Equipment and Calibration History

History for Gill WM 1590 153803

Rec No.DateLoc FromTitleLoc ToNotes
5088 2021-02-03 Lab none Installed Bouldin Corn Anemometer Installed on the BC tower to replace sensors damaged when the tower fell last week
4950 2020-11-12 Sherman Wetland-Temp1 Anemometer Removed Lab none Removed with decommissioning of Sherman Wetland Temp tower. Brought back to the lab for cleanup.
4332 2020-01-30 East Pond Anemometer Moved Sherman Wetland-Temp1 Anemometer Removed with the decommissioning of the EP tower and installed at the new SW Temporary site 1250m upwind of the Permanent Tower.
3560 2018-06-28 East Pond Anemometer Moved East Pond Anemometer Sensor moved up 26cm to try to address rotation angle issues. The water level was 68cm and the sensor was 540cm above water level.
3403 2018-04-17 Lab none Installed East Pond Anemometer Installed at EP tower as part of initial site set up.
3169 2017-10-19 East End-Temp1 Anemometer Removed Lab none Removed with decommissioning of East End Open Water tower (temporary tower 1)
2809 2017-06-01 Lab none Installed East End-Temp1 Anemometer Installed at open water temporary tower in East End wetland
2614 2017-02-10 Lab none Upgrade Lab none Upgraded firmware to correct for W bug.

Old SN: Y153803
New SN: X153803
Upgrade file: Y153802-Y153806-m5.gcu
Old firmware: 2329-601-03
New firmware: 2329-701-01

All Checksum fields report PASS.

Log file: WMUpgradeLog_Y153803_170310_132625.txt

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Output rate : P1 1 Hz
Output rate : P8 20 Hz

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER========================

Serial Number : X153803

---- SW VERSION--------------------------------------------
Anemometer Firmware Version: 2329-701

---- HW DETAILS OF ANEMOMETER------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER----------------------
Baud Rate : 19200
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : SOS
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Speed of Sound
Synchronised Polling : Off


2604 2017-01-31 Sherman Wetland Anemometer Removed Lab none Removed from Sherman Wetland tower to be replaced with sonic sn 120906 which had a firmware update fixing the w-bug
2287 2016-05-24 Lab none Installed Sherman Wetland Anemometer Installed on the Sherman Wetland tower at the initial set up of the site.