Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP222
Serial No:Z2720012
Tag No:
Acquisition Date:0000-00-00
Project:DOE-TCP
Location on 2017-08-25:Lab, -16cm
Notes: 0-1% range

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
3021 2017-08-25 Tonzi Notes Lab none Removed CO2 probe from Tonzi Tree2 profile at -2 cm. It was sn Z2720014, not sn Z2720012. Not sure where sn Z2720012 is.
2194 2016-02-23 Lab none Installed Tonzi CO2 Probe Tree2 Profile -0.02 Installed in the Tonzi Tree 2 profile at -2cm after lab calibration.
2164 2016-02-04 Tonzi CO2 Probe Tree2 Profile -0.04 Removed Lab none Removed from Tonzi Tree 2 profile at -4cm for calibration
2077 2014-12-09 Tonzi CO2 Probe Tree1 Profile -0.04 Removed Lab none this probe was at Tonzi Tree2 profile at -4cm, notes say it was swapped but no serial number of the new probe
2058 2014-03-27 Tonzi CO2 Probe Open Profile -0.02 Moved Tonzi CO2 Probe Tree2 Profile -0.04 This probe in the Tonzi Open profile was moved from the -2cm location to the Tree2 profile -4cm
1990 2014-01-29 Lab none Installed Tonzi CO2 Probe Open Profile -0.02 Installed at Tonzi in the Open profile at -2cm with new U-shaped tops on the PVC enclosures
1431 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*2.256 + 934.1; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.
1181 2014-01-24 Tonzi CO2 Probe Open Profile -0.02 Removed Lab none Removed from Tonzi open at -2cm for lab calibration
1133 2012-03-02 Lab none Installed Tonzi CO2 Probe Tree2 Profile -0.02 Installed at Tonzi in the Tree2 set at -2cm
4639 2012-02-16 Lab none Installed Tonzi CO2 Probe Open Profile -0.02 Installed at Tonzi in the Open set at -2cm. This was supposed to be sn X3830001 but that probe was not working and so was returned to the lab.
1408 2012-02-06 Lab none Calibration Lab none Trueppm = Probeppm*1.029 + 14.22; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan.
1371 2010-03-02 Lab none Calibration Lab none Trueppm = Probeppm*1.083 + -6.793; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, 597ppm, and 2238ppm. Ta=22.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Jaclyn.
4638 2010-01-08 Tonzi CO2 Probe Open Profile -0.16 Location Tonzi CO2 Probe Open Profile -0.16 Pulled out for ambient air reading and put back.
4548 2008-03-06 Lab none Installed Tonzi CO2 Probe Tree1 Profile -0.16 Installed in the Wast under tree profile at -16cm. mV and ppm readings in air were recorded before installation.
2027 2005-02-17 Tonzi CO2 Probe Open Profile -0.08 Location Tonzi CO2 Probe Open Profile -0.08 Location noted at Tonzi in Open (East) profile at -8cm