Equipment and Calibration History

History for Vaisala GMP343 N2210009

Rec No.DateLoc FromTitleLoc ToNotes
6746 2024-06-19 Lab none Installed Tonzi CO2 Probe Tower Profile 6 Reinstalled in location #3 (6m) of the Tower co2 profile.
6740 2024-06-18 Factory none Received Lab none Received after repair at Vaisala
6741 2024-05-28 Factory none Calibration Factory none Device failed in calibration due to faulty sensor.
Action(s) taken
Sensor module, body of the GMP343 and filter changed.
Operation tested, adjustment made and calibrated.
Calibration certificate H30-24220020 issued.
6739 2024-05-10 Lab none Shipped Factory none Shipped to Vaisala on RMA 00673130
6695 2024-04-29 Lab none Tested Lab none I connected a DB9 connector white to pin 2, green to pin 3, and power to a lithium power pack, pink to +12V and gray to G and a spare brown wire between G and pin5 on the DB9. This allowed me to communicate with the probe via a serial terminal. I managed to get it to tell me its error notices and it confessed that its IR source has failed. Vaisala's web site has a automated repair quote process that had a flat repair fee for about 0 and a calibration fee of about 0. I'll send it for repair.
6686 2024-04-22 Tonzi CO2 Probe Tower Profile 6 Removed Lab none Not sending sensible data. Not sure why. Brought in for testing.
6582 2024-02-27 Lab none Installed Tonzi CO2 Probe Tower Profile 6 Reinstalled in the upper-mid location of the Tower co2 profile after testing in the lab.
6562 2024-02-09 Tonzi CO2 Probe Tower Profile 6 Removed Lab none Removed from the upper mid location on the Tower due to bad readings and to be calibrated
3295 2018-01-15 Lab none Installed Tonzi Installed at the 3rd lowest location in the tower co2 profile - not running yet due to data logger failure.