Rec No. | Date | Loc From | Title | Loc To | Notes |
6876 |
2024-08-21 |
Vaira Forced Diffusion West |
Removed |
Lab none |
Removed from Vaira West location due to bad CO2 readings |
5625 |
2021-12-02 |
Lab none |
Installed |
Vaira Forced Diffusion West |
Installed at Varia soil co2 west to replace the chamber that was removed last time for bad data. |
5590 |
2021-11-23 |
Mayberry Forced Diffusion Boom |
Removed |
Lab none |
Removed from MB floating boom and brought back to the lab for testing. It underwent a lot of power cycles because of a short in the system and I want to test it before we deploy it again. |
5402 |
2021-07-21 |
Lab none |
Installed |
Mayberry Forced Diffusion Boom |
Installed at Mayberry to replace sn FD20190027 because 27's numbers looked bad. |
5198 |
2021-03-15 |
Factory none |
Returned |
Lab none |
Returned from Eosense for high temp start problem and noisy CO2 readings. It had a failed clock crystal causing start up issues at high temperature, it was given custom firmware to bypass the component. It was calibrated for CO2 concentration and CO2 flux |
5197 |
2021-01-12 |
Lab none |
Shipped |
Factory none |
Sent to Eosense on RMA #210105 to check on its high temperature start up problem and some noise in the CO2 readings. |
5045 |
2020-12-18 |
Bouldin Alfalfa Forced Diffusion A |
Removed |
Lab none |
Removed from Bouldin Alfalfa. This one has high temperature start up issues and was running on a relay to restart it every night. I removed the relay too but did not update the CR1000 program. |
4830 |
2020-10-05 |
Lab none |
Installed |
Bouldin Alfalfa Forced Diffusion A |
Installed at Bouldin Alfalfa replacing sn FD20180048 that has low CO2 concentration readings. Also added a relay to restart this chamber at midnight to avoid high temperature start up problem. |
4776 |
2020-09-23 |
Bouldin Corn Forced Diffusion A |
Removed |
Lab none |
Removed and brought back to the lab for testing because it would not connect in the field |
4742 |
2020-09-10 |
Lab none |
Installed |
Bouldin Corn Forced Diffusion A |
Reinstalled after having been returned to the lab when it would not start due to being hot. (Note by Daphne 2021-08-11: I'm pretty sure this chamber was installed under the tower since we were waiting for our field to be harvested at this time). |
4736 |
2020-09-01 |
Bouldin Corn Forced Diffusion A |
Removed |
Lab none |
Removed for testing in the lab because it was not responding in the field. Probably its high temperature non-start problem. |
4628 |
2020-07-21 |
Lab none |
Installed |
Bouldin Corn Forced Diffusion A |
Installed in the normal density corn in front of the tower. |
4522 |
2020-06-11 |
Lab none |
Tested |
Lab none |
This unit was non-responsive in the field. In the lab it started right up and I was able to collect and clear its data. These units have had an issue with not starting up when they are too hot though this is supposed to be fixed in newer units. Maybe this unit has the too-hot-to-start problem and this combined with the power outage observed at Bouldin Corn to knock it out. Maybe I'll ask Camilo to put it back on Monday. |
4520 |
2020-06-09 |
Bouldin Corn Forced Diffusion A |
Removed |
Lab none |
Removed from the field because it was not responding. Cable tested okay. Maybe go too hot. |
4511 |
2020-05-29 |
Lab none |
Installed |
Bouldin Corn Forced Diffusion A |
Reinstalled in normal density corn just in front of tower |
4421 |
2020-04-17 |
Bouldin Corn Forced Diffusion A |
Removed |
Lab none |
Removed from BC and brought back to the lab because there will be tractor work on the field soon. |
4323 |
2020-01-07 |
Lab none |
Installed |
Bouldin Corn Forced Diffusion A |
Reinstalled at Bouldin Corn after being removed when the field was flooded. Installed near previous location on 12 inch collars. |
4732 |
2019-12-20 |
Bouldin Corn Forced Diffusion A |
Removed |
Lab none |
Removed from the field by Tyler due to being partially submerged by winter flooding. |
4731 |
2019-11-14 |
Lab none |
Installed |
Bouldin Corn Forced Diffusion A |
Installed in the corn field near Tyler's northern soil gas ports |