Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6038 | 2022-10-20 | Lab none | Installed | Tonzi PAR Sensor Reflected | Installed at Tonzi to replace sn 030382 that has been spiky. |
6035 | 2022-10-19 | Experiment none | Removed | Lab none | I removed this sensor from the bike kit to use at Tonzi. I also spliced about 3m of extra length on to its cable. |
5932 | 2022-08-10 | Lab none | Installed | Experiment none | Installed on the back of Robert's e-bike for his urban heat island transect in Oakley |
5669 | 2022-02-02 | East End PAR Sensor Undercanopy B | Removed | Lab none | Removed from lower location south of the tower at East End. This sensor was disconnected from the data logger two weeks before |
5230 | 2021-03-30 | Lab none | Installed | East End none | Installed in EE wetland to measure undercanopy PAR. This sensor is S of the tower mounted on a grounding rod. The sensor is sitting in an open patch a few cm above the water surface. |
5113 | 2021-02-10 | Cal Kit PAR Sensor Incoming | Removed | Lab none | Removed from TZ Tower after three weeks of inter comparison with the AMP standard |
5057 | 2021-01-21 | Lab none | Installed | Cal Kit PAR Sensor Incoming | Installed at TZ as part of the inter-comparison kit to test PAR, T/RH on the top of the Tower. |
4935 | 2020-10-26 | Lab none | Repair | Lab none | The cable insulation on this sensor was rotted. I cut the cable off near the sensor and spliced on a new cable. I also added an aluminum bracket to provide strain relief and used liquid electrical tape to seal the hole where the cable enters the sensor. The white wire (+) had a nick in its insulation right at the point where it entered the sensor that I covered with heat shrink. |
4804 | 2020-09-29 | Vaira PAR Sensor Incoming Cal | Removed | Lab none | Removed from the tower (incoming cal) for repair - the cable is all rotted |
4803 | 2020-09-16 | Vaira PAR Sensor Incoming | Moved | Vaira PAR Sensor Incoming Cal | Moved from the main incoming position to the incoming cal position in preparation to being removed. |
33 | 2010-01-08 | Vaira PAR Sensor Incoming | Location | Vaira PAR Sensor Incoming | This instrument is installed in the upward diretion collecting incoming PAR. Calibration factor on instrument is 5.03 µV/µmol/s.m2 |