Equipment and Calibration History

History for Texas Electronics TE525MM 65548-615

Rec No.DateLoc FromTitleLoc ToNotes
5581 2021-11-18 Bouldin Corn Precipitation Calibration Bouldin Corn Precipitation I adjusted the screw 1 turn CCW before testing b/c I forgot to adjust it after the previous test. We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 49 tips and I called it good until next year.
5574 2021-11-05 Bouldin Corn Precipitation Tested Bouldin Corn Precipitation We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 52 tips but forgot to adjust the screws.
5541 2021-10-21 Bouldin Corn Precipitation Tested Bouldin Corn Precipitation We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 54 tips and adjusted its set screws 1.5 turns CCW (1/2 turn of both screws is a 2-3% adjustment).
5018 2020-12-03 Bouldin Corn Precipitation Tested Bouldin Corn Precipitation Plugged tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket (1/2 of the bottle at a time) which should result in 100 tips. Got 117 tips. Did not adjust - will retest with better water metering system.
4969 2020-11-19 Bouldin Corn Precipitation Bouldin Corn Precipitation Plugged tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100 tips. Got 116 tips. Did not adjust - will retest with better water metering system.
4306 2019-12-13 Bouldin Corn Precipitation Tested Bouldin Corn Precipitation Plugged tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100 tips. Got 102 tips and did not adjust the bucket.
4295 2019-12-13 Lab none Installed Bouldin Corn Precipitation Re-installed at Bouldin Corn after re-assembling tower.
4263 2019-12-02 Bouldin Corn Precipitation Removed Lab none Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26.
4119 2019-10-22 Bouldin Corn Precipitation Calibration Bouldin Corn Precipitation After cleaning the tipping bucket we connected it to a CR23x and dripped 475ml of water through the bucket which should result in 100tips. We got 99 tips – no adjustment needed.
3742 2018-10-18 Bouldin Corn Precipitation Calibration Bouldin Corn Precipitation Got 94 of 100 tips. Adjusted a total of 1 turn CW.

2018-10-18: Got 94 of 100 tips - adjusted screws 0.5 turns CW
2018-10-25: Got 97 tips and adjusted set screws 0.5 turn CW.
2018-11-01: 99 of 100 tips - no adjustment.
2726 2017-04-26 Lab none Installed Bouldin Corn Precipitation Installed on the top southeast corner of the scaffolding at the Bouldin Corn site with the initial set up of the tower.