Other sites visited today: Twitchell Alfalfa | East End | Mayberry | Twitchell Rice

West Pond: 2015-10-29 12:20:00 (DOY 302)
Author: Joe Verfaillie
Others: Sara

Summary: Standard data collection, adjust 7550 clock and band width, camera clock needs battery.

West Pond
2015-10-29

I arrived about 12:20 with Sara. It was mild and windy with a clear blue sky. The inlet was running and water level was normal.

The 7500 read: 388.4ppm CO2, 335.7mmol/m3, 88.0SS, 23.27C, 101.15kPa
The 7700 read: 1.93ppm CH4, 39.5RSSI – 72.7 after cleaning
Optics RH was 24%

I checked the 7550 settings. The clock was off. At 11:27 I reset it from 11:21.
I changed the band width from 5Hz to 10Hz since we are collecting at 20Hz.

0 photos found

Graphs display:
8 sets found

Can't check WP_DO
1 WP_DO graphs found

WestPond DO

14 WP_met graphs found

Battery

Temperatures

Solar Radiation

RH and Pressure

Water Level and Precipitation

Conductivity

Temperatures Set A

Temperatures Set B

Temperature Set C

Water CO2 and Temp

West Pond Water CO2

WP miniDOT Dissolved Oxygen

SRS NDVI Channels

NDVI

5 WP_flux graphs found

Winds

LI7500 CO2 and H2O

LI7700 CH4 and RSSI

West Pond Flux Data Counts

Air Temperatures

Can't check MBWPEESWGT_mixed
Can't check MBEEWPSWEPGT_cam
Can't check MBSWEEWPBCBA
1 WP_processed graphs found

WP CO2 Fluxes

No data found for WP_arable.

 
 

Login
Select Notes:

This material is based upon work supported by the National Science Foundation and US Department of Energy. Any opinions, findings, conclusions, or recommendations expressed in the material are those of the author(s) and do not necessarily reflect the views of the supporters.