History for Vaisala GMP221 K1340006
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6029 | 2022-10-05 | Tonzi-SoilCO2 CO2 Probe Open Profile -0.08 | Removed | Lab none | The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -8cm |
3056 | 2017-09-14 | Lab none | Installed | Tonzi CO2 Probe Open Profile -0.08 | Installed at Tonzi in the Open profile at -8cm after lab calibration. This is a new installation with new holes dug. |
3011 | 2017-08-25 | Tonzi CO2 Probe Open Profile -0.08 | Removed | Lab none | Removed from Tonzi Open profile at -8 cm position for calibration |
2190 | 2016-02-23 | Lab none | Installed | Tonzi CO2 Probe Open Profile -0.08 | Installed in the Tonzi Open profile at -8cm after lab calibration. |
1870 | 2014-09-08 | Twitchell Alfalfa | Removed | Lab none | Removed from Alfalfa Set B -8cm because of low readings. Replaced by F4960018 |
1529 | 2014-08-27 | Lab none | Installed | Twitchell Alfalfa | Installed in profile B at 8cm to replace sn F4960018 due to zero reading. |
2093 | 2014-06-17 | Sherman Island | Removed | Sherman Island | Removed from Sherman profile A at -16cm due to low readings. Replaced by sn G0150037 |
2092 | 2014-04-10 | Lab none | Installed | Sherman Island | Installed in Sherman profile A at -16cm to replace sn G0150035 |
1453 | 2014-04-08 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.002 + -37.28; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.1C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |