Rec No. | Date | Loc From | Title | Loc To | Notes |
6979 |
2024-10-29 |
Lab none |
Installed |
Tonzi Infrared Temperature StemFlux East |
Installed east of the Floor tower to measure tree trunk temperatures for Kuno's stem flux experiment |
6948 |
2024-10-23 |
Gilbert Tract |
Removed |
Lab none |
Removed from "Northside" station in Gilbert Tract at the conclusion of Robert's thermocouple experiment |
6857 |
2024-09-19 |
Lab none |
Installed |
Gilbert Tract none |
Installed at the Gilbert Tract "northside" station for Robert's temperature campaign |
6333 |
2022-11-17 |
Tonzi Infrared Temperature Ref Panel A |
Removed |
Lab none |
Removed from Tonzi at the conclusion of the IRT calibration along with CR23X sn 2301 |
6040 |
2022-10-20 |
Lab none |
Installed |
Tonzi Infrared Temperature Ref Panel A |
Install at the Tonzi cal panel for calibration. But I damaged the data logger so not collecting data until next time. |
6036 |
2022-10-19 |
Experiment none |
Removed |
Lab none |
I removed this sensor from the bike kit to calibrate it against the cal panel at Tonzi. |
5935 |
2022-08-10 |
Lab none |
Installed |
Experiment |
Installed on the back of Robert's e-bike for his urban heat island transect in Oakley |
5891 |
2022-06-07 |
Tonzi Infrared Temperature Grass B |
Removed |
Lab none |
Disconnected and packed up on 05/26 but brought back to the lab today - no longer needed |
5301 |
2021-06-02 |
Tonzi Infrared Temperature Ref Panel B |
Moved |
Tonzi Infrared Temperature Grass B |
Moved from pointing at the cal panel to pointing at the grass. It was moved to the cal panel to replace the IRT there for cleaning. I realized later that the grass is probably more important than cal panel but there should be a mini inter-comparison now. |
5257 |
2021-05-04 |
Lab none |
Installed |
Tonzi Infrared Temperature Grass B |
Reinstalled after cows chewed the cable and the first attempt at a splice had a bad solder joint. |
5251 |
2021-04-29 |
Lab none |
Repair |
Lab none |
A few weeks ago I spliced a new wire on to the circular connector at the end of the cable where a cow chewed it off. I tried to install it in the field on 2021-04-01 and thought I was missing a resistor. On 2021-04-18 I was sure that it did not need a resistor but it still didn't work. I tested the sensor on another cable in the field and it was fine so my splice must be bad. Today I found the the black wire was not spliced well. I fixed the splice and the senor tested okay in the lab. |
5221 |
2021-03-17 |
Tonzi Infrared Temperature Grass B |
Removed |
Lab none |
A cow ate the wire from the sensor to the data logger box. Removed for repair in the lab. |
3992 |
2019-05-30 |
Lab none |
Installed |
Tonzi |
Installed near the Floor solar panels looking at the grass to the north. Footprint of about 1.23m x 2.5m diameter |
3932 |
2019-04-16 |
Tonzi |
Removed |
Lab none |
Removed with the wooden cal panel for the FLIR because cows chewed the TC wires and the new cal panel is better |
3477 |
2018-06-01 |
Lab none |
Installed |
Tonzi |
Installed near the Floor Tower on a CR10X to give temperatures of a black panel as a reference for the thermal camera. |