Equipment and Calibration History

View Equipment |

View history entry 2322 for Licor LI-820 sn: CGA-556

Date: 2016-06-16
From: Tonzi-Tower -
To: Tonzi-Tower -
Title: Calibration
Notes: Last calibration was 2016-05-26 Cell T = 52.2, Pressure = 92.0 Zero Air offset = 5.0ppm Span CO2 405ppm offset = 0.0ppm Zero 0.9461 -> 0.9472 Span 0.8499 -> 0.8496
Entered by: Joe Verfaillie