Rec No. | Date | Loc From | Title | Loc To | Notes |
3021 |
2017-08-25 |
Tonzi |
Notes |
Lab none |
Removed CO2 probe from Tonzi Tree2 profile at -2 cm. It was sn Z2720014, not sn Z2720012. Not sure where sn Z2720012 is. |
2194 |
2016-02-23 |
Lab none |
Installed |
Tonzi CO2 Probe Tree2 Profile -0.02 |
Installed in the Tonzi Tree 2 profile at -2cm after lab calibration. |
2164 |
2016-02-04 |
Tonzi CO2 Probe Tree2 Profile -0.04 |
Removed |
Lab none |
Removed from Tonzi Tree 2 profile at -4cm for calibration |
2077 |
2014-12-09 |
Tonzi CO2 Probe Tree1 Profile -0.04 |
Removed |
Lab none |
this probe was at Tonzi Tree2 profile at -4cm, notes say it was swapped but no serial number of the new probe |
2058 |
2014-03-27 |
Tonzi CO2 Probe Open Profile -0.02 |
Moved |
Tonzi CO2 Probe Tree2 Profile -0.04 |
This probe in the Tonzi Open profile was moved from the -2cm location to the Tree2 profile -4cm |
1990 |
2014-01-29 |
Lab none |
Installed |
Tonzi CO2 Probe Open Profile -0.02 |
Installed at Tonzi in the Open profile at -2cm with new U-shaped tops on the PVC enclosures |
1431 |
2014-01-27 |
Lab none |
Calibration |
Lab none |
Trueppm = Probeppm*2.256 + 934.1; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
1181 |
2014-01-24 |
Tonzi CO2 Probe Open Profile -0.02 |
Removed |
Lab none |
Removed from Tonzi open at -2cm for lab calibration |
1133 |
2012-03-02 |
Lab none |
Installed |
Tonzi CO2 Probe Tree2 Profile -0.02 |
Installed at Tonzi in the Tree2 set at -2cm |
4639 |
2012-02-16 |
Lab none |
Installed |
Tonzi CO2 Probe Open Profile -0.02 |
Installed at Tonzi in the Open set at -2cm. This was supposed to be sn X3830001 but that probe was not working and so was returned to the lab. |
1408 |
2012-02-06 |
Lab none |
Calibration |
Lab none |
Trueppm = Probeppm*1.029 + 14.22; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
1371 |
2010-03-02 |
Lab none |
Calibration |
Lab none |
Trueppm = Probeppm*1.083 + -6.793; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, 597ppm, and 2238ppm. Ta=22.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Jaclyn. |
4638 |
2010-01-08 |
Tonzi CO2 Probe Open Profile -0.16 |
Location |
Tonzi CO2 Probe Open Profile -0.16 |
Pulled out for ambient air reading and put back. |
4548 |
2008-03-06 |
Lab none |
Installed |
Tonzi CO2 Probe Tree1 Profile -0.16 |
Installed in the Wast under tree profile at -16cm. mV and ppm readings in air were recorded before installation. |
2027 |
2005-02-17 |
Tonzi CO2 Probe Open Profile -0.08 |
Location |
Tonzi CO2 Probe Open Profile -0.08 |
Location noted at Tonzi in Open (East) profile at -8cm |