Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: cal_KippZonen_CNR4_sn121038_20191117.pdfReport: SW 4-way comparison - 2021 - 2021-02-25
Report: Bouldin Corn - Radiometer Calibration 2021 - 2021-06-09
Report: Hill Slough - Radiometer Calibration 2021 - 2021-06-22
Report: West Pond - Radiometer Calibration 2021 - 2021-06-22
Report: East End - Radiometer Calibration 2021 - 2021-07-21
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5561 | 2021-11-08 | Lab none | Returned | Returned none | Returned to Sebastien of AMP in exchange for a freshly calibrated 4-way radiometer CNR4 sn 120788 |
5569 | 2021-11-05 | Cal Kit Net Radiometer 4-way | Removed | Lab none | Removed from BA tower at the conclusion of the rad calibration and brought back to the lab. This sensor will go back to AMP. |
5535 | 2021-10-21 | Cal Kit Net Radiometer 4-way | Moved | Cal Kit Net Radiometer 4-way | Moved from GT to BA as part of the radiometer cal kit to begin the calibration at BA |
5457 | 2021-09-15 | Mayberry | Moved | Gilbert Tract none | Moved from MB to the new Gilbert Tract site to continue the radiometer intercomparison there. |
5425 | 2021-08-17 | East End | Moved | Mayberry none | Moved from EE to MB after completion of radiometer intercomparison at EE |
5389 | 2021-07-21 | West Pond | Moved | East End none | Moved from West Pond to East End at the conclusion of the radiometer intercomparison at WP |
5346 | 2021-06-22 | Hill Slough | Moved | West Pond none | Moved from HS to WP as part of the radiometer calibration kit at the conclusion of the 2-week intercomparison at HS. |
5321 | 2021-06-09 | Cal Kit Net Radiometer 4-way | Moved | Cal Kit Net Radiometer 4-way | Moved from BC to HS as part of rad cal kit |
5308 | 2021-05-27 | Cal Kit Net Radiometer 4-way | Moved | Cal Kit Net Radiometer 4-way | Moved to Bouldin Corn (BC) as part of the cal kit |
5205 | 2021-03-16 | Sherman Wetland Net Radiometer 4-way Cal B | Removed | Lab none | Removed from SW at the conclusion of the 4-way intercomparison test and brought back to the lab |
5138 | 2021-02-26 | Factory none | Installed | Sherman Wetland Net Radiometer 4-way Cal B | Installed on the Sherman Wetland tower for inter-comparison. This is the AMP standard. It is clamped near the end of the square tube of the radiation boom pointing south. |
5119 | 2019-11-17 | Factory none | Calibration | Factory none | Initial calibration by Kipp and Zonen. LW on 2019-11-17: LW_in = 6.83 uV/(W/m2), LW_out = 5.48 uV/(W/m2). SW on 2019-11-15: SW_in = 14.56 uV/(W/m2), SW_out = 14.17 uV/(W/m2) |