Tower Reinstalled in the Tower CO2 profiles system, but the pump was not working and was removed. The IRGA was left unpowered.
2204 2016-02-09 Lab none Calibration Lab none Last calibration was 09-17-2015
Cleaned cell before calibration
Cell T = 52.2, Pressure = 100.9
Zero Air offset = -20.0 railed
Span God CO2 421.22ppm offset = 11.28ppm
Zero 0.9477 -> 0.9423
Span 0.8394 -> 0.8220
1735 2015-03-02 Tonzi-Tower Calibration Tonzi-Tower Calibrated on tower. 52.1C, 93.7kPa, last cal 2015-02-04. N2 1200psi, CO2 405ppm 1750psi.
zero 0.6ppm, 0.9040
span -0.8ppm, 0.8395
1678 2015-01-21 Lab none Installed Tonzi-Tower Reinstalled in the CO2 profile system after swap of cell with LI-800 in the lab.
1563 2014-09-11 Lab none Installed Tonzi-Tower Reinstalled in the CO2 profile system on the Tonzi Tower. The IRGA seemed to be working but something wrong with the plumbing made it impossible to calibrate on the tower.
1562 2014-09-09 Lab none Maintenance Lab none I cleaned the cell and replaced the internal inlet tubing. The tubing had some dirty looking deposit inside it. I calibrated the IRGA: P = 100.3kPa, CellT = 52.2C, Zero = 0.9656, Span = 0.8022
1564 2014-08-21 Tonzi-Tower Removed Lab none Removed from the tower due to dirty cell that could not be cleaned well enough in the field.