Rec No. | Date | Loc From | Title | Loc To | Notes |
2037 |
2015-11-03 |
Tonzi CO2 Probe Tree1 Profile -0.02 |
Removed |
Lab none |
Removed from the Tonzi Tree 1 profile at -2cm due to bad readings. Replaced by sn L2010069 |
2073 |
2014-12-09 |
Tonzi CO2 Probe Tree1 Profile -0.02 |
Location |
Tonzi CO2 Probe Tree1 Profile -0.02 |
Tonzi Tree1 profile at -2cm |
2023 |
2014-03-18 |
Tonzi CO2 Probe Tree1 Profile -0.04 |
Moved |
Tonzi CO2 Probe Tree1 Profile -0.02 |
At Tonzi in the Tree 1 profile this probe was moved from the -4cm location to -2cm |
1994 |
2014-01-29 |
Lab none |
Installed |
Tonzi CO2 Probe Tree1 Profile -0.04 |
Installed at Tonzi in the Tree 1 profile at -4cm with new U-shaped tops on the PVC enclosures |
1435 |
2014-01-27 |
Lab none |
Calibration |
Lab none |
Trueppm = Probeppm*1.287 + 388; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
1179 |
2014-01-24 |
Tonzi CO2 Probe Tree1 Profile -0.04 |
Removed |
Lab none |
Removed from Tonzi tree #1 at -4cm for lab calibration |
1131 |
2012-02-16 |
Lab none |
Installed |
Tonzi CO2 Probe Tree1 Profile -0.04 |
Installed at Tonzi in the Tree1 set at -4cm |
1406 |
2012-02-06 |
Lab none |
Calibration |
Lab none |
Trueppm = Probeppm*1.063 + -34.27; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan. |
1369 |
2010-03-02 |
Lab none |
Calibration |
Lab none |
Trueppm = Probeppm*1.078 + -11.78; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, 597ppm, and 2238ppm. Ta=22.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Jaclyn. |
4636 |
2010-01-08 |
Tonzi CO2 Probe Open Profile -0.04 |
Location |
Tonzi CO2 Probe Open Profile -0.04 |
Pulled out for ambient air reading and put back. |
4546 |
2008-03-06 |
Lab none |
Installed |
Tonzi CO2 Probe Tree1 Profile -0.04 |
Installed in the Wast under tree profile at -4cm. mV and ppm readings in air were recorded before installation. |