Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP222
Serial No:Z2720011
Tag No:
Acquisition Date:0000-00-00
Project:DOE-TCP
Location on 2015-06-09:Lab, -8cm
Notes: 0-1% range

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
1885 2015-06-09 Twitchell Alfalfa Removed Lab none Removed from forced diffusion ambient cell with upgrade of soil CO2 system
1599 2014-10-09 Lab none Installed Twitchell Alfalfa Installed in the FD ambient chamber to replace sn Z2720014 that had high readings.
1524 2014-08-27 Twitchell Alfalfa Removed Lab none Removed from Forced Diffusion probe 1 because it was reading zero. Replaced by sn Z2720014
1867 2014-08-13 Lab none Installed Twitchell Alfalfa Installed near the radiometers at Alfalfa in forced diffusion ambient cell
1460 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*1.176 + -287.8; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
1302 2014-05-13 Sherman Island Removed Lab none Removed from Sherman Set B at -2cm to be replaced by a 3% probe sn F4960019.
1243 2014-01-29 Lab none Installed Sherman Island Reinstalled after lab calibration Set B at -2cm
1434 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*1.11 + 190.2; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.
1180 2014-01-24 Tonzi CO2 Probe Tree1 Profile -0.08 Removed Lab none Removed from Tonzi tree #1 at -8cm for lab calibration
1132 2012-02-16 Lab none Installed Tonzi CO2 Probe Tree1 Profile -0.08 Installed at Tonzi in the Tree1 set at -8cm
1407 2012-02-06 Lab none Calibration Lab none Trueppm = Probeppm*0.9639 + -59.19; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan.
1368 2010-03-02 Lab none Calibration Lab none Trueppm = Probeppm*0.9871 + -73.58; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, 597ppm, and 2238ppm. Ta=22.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Jaclyn.
4637 2010-01-08 Tonzi CO2 Probe Open Profile -0.08 Location Tonzi CO2 Probe Open Profile -0.08 Pulled out for ambient air reading and put back.
4547 2008-03-06 Lab none Installed Tonzi CO2 Probe Tree1 Profile -0.08 Installed in the Wast under tree profile at -8cm. mV and ppm readings in air were recorded before installation.
2026 2005-02-17 Tonzi CO2 Probe Open Profile -0.04 Location Tonzi CO2 Probe Open Profile -0.04 Location noted at Tonzi in Open (East) profile at -4cm