Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: cal_Apogee_SI-131-SS_sn7385_20180409.pdfPackingSlip: ps_Apogee_20180413.pdf
Report: Testing of inexpensive DFRobot IRT sensors - 2019-05-10
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6979 | 2024-10-29 | Lab none | Installed | Tonzi Infrared Temperature StemFlux East | Installed east of the Floor tower to measure tree trunk temperatures for Kuno's stem flux experiment |
6948 | 2024-10-23 | Gilbert Tract | Removed | Lab none | Removed from "Northside" station in Gilbert Tract at the conclusion of Robert's thermocouple experiment |
6857 | 2024-09-19 | Lab none | Installed | Gilbert Tract none | Installed at the Gilbert Tract "northside" station for Robert's temperature campaign |
6333 | 2022-11-17 | Tonzi Infrared Temperature Ref Panel A | Removed | Lab none | Removed from Tonzi at the conclusion of the IRT calibration along with CR23X sn 2301 |
6040 | 2022-10-20 | Lab none | Installed | Tonzi Infrared Temperature Ref Panel A | Install at the Tonzi cal panel for calibration. But I damaged the data logger so not collecting data until next time. |
6036 | 2022-10-19 | Experiment none | Removed | Lab none | I removed this sensor from the bike kit to calibrate it against the cal panel at Tonzi. |
5935 | 2022-08-10 | Lab none | Installed | Experiment | Installed on the back of Robert's e-bike for his urban heat island transect in Oakley |
5891 | 2022-06-07 | Tonzi Infrared Temperature Grass B | Removed | Lab none | Disconnected and packed up on 05/26 but brought back to the lab today - no longer needed |
5301 | 2021-06-02 | Tonzi Infrared Temperature Ref Panel B | Moved | Tonzi Infrared Temperature Grass B | Moved from pointing at the cal panel to pointing at the grass. It was moved to the cal panel to replace the IRT there for cleaning. I realized later that the grass is probably more important than cal panel but there should be a mini inter-comparison now. |
5257 | 2021-05-04 | Lab none | Installed | Tonzi Infrared Temperature Grass B | Reinstalled after cows chewed the cable and the first attempt at a splice had a bad solder joint. |
5251 | 2021-04-29 | Lab none | Repair | Lab none | A few weeks ago I spliced a new wire on to the circular connector at the end of the cable where a cow chewed it off. I tried to install it in the field on 2021-04-01 and thought I was missing a resistor. On 2021-04-18 I was sure that it did not need a resistor but it still didn't work. I tested the sensor on another cable in the field and it was fine so my splice must be bad. Today I found the the black wire was not spliced well. I fixed the splice and the senor tested okay in the lab. |
5221 | 2021-03-17 | Tonzi Infrared Temperature Grass B | Removed | Lab none | A cow ate the wire from the sensor to the data logger box. Removed for repair in the lab. |
3992 | 2019-05-30 | Lab none | Installed | Tonzi | Installed near the Floor solar panels looking at the grass to the north. Footprint of about 1.23m x 2.5m diameter |
3932 | 2019-04-16 | Tonzi | Removed | Lab none | Removed with the wooden cal panel for the FLIR because cows chewed the TC wires and the new cal panel is better |
3477 | 2018-06-01 | Lab none | Installed | Tonzi | Installed near the Floor Tower on a CR10X to give temperatures of a black panel as a reference for the thermal camera. |