Equipment and Calibration

View History

Equipment Entry

Instrument:Interface
Manufacture:Licor
Model No:LI-7550
Serial No:AIU-0650
Tag No:NULL
Acquisition Date:2012-02-07
Project:DWR-Wetland
Location on 2021-03-25:Hill Slough, NULL
Notes: Part of GHG1 system with LI-7700 sn TG1-0223 and LI-7500A sn 75H-2176

Documents

Calibration: cal_Licor_LI-7550_snAIU-0650_20111005.pdf
PackingSlip: ps_Licor_20120207.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
5224 2021-03-25 Hill Slough Interface LI-7550 Upgrade Hill Slough Interface LI-7550 The firmware was upgraded from v8.5.5 to v8.8.32. No problems noticed.
5180 2021-03-11 Lab none Installed Hill Slough Interface LI-7550 Installed at Hill Slough with site setup. Renamed from SWTT-0650 to HS-0650.
4953 2020-11-12 Sherman Wetland-Temp1 Interface LI-7550 Removed Lab none Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup.
4335 2020-01-30 East Pond Gas Analyzer CO2/H2O Moved Sherman Wetland-Temp1 Gas Analyzer CO2/H2O Removed with the decommissioning of the EP tower and installed at the new SW Temporary site 1250m upwind of the Permanent Tower.
3417 2018-04-19 Lab none Installed East Pond Gas Analyzer CO2/H2O Installed on the East Pond tower to replace sn AIU-0666 that is supposed to be used in the lab for calibrations.
3166 2017-10-19 East End-Temp1 Interface LI-7550 Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
2812 2017-06-01 Lab none Installed East End-Temp1 Interface LI-7550 Installed at open water temporary tower in East End wetland
2769 2017-05-15 Lab none Upgrade Lab none Updated the embedded software in the LI-7550 from 8.0.0 to 8.5.0.
2757 2017-05-11 Vaira Removed Lab none Removed from Vaira because the Ameriflux intercomparison finished and because we received new 7500A and sonics from Ameriflux. This interface was replaced by interface AIU-1870.
2700 2017-04-10 Twitchell Rice Interface LI-7500 Installed Vaira Interface LI-7500 Added to the Vaira eddy tower for the Ameriflux inter-comparison and the transition to newer LI-7500A eddy systems.
2237 2016-04-13 Twitchell Rice Interface LI-7500 Upgrade Twitchell Rice Interface LI-7500 Updated the 7550 firmware from v7.3.0 to v8.0.0 skiping version 7.5.0
Relevant 7.5.0 changes: Main Window is now resizable, Added serial number entry for sonics, Help update, Fixed timeout when saving config files
Relevant 8.0.0 changes: Separate bandwidth settings for analog inputs and gas channels, Detection of new RS instruments, Additional calibration coefficients, Added sonic software version entry, Added chopper temp logging
1590 2014-10-02 Twitchell Rice Interface LI-7500 Upgrade Twitchell Rice Interface LI-7500 Updated the 7550 firmware from v7.0.1 to v7.3.0. Fixed bug that can cause the instrument to restart unexpectedly, added 7700 data to interface can view/log power Vin, other minor bug fixes.
1589 2014-03-07 Twitchell Rice Interface LI-7500 Upgrade Twitchell Rice Interface LI-7500 Updated the 7550 firmware from v6.5.5 to v7.0.1. Improved interface, support for SmartFlux, bug fixes.
1142 2013-11-21 Twitchell Rice Interface LI-7500 Upgrade Twitchell Rice Interface LI-7500 The firmware of the 7550 box was upgraded from v6.5.2 to v6.5.5 to fix bugs with the USB memory overwrite and synchronizing the 7700 clock.
873 2013-01-03 Lab none Installed Twitchell Rice Interface LI-7500 Istalled on the Twitchell Rice tower to replace AIU-0674 for repair of the condensation problem.
866 2012-12-21 Factory Returned Lab none Returned from Licor with 75H-2180 after repair for the condensation problem. The external connector circuit board at the bottom of the box has been completely sealed. New firmware v6.0.0 has been loaded on the box and requires new Windows interface software. Pressure Sensor sn PX-2820 A0=57.729, A1=15.323.
856 2012-11-20 Lab none Repair Factory Returned to Licor to for repair of condensation problem. This unit did not fail in the field, but showed corrosion on the external connector pins.
850 2012-10-08 West Pond Interface LI-7550 Removed Lab none Removed from West Pond to be returned to Licor for repair of condensation probelm. This unit had not failed yet. Replaced by sn 0397.
584 2012-05-02 Lab none Installed West Pond Interface LI-7550 Installed on the West Pond tower at mid-level on the scaffolding.