Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP222
Serial No:Z2720014
Tag No:
Acquisition Date:2004-06-29
Project:DOE-TCP
Location on 2021-08-11:Lab, -8cm
Notes: 0-1% range

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
5411 2021-08-11 Tonzi-SoilCO2 CO2 Probe Tree2 Profile -0.02 Removed Lab none Removed from -2cm in the soil co2 profile near LiuKang's tree. This profile has not been running for a while due to a break in the power line. This style of measurements are being replaced by forced diffusion chambers.
3062 2017-09-14 Lab none Installed Tonzi CO2 Probe Tree2 Profile -0.02 Installed at Tonzi in the Tree 2 profile at -2cm after lab calibration. This is a new installation with new holes dug.
3019 2017-08-25 Tonzi CO2 Probe Tree2 Profile -0.02 Removed Lab none Removed from Tonzi Tree 2 profile at -2 cm position for calibration
2163 2016-02-04 Tonzi CO2 Probe Tree2 Profile -0.02 Removed Lab none Removed from Tonzi Tree 2 profile at -2cm for calibration
1598 2014-10-09 Twitchell Alfalfa Removed Lab none Removed from the FD ambient chamber due to high readings. Replaced by sn Z2720011.
1525 2014-08-27 Lab none Installed Twitchell Alfalfa Installed in Forced Diffusion probe 1 to replace sn Z2720011 due to zero reading
1456 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*1.248 + 0.9176; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
1455 2014-04-08 Lab none Calibration Lab none Trueppm = Probeppm*1.027 + 0.2028; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.1C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
1246 2014-03-13 Sherman Island Removed Lab none Removed from the Sherman profile B at -2cm after failing field test at atmospheric levels.
1240 2014-01-29 Lab none Installed Sherman Island Reinstalled after lab calibration Set A at -2cm
1436 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*1.076 + -9.057; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.
1183 2014-01-24 Tonzi CO2 Probe Open Profile -0.08 Removed Lab none Removed from Tonzi open at -8cm for lab calibration
1137 2012-02-16 Lab none Installed Tonzi CO2 Probe Open Profile -0.08 Installed at Tonzi in the Open set at -8cm
1366 2010-03-01 Lab none Calibration Lab none Trueppm = Probeppm*1.04 + -83.07; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, 597ppm, and 2238ppm. Ta=20.2C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Jaclyn.
4633 2010-01-08 Tonzi CO2 Probe Tree1 Profile -0.08 Location Tonzi CO2 Probe Tree1 Profile -0.08 Pulled out for ambient air reading and put back.
4543 2008-03-06 Lab none Installed Tonzi CO2 Probe Open Profile -0.08 Installed in the East open grass profile at -8cm. mV and ppm readings in air were recorded before installation.
2025 2005-02-17 Tonzi CO2 Probe Open Profile -0.02 Location Tonzi CO2 Probe Open Profile -0.02 Location noted at Tonzi in Open (East) profile at -2cm