Equipment and Calibration History

View Equipment |

View history entry 4935 for Kipp & Zonen PAR-LITE sn: 990025

Date: 2020-10-26
From: Lab - none
To: Lab - none
Title: Repair
Notes: The cable insulation on this sensor was rotted. I cut the cable off near the sensor and spliced on a new cable. I also added an aluminum bracket to provide strain relief and used liquid electrical tape to seal the hole where the cable enters the sensor. The white wire (+) had a nick in its insulation right at the point where it entered the sensor that I covered with heat shrink.
Entered by: Joe Verfaillie