Rec No. | Date | Loc From | Title | Loc To | Notes |
5556 |
2021-11-04 |
Hill Slough Precipitation |
Adjusted |
Hill Slough Precipitation |
The screws were adjusted 1/2 turn in (CW) before testing. The test got 51/50 tips - okay. |
5778 |
2021-10-26 |
Hill Slough Precipitation |
Tested |
Hill Slough Precipitation |
Got 53/50 tips. I meant to turn the screws 1 turn CCW but I think I may have turned them 2 turns CCW. Needs a retest next time anyways. |
5216 |
2021-03-16 |
Lab none |
Installed |
Hill Slough Precipitation |
Installed on the SE corner of the scaffolding |
5040 |
2020-12-18 |
Tonzi none |
Removed |
Lab none |
Removed from OSU SP1 for experiments in the lab. The cable was disconnected inside the tipping bucket and left in place in the field. |
4125 |
2019-10-24 |
Tonzi |
Calibration |
Tonzi |
I connected the bucket to a CR23x dripping 475ml of water through the bucket which should result in 100tips. I got 104 tips. I did not adjust it. |
3737 |
2018-10-17 |
Tonzi |
Calibration |
Tonzi |
Got 88 of 100 tips - adjusted screws 2.5 turns CW |
3657 |
2012-01-09 |
Lab none |
Installed |
Tonzi |
Installed on the ground at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ. |