Equipment and Calibration

View History

Equipment Entry

Instrument:Water Level
Manufacture:Campbell Sci.
Model No:CS450
Serial No:20010026
Tag No:NULL
Acquisition Date:2009-05-05
Project:DWR-Rice
Location on 2024-04-04:Mayberry, NULL
Notes: NULL

Documents

No documents found

Report: Water Level sensor calibration with staff gauge - 2018-04-24
Report: GT Water Level Comparison and Correction - 2021-09-09

History

Rec No.DateLoc FromTitleLoc ToNotes
6644 2024-04-04 Lab none Installed Mayberry Water Level Installed at MB off of the northeast corner of the scaffolding. Water level data from the HYDROS21 CTD sn H21G200002263 hasn't worked since Feb, but cond/temp seem ok.
6636 2024-03-25 West Pond Water Level Removed Lab none Removed from WP and brought back to the lab. This sensor is needed at Mayberry, so we are slowly decommissioning West Pond.
6364 2023-08-08 Lab none Installed West Pond Water Level Installed at West Pond to replace sn 20010769, which died after 11 years of service. The old sensor is still installed, but not wired to the datalogger. The new sensor is installed temporarily hangingin the water by the staff gauge. We will install a well when the water level drops.
5671 2022-02-02 Gilbert Tract Water Level Removed Lab none Removed from Gilbert Tract as its measurements are now redundant with the HYDROS21 water level sensor.
5515 2021-09-02 Sherman Wetland Water Level Moved Gilbert Tract Water Level Moved from SW to GT with the decomissioning of SW and the beginning of GT. Not installed until 09/09
5483 2020-08-28 Lab none Installed Sherman Wetland Water Level Installed after repair at Campbell -90cm from top of pipe -45cm from ground surface.
4667 2020-08-20 Factory none Received Lab none Report from CSI: Sensor communicates ok and draws normal current, it does read OOT so it will need a new cal and the rotted 20' cable replaced.
Repair: Cleaned up the sensor element, updated OS, calibrated the sensor and had a new 20'cable made.
4658 2020-07-17 Lab none Shipped Factory none Shipped back to Campbell for repair on RMA # 32203
4657 2020-07-09 Sherman Wetland Water Level Removed Lab none Removed from SW and brought back to the lab because this sensor was causing the met system to short
2930 2017-06-23 Lab none Installed Sherman Wetland Water Level Installed at Sherman Wetland tower
1846 2015-05-12 Sherman Island Removed Lab none Removed with the decommissioning of the Sherman Cow Pasture site.
1071 2013-09-05 Lab none Installed Sherman Island Installed at Sherman Island in newly dig well with perforated PVC tube just south of the computer box. The water pressure sensor is installed 124 cm below the top of the PVC pipe, and the top of the PVC pipe extends 13 cm above the ground surface. At the time of install, the water table was measured at 82 cm below ground and the water pressure reading was 0.43.
880 2013-01-03 Lab none Installed DWR Corn Installed in the original well hole in the corn field. See fieldnotes of details.
879 2012-12-18 Lab none Repair Lab none I pulled apart the existing splice in the cable, cut a small out and made a new splice. After wards I tested it in the lab and it seemed to function well.
858 2012-11-29 Mayberry Water Level Removed Lab none This sensor was removed from the Tule tower due to bad readings. Replaced by brand new CS451 sn20010046 sensor.
817 2012-10-25 Twitchell Rice Water Level Moved Mayberry Water Level Removed from the Twitchell tower since the well was dry and moved to the Mayberry Tule tower to replace the water pressure sensor that failed there.
393 2011-07-14 Twitchell Rice Water Level Installed Lab none Reinstalled in the field after repairs to the cable where otters had chewed on it. I spliced the wires and the tube that provides the atmospheric pressure reference.
394 2011-07-07 Twitchell Rice Water Level Removed Lab none This senor was removed from the Twitchell tower after it was discovered that otters had chewed the cable.
1323 2009-05-21 Lab none Installed Twitchell Rice Water Level Installed at the Twitchell Rice site