History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5296 | 2021-06-02 | Tonzi-Floor Interface LI-7550 | Upgrade | Tonzi-Floor Interface LI-7550 | The firmware was upgraded from v8.5.0 to v8.8.32 for standardization with the Delta Licor systems |
3979 | 2019-05-07 | Lab none | Installed | Tonzi-Floor Interface LI-7500 | Installed on the Floor tower to replace the one that was moved to Tonzi Tower last week. |
3416 | 2018-04-19 | East Pond Gas Analyzer CO2/H2O | Removed | Lab none | Removed from East Pond because it is supposed to be use in the lab for calibrations. |
3400 | 2018-04-17 | Lab none | Installed | East Pond Gas Analyzer CO2/H2O | Installed at EP tower as part of initial site set up. |
3356 | 2018-03-09 | Lab none | Maintenance | Lab none | Cleaned out the box of bug detritus and replaced the gasket on the front lid. This box is missing the pressure port (the hole is covered by electrical tape) and the reset button is missing, so this box should stay in the lab for now for calibration purposes. |
3315 | 2018-01-31 | West Pond Interface LI-7550 | Removed | Lab none | Brought back to the lab for some TLC. It was replaced by AIU-1570. |
3306 | 2018-01-16 | West Pond Interface LI-7550 | Upgrade | West Pond Interface LI-7550 | Downgraded embedded Licor software from v8.7.5 to 8.5.0 because 8.7.5 caused mismatch problems between data columns and headers |
2240 | 2016-04-13 | West Pond Interface LI-7550 | Upgrade | West Pond Interface LI-7550 | Updated the 7550 firmware from v7.3.0 to v8.0.0 skiping version 7.5.0 Relevant 7.5.0 changes: Main Window is now resizable, Added serial number entry for sonics, Help update, Fixed timeout when saving config files Relevant 8.0.0 changes: Separate bandwidth settings for analog inputs and gas channels, Detection of new RS instruments, Additional calibration coefficients, Added sonic software version entry, Added chopper temp logging |
1592 | 2014-10-02 | West Pond Interface LI-7550 | Upgrade | West Pond Interface LI-7550 | Updated the 7550 firmware from v7.0.1 to v7.3.0. Fixed bug that can cause the instrument to restart unexpectedly, added 7700 data to interface can view/log power Vin, other minor bug fixes.
|
1542 | 2014-09-08 | Lab none | Installed | West Pond Interface LI-7550 | Installed on the West Pond tower with head sn 75H-2180. There was now 7550-7500A pair on the tower because of issues with swapping the head on 9/4. This box has a new door gasket. |
1591 | 2014-02-28 | Lab none | Upgrade | Lab none | Upgraded firmware from v6.5.5 to v7.0.1 - FPGA v6.0.3 not needing update. Improved interface, support for SmartFlux, bug fixes. |
1055 | 2013-08-21 | West Pond Interface LI-7550 | Removed | Lab none | Removed from the West Pond tower. This box has been suffer an ant infestation for many weeks. It was removed to be de-ant-ified, cleaned up and tested in the lab. |
875 | 2013-01-03 | Lab none | Installed | West Pond Interface LI-7550 | Installed on the West Pond tower with head 0065 to replace AIU-0397 for repair of condensation problem. |
868 | 2012-12-21 | Factory | Returned | Lab none | Returned from Licor with 75H-0065 after repair of condensation problem. The external connector board has been completely sealed and new firmware v6.0.0 has been loaded to the box. Pressure Sensor sn PX-2763: A0=57.595, A1=15.231 |
849 | 2012-11-02 | Lab none | Shipped | Factory | Shipped to Licor for repair of condensation problem, RMA 172095 |
829 | 2012-10-30 | Mayberry Gas Analyzer CO2/H2O | Removed | Lab none | Removed from the Mayberry Tower due to corrosion on the connector circuit board. Replaced by sn 0614 |
547 | 2012-04-24 | Lab none | Installed | Mayberry Gas Analyzer CO2/H2O | Installed on the Mayberry eddy tower to swap out LI-7500 sn 1594 |