Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP222
Serial No:K1310102
Tag No:
Acquisition Date:2014-04-03
Project:DOE
Location on 2022-10-05:Lab,
Notes: 0-1% range

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
6030 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.16 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -16cm
4172 2019-11-19 Lab none Installed Tonzi CO2 Probe Open Profile -0.16 Installed in the deepest (-16cm) location in the Open soil co2 profile to replace a 3% probe that was railed. This is a 1% probe.
3207 2017-11-29 Vaira CO2 probe East Profile -0.04 Removed Lab none Removed from the 4cm location in the East profile at Vaira. This probe was giving erratic values
3122 2017-10-05 Lab none Installed Vaira CO2 probe East Profile -0.04 Installed at Vaira in the East profile at the -4cm location
3009 2017-08-25 Tonzi CO2 Probe Open Profile -0.02 Removed Lab none Removed from Tonzi Open profile at -2 cm position for calibration
2188 2016-02-23 Lab none Installed Tonzi CO2 Probe Open Profile -0.02 Installed in the Tonzi Open profile at -2cm after lab calibration.
2157 2016-02-04 Tonzi CO2 Probe Open Profile -0.02 Removed Lab none Removed from Tonzi Open profile at -2cm for calibration
2070 2014-12-09 Tonzi CO2 Probe Open Profile -0.02 Location Tonzi CO2 Probe Open Profile -0.02 Tonzi Open profile at -2cm
2009 2014-04-22 Lab none Installed Vaira CO2 probe West Profile -0.04 Installed in the Varia West profile at -4cm replacing sn Y1720006
2017 2014-04-08 Lab none Installed Tonzi CO2 Probe Open Profile -0.08 Installed at Tonzi in the Open profile at -8cm. not clear if it replaced another probe.
1445 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.014 + 42.61; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.