Rec No. | Date | Loc From | Title | Loc To | Notes |
7004 |
2024-10-24 |
Lab none |
Installed |
Rush Ranch Interface |
Installed on the main tower at Rush Ranch |
6591 |
2024-02-28 |
West Pond Interface LI-7550 |
Removed |
Lab none |
Removed from WP as part of its decomissioning |
5051 |
2021-01-17 |
West Pond Interface LI-7550 |
Upgrade |
West Pond Interface LI-7550 |
Upgraded embedded software from v8.5.0 to v8.8.32 in hopes this will solve the persistent 7700 sync issue |
3316 |
2018-01-31 |
Lab none |
Installed |
West Pond Interface LI-7550 |
Installed at WP to replace AIU-0666 that was removed to fix door gasket, pressure port, circuit board cover. |
3308 |
2018-01-29 |
Lab none |
Upgrade |
Lab none |
Downgraded from embedded Licor software v8.7.5 to 8.5.0. Version 8.7.5 was causing issues with header/data mismatch. I checked 7550 time and time zone and CO2 and CH4 logging config. |
3107 |
2017-10-04 |
Bouldin Corn Interface LI-7550 |
Removed |
Lab none |
Removed from the Bouldin Corn tower for testing in the lab for bad headers and 7700 drop outs during wash cycle. Replaced by sn AIU-0614. |
3154 |
2017-08-23 |
Bouldin Corn Interface LI-7550 |
Upgrade |
Bouldin Corn Interface LI-7550 |
Upgraded the 7550 firmware from v8.0.0 to v8.7.5. I confirmed the 7550 time, time zone, sonic analog settings, and CO2 and CH4 logging config. |
2715 |
2017-04-26 |
Lab none |
Installed |
Bouldin Corn Interface LI-7550 |
Installed at the Bouldin Corn site with the initial set up of the tower. |
2364 |
2016-06-22 |
Factory |
Returned |
Lab none |
Returned from Licor RMA 00195802: "Inspected the unit and found the front control panel and the logic board corroded and the detector temperature control circuit inoperative. Replaced defective parts, removed corrosion, environmental tested and verified the operation [of] the unit." |
2363 |
2016-06-02 |
Lab none |
Shipped |
Factory |
Sent to Licor on RMA 00195802 for repair. The connector board on the bottom of the box got wet and the pins of the power connecters rotted. Licor says the board is all one piece and probably will have to be replaced. |
2138 |
2015-12-23 |
Twitchell Rice-Trailer |
Removed |
Lab none |
Removed with the temporary tower set up near the building at the rice project for the LGR COS experiment. This box had a significant amount of water inside. There was some corrosion on the power connectors but it looks like it was running up until I turned it off. I blew out all the connectors and stuffed it full of desiccant in the lab. |
2116 |
2015-12-08 |
Lab none |
Installed |
Twitchell Rice-Trailer |
Installed near the trailer at Twitchell Rice as a temporary site to test a COS analyzer from LGR |