Rec No.DateLoc FromTitleLoc ToNotes
5556 2021-11-04 Hill Slough Precipitation Adjusted Hill Slough Precipitation The screws were adjusted 1/2 turn in (CW) before testing. The test got 51/50 tips - okay.
5778 2021-10-26 Hill Slough Precipitation Tested Hill Slough Precipitation Got 53/50 tips. I meant to turn the screws 1 turn CCW but I think I may have turned them 2 turns CCW. Needs a retest next time anyways.
5216 2021-03-16 Lab none Installed Hill Slough Precipitation Installed on the SE corner of the scaffolding
5040 2020-12-18 Tonzi none Removed Lab none Removed from OSU SP1 for experiments in the lab. The cable was disconnected inside the tipping bucket and left in place in the field.
4125 2019-10-24 Tonzi Calibration Tonzi I connected the bucket to a CR23x dripping 475ml of water through the bucket which should result in 100tips. I got 104 tips. I did not adjust it.
3737 2018-10-17 Tonzi Calibration Tonzi Got 88 of 100 tips - adjusted screws 2.5 turns CW
3657 2012-01-09 Lab none Installed Tonzi Installed on the ground at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.