Equipment and Calibration History

View Equipment |

View history entry 2745 for Vaisala HMP45AC sn: W0730036

Date: 2017-04-03
From: Lab - none
To: Factory -
Title: Serviced
Notes: Shipped to Campbell Sci for calibration. CSI Technician notes: Tested & inspected. Replaced o-rings. RH chip has corrosion damage. Took before readings with original RH chip. Replaced RH chip & calibrated with documentation.
Entered by: Daphne Szutu