Rec No. | Date | Loc From | Title | Loc To | Notes |
5006 |
2020-12-01 |
Vaira CO2 probe East Profile -0.02 |
Removed |
Lab none |
The battery bank has died and it has been decided to focus on the forced diffusion soil CO2 and Eosense GP profile. The rest of the soil co2 equipment will be removed. |
3126 |
2017-09-21 |
Vaira CO2 probe North Profile -0.04 |
Moved |
Vaira CO2 probe East Profile -0.02 |
Moved from North profile at -4cm to East profile at -2cm |
3070 |
2017-09-14 |
Lab none |
Installed |
Vaira CO2 probe East Profile -0.04 |
Reinstalled at Vaira in the North profile at -4cm after lab calibration. |
3028 |
2017-08-25 |
Vaira CO2 probe East Profile -0.04 |
Removed |
Lab none |
Removed from Vaira from East profile -4 cm position for calibration |
2175 |
2016-02-23 |
Lab none |
Installed |
Vaira CO2 probe East Profile -0.04 |
Installed in the Vaira East profile at -4cm after lab calibration. |
2158 |
2016-02-04 |
Tonzi CO2 Probe Open Profile -0.04 |
Removed |
Lab none |
Removed from Tonzi Open profile at -4cm for calibration |
2071 |
2014-12-09 |
Tonzi CO2 Probe Open Profile -0.04 |
Location |
Tonzi CO2 Probe Open Profile -0.04 |
Tonzi Open profile at -4cm |
2007 |
2014-04-22 |
Lab none |
Installed |
Vaira CO2 probe West Profile -0.02 |
Installed in the Varia West profile at -2cm replacing sn Y0850004 |
2016 |
2014-04-08 |
Lab none |
Installed |
Tonzi CO2 Probe Open Profile -0.04 |
Installed at Tonzi in the Open profile at -4cm. not clear if it replaced another probe. |
1446 |
2014-04-07 |
Lab none |
Calibration |
Lab none |
Trueppm = Probeppm*1.094 + 0.7501; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |