Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP222
Serial No:K1350101
Tag No:
Acquisition Date:2014-04-03
Project:DOE
Location on 2020-12-01:Lab,
Notes: 0-1% range

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
5003 2020-12-01 Vaira CO2 probe North Profile -0.04 Removed Lab none The battery bank has died and it has been decided to focus on the forced diffusion soil CO2 and Eosense GP profile. The rest of the soil co2 equipment will be removed.
3123 2017-10-05 Lab none Installed Vaira CO2 probe North Profile -0.04 Installed at Vaira in the North profile at the -4cm location
3024 2017-08-25 Vaira CO2 probe West Profile -0.08 Removed Lab none Removed from Vaira from West profile -8 cm position for calibration
2179 2016-02-23 Lab none Installed Vaira CO2 probe West Profile -0.08 Installed in the Vaira West profile at -8cm after lab calibration.
2153 2016-02-04 Vaira CO2 probe West Profile -0.08 Removed Lab none Removed from Vaira West profile at -8cm for calibration
2063 2014-12-09 Vaira CO2 probe West Profile -0.08 Location Vaira CO2 probe West Profile -0.08 Vaira West profile at -8cm
2011 2014-04-22 Vaira CO2 probe West Profile -0.02 Installed Lab none Installed in the Varia West profile at -2cm replacing sn G0230062
1447 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.014 + 51.82; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.