Rec No. | Date | Loc From | Title | Loc To | Notes |
6480 |
2023-12-01 |
Factory none |
Received |
Lab none |
Received back from Licor. This 7550 sn AIU-1866 went back to Licor for a power problem. Licor replaced the 7550 main power supply under warranty, subjected the units to environmental tests and verified the operation of the units. It was shipped with 7500A sn 75H-2669 (nothing wrong with the 7500A, it was just the sensor attached when the box died) |
6468 |
2023-11-09 |
Lab none |
Shipped |
Factory none |
Shipped back to Licor for repair. It was sent with 7500A sn 75H-2669. |
6456 |
2023-10-26 |
Gilbert Tract Interface LI-7550 |
Removed |
Lab none |
Removed because of a blown fuse. We tried replacing the fuse, but it blew immediately. |
6294 |
2023-05-17 |
Lab none |
Installed |
Gilbert Tract Interface LI-7550 |
Installed at GT after factory repair to replace sn AIU-1039, the AMP 7550. |
6271 |
2023-04-27 |
Factory none |
Received |
Lab none |
Received back from Licor after repair. The main power supply was found defective and replaced. |
6231 |
2023-03-03 |
Lab none |
Shipped |
Factory none |
Shipped to Licor for repair |
6203 |
2023-03-02 |
Bouldin Corn Interface LI-7550 |
Removed |
Lab none |
Unit is receiving 24V but no LED lights and couldn't connect. Removed from BC tower and brought back to the lab for testing. |
5243 |
2021-04-15 |
Bouldin Corn Interface LI-7550 |
Upgrade |
Bouldin Corn Interface LI-7550 |
Upgraded firmware from v8.5.0 to v8.8.32 |
5090 |
2021-02-03 |
Lab none |
Installed |
Bouldin Corn Interface LI-7550 |
Installed on the BC tower to replace sensors damaged when the tower fell last week |
4104 |
2019-09-04 |
Bouldin Alfalfa-ReleaseEx Interface LI-7550 |
Removed |
Lab none |
Removed from Bouldin Alfalfa with the end of the CH4 release experiment. This unit was on the secondary tower |
4086 |
2019-08-16 |
Lab none |
Installed |
Bouldin Alfalfa-ReleaseEx Interface LI-7550 |
Installed for CH4 release experiment. This unit was part of a secondary, lower eddy system |
3987 |
2019-05-31 |
Factory |
Received |
Lab none |
Received back from Licor. They replaced the RF chip in the power supply. |
3981 |
2019-05-09 |
Lab none |
Shipped |
Factory |
Shipped back to Licor for them to inspect. |
3969 |
2019-05-01 |
Lab none |
Tested |
Lab none |
Plugged it in to the 24V wall power and it was dead. Will e-mail Licor. |
3972 |
2019-04-30 |
Tonzi-Tower Interface LI-7550 |
Removed |
Lab Interface LI-7550 |
Removed from Tonzi Tower due to power issues. Power light was flickering with audible relay click when a head was attached. Seemed to boot without head. Might have caused 5A Tower load breaker to pop on 4-23. Brought in for testing - replaced with Floor box sn AIU-1869 |
3369 |
2018-03-06 |
Tonzi-Tower Interface LI-7550 |
Upgrade |
Tonzi-Tower Interface LI-7550 |
Downgraded from embedded software version 8.7.5 to version 8.5.0 to standardize our systems. We had to downgrade other interfaces to 8.5.0 to solve the data-header mismatch we had with version 8.7.5. |
2973 |
2017-08-17 |
Tonzi-Tower |
Upgrade |
Tonzi-Tower |
The firmware was upgraded from v8.5.0 to v8.7.5 |
2766 |
2017-05-11 |
Lab none |
Installed |
Tonzi-Tower |
Installed on the Tower one level below the top on the southwest corner post. |