Equipment and Calibration History

History for Vaisala GMP222 Y1950005

Rec No.DateLoc FromTitleLoc ToNotes
2043 2015-11-03 Vaira CO2 probe North Profile -0.08 Removed Lab none Removed from Vaira North profile at -8cm due to bad readings. Replaced by sn L2010070
2064 2014-12-09 Vaira CO2 probe North Profile -0.08 Location Vaira CO2 probe North Profile -0.08 Vaira North profile at -8cm - may have been at -4cm
2013 2014-04-22 Vaira CO2 probe North Profile -0.04 Moved Vaira CO2 probe North Profile -0.08 At Vaira North profile, moved from -4cm to -8cm.
1997 2014-01-29 Lab none Installed Vaira CO2 probe North Profile -0.04 Installed at Vaira in the North profile as -4cm with new U-shaped tops on the PVC enclosures.
1444 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*1.073 + -174.1; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.
1190 2014-01-24 Vaira CO2 probe North Profile -0.04 Removed Lab none Removed from Vaira north plot at -4cm for lab calibration
1124 2012-03-21 Lab none Installed Vaira CO2 probe North Profile -0.04 Installed at Vaira in the north set at -4cm
1399 2012-01-30 Lab none Calibration Lab none Trueppm = Probeppm*1.08 + -104.4; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=27.4C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan.
1372 2011-02-01 Lab none Calibration Lab none Trueppm = Probeppm*1.079 + -120.4; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, and 597ppm. Ta=25.5C, P=102.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie.
243 2010-10-27 Sherman Island Removed Lab none This probe was removed from the Sherman site for testing because it was reporting >6V at 1.23% which was inconsistent with the label on the transmitter that said it was a 3% probe. However, this is a 1% probe so it is probably fine.
1355 2010-02-03 Lab none Calibration Lab none Trueppm = Probeppm*1.103 + -98.76; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, 597ppm, and 2238ppm. Ta=24.1C, P=101.2kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Jaclyn.
5 2009-12-01 Lab none Calibration Lab none Vaisala factory calibration 0-1% - certificate on file
Problem: Out of tolerance.
Action: Updated coefficients and adjusted unit.