5088 | 2021-02-03 | Lab none | Installed | Bouldin Corn Anemometer | Installed on the BC tower to replace sensors damaged when the tower fell last week |
4950 | 2020-11-12 | Sherman Wetland-Temp1 Anemometer | Removed | Lab none | Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup. |
4332 | 2020-01-30 | East Pond Anemometer | Moved | Sherman Wetland-Temp1 Anemometer | Removed with the decommissioning of the EP tower and installed at the new SW Temporary site 1250m upwind of the Permanent Tower. |
3560 | 2018-06-28 | East Pond Anemometer | Moved | East Pond Anemometer | Sensor moved up 26cm to try to address rotation angle issues. The water level was 68cm and the sensor was 540cm above water level. |
3403 | 2018-04-17 | Lab none | Installed | East Pond Anemometer | Installed at EP tower as part of initial site set up. |
3169 | 2017-10-19 | East End-Temp1 Anemometer | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
2809 | 2017-06-01 | Lab none | Installed | East End-Temp1 Anemometer | Installed at open water temporary tower in East End wetland |
2614 | 2017-02-10 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug. Old SN: Y153803 New SN: X153803 Upgrade file: Y153802-Y153806-m5.gcu Old firmware: 2329-601-03 New firmware: 2329-701-01 All Checksum fields report PASS. Log file: WMUpgradeLog_Y153803_170310_132625.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Output rate : P1 1 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X153803 ---- SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-701 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound Synchronised Polling : Off |
2604 | 2017-01-31 | Sherman Wetland Anemometer | Removed | Lab none | Removed from Sherman Wetland tower to be replaced with sonic sn 120906 which had a firmware update fixing the w-bug |
2287 | 2016-05-24 | Lab none | Installed | Sherman Wetland Anemometer | Installed on the Sherman Wetland tower at the initial set up of the site. |