Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: cal_KippZonen_NRLite_sn990350_19990729.pdfReport: Oxford Tract Radiometer Comparison - 09/2018 - 2019-09-16
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6089 | 2022-11-02 | Bouldin Alfalfa-ADEX Net Radiometer Edge | Removed | Lab none | Removed from top of the Edge tower at 3.34m alg at the end of the 2022 advection experiment at BA |
5851 | 2022-04-25 | Lab none | Installed | Bouldin Alfalfa-ADEX Net Radiometer Edge | Installed on top of the Edge tower at 334m alg for the advection experiment at BA |
5757 | 2022-03-31 | Hill Slough Net Radiometer | Removed | Lab none | Removed from HS because we are done comparing the NR01 data to it. Planning to use this sensor for the advection setup this summer. |
5183 | 2021-03-11 | Lab none | Installed | Hill Slough Net Radiometer | Installed at Hill Slough with site setup. Temporary Rnet measurement until we have finished testing the 4-way radiometer. |
4192 | 2019-10-01 | Lab none | Calibration | Lab none | New calibration factor of 13.0 uV/(W/m2).
Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 09/2018" for details. https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=27 |
4022 | 2019-07-25 | Mayberry Net Radiometer Water | Removed | Lab none | Removed from MB (Rnet_water) and brought back to the lab for repair (broken cable). It was replaced by sn 990351. |
326 | 2011-03-01 | Lab none | Installed | Mayberry Net Radiometer Water | In stalled on the Mayberry tule tower over the water. |
289 | 2011-02-03 | Tonzi | Removed | Lab none | Removed from under the trees near the floor tower. |
101 | 2010-03-16 | Lab none | Installed | Tonzi | Installed at the request of Hideki under the oak canopy near the radiation tram at Tonzi. |
3807 | 2002-02-08 | Lab none | Installed | Tonzi Gas Analyzer CO2/H2O | Installed inside the orange fence on the floor met CR10x data logger. |
4193 | 1999-07-29 | Factory | Calibration | Factory | Initial Factory calibration 13.5uV/W/m2 |