Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: cal_Vaisala_HMP45AC_snW0730036_20010307.pdfCalibration: cal_Vaisala_HMP45AC_snW0730036_20170414.pdf
PackingSlip: ps_Campbell_20010521.pdf
Report: SW AirT/RH comparison 2020 - 2020-09-10
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6185 | 2023-02-14 | Gilbert Tract RH & Temp | Removed | Lab none | Removed from GT tower because RH has been maxing out last winter. Replaced by Hygrovue10 sn E3452 |
5509 | 2021-09-02 | Sherman Wetland RH & Temp | Moved | Gilbert Tract RH & Temp | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
3853 | 2019-01-24 | Lab none | Installed | Sherman Wetland RH & Temp | Installed at SW to replace T2130005, which was brought back to the lab for inspection. |
3846 | 2019-01-23 | Lab none | Maintenance | Lab none | Spliced on a longer cable so total length is ~20'. New wiring colors are: Orange-Temperature, Blue-RH, White-SignalReference, Red-12V, Black-G, Bare/Shield-AG |
3682 | 2018-08-23 | Sherman Wetland RH & Temp | Removed | Lab none | Removed from SW because its cable was to short to allow it to be mounted in the correct place. |
3678 | 2018-08-07 | Lab none | Installed | Sherman Wetland RH & Temp | Installed at SW to replace HMP60 sn L2310736. The HMP60 was reading numbers lower than reasonable. |
3252 | 2017-12-19 | Twitchell Alfalfa | Removed | Lab RH & Temp | Removed from the HMP inter-comparison at Twitchell Alfalfa |
3270 | 2017-12-18 | Lab none | Installed | Twitchell Alfalfa | Installed at Twitchell Alfalfa for an inter-comparison of HMPs |
3176 | 2017-10-19 | East End-Temp1 RH & Temp | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
2891 | 2017-06-01 | Lab none | Installed | East End-Temp1 RH & Temp | Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
2746 | 2017-05-05 | Factory | Calibration | Lab none | Sensor received from Campbell Sci with Certificate of Calibration. Next calibration due 15-Apr-2018. |
2745 | 2017-04-03 | Lab none | Serviced | Factory | Shipped to Campbell Sci for calibration. CSI Technician notes:
Tested & inspected. Replaced o-rings. RH chip has corrosion damage. Took before readings with original RH chip. Replaced RH chip & calibrated with documentation. |
1837 | 2015-05-12 | Sherman Island RH & Temp | Removed | Lab none | Removed with the decommissioning of the Sherman Cow Pasture site. |
40 | 2006-11-13 | Factory | Calibration | Factory | Factory calibration |